cobas® NHP Negative Control Kit 1.0Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00875197005127), received between Sep 18, 2025 and Sep 19, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Positive Result2
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 19, 2025False Positive Result; False Positive Result

    THE INITIAL REPORTER ALLEGED UNEXPECTED REACTIVE RESULTS FOR HIV WHEN USING THE COBAS MPX ASSAY ON THE COBAS 6800 INSTRUMENT. THE CUSTOMER REPORTED TWO SAMPLES THAT WERE INITIALLY REACTIVE FOR HIV WITH CYCLE THRESHOLD (CT) VALUES OF 39.1 AND 42.2. THE HIV GROWTH CURVE FOR BOTH SAMPLES SHOWED LATE AND MINIMAL AMPLIFICATION. SEROLOGY TESTING FOR BOTH SAMPLES WAS NON-REACTIVE. THE COBAS MPX ASSAY WAS REPEATED ON THE SAME SAMPLES, AND THE RESULTS WERE NON-REACTIVE. THE SAMPLES WERE REPORTED AS REACTIVE, AND THE BLOOD DONATIONS WERE DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23094355 ↗
  • MalfunctionSep 18, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE INITIAL REPORTER ALLEGED UNEXPECTED REACTIVE RESULTS FOR HIV IN BOTH SAMPLE POOLS AND NEGATIVE CONTROLS WHEN USING THE COBAS® 6800/8800 MPX ASSAY AND KIT COBAS 6800/8800 NHP NEG RMC IVD ON THE COBAS 8800 ANALYZER. TWO SAMPLE POOLS WERE REPORTED AS REACTIVE FOR HIV. THE FIRST POOL, TESTED ON INSTRUMENT 5149 AT 06:18, HAD A CYCLE THRESHOLD (CT) VALUE OF 37.59. THE SECOND POOL, ALSO TESTED ON INSTRUMENT 5149 AT 08:36, HAD THE SAME CT VALUE OF 37.59. BOTH POOLS WERE SUBSEQUENTLY RESOLVED ON INSTRUMENT 5100. FOR THE FIRST POOL, ALL RESOLUTION TESTING WAS NON-REACTIVE. FOR THE SECOND POOL, TWO BATCHES WERE TESTED: BATCH 12606 SAMPLES WERE ALL NON-REACTIVE, AND BATCH 12608, REPEATED ON INSTRUMENT 5149, WAS ALSO NON-REACTIVE. FINAL RESOLUTION RESULTS FOR BOTH POOLS WERE NON-REACTIVE, CONSISTENT WITH NEGATIVE SEROLOGY RESULTS FOR ALL SAMPLES. THE FINAL RESULTS FOR ALL SAMPLES WERE NON-REACTIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23086578 ↗

Other ASSAY,HYBRIDIZATION AND/OR NUCLEIC ACID AMPLIFICATION FOR DETECTION OF HEPATITIS C RNA,HEPATITIS C VIRUS devices

All ASSAY,HYBRIDIZATION AND/OR NUCLEIC ACID AMPLIFICATION FOR DETECTION OF HEPATITIS C RNA,HEPATITIS C VIRUS devices →