COMPREHENSIVE® SHOULDER SYSTEM 407393Recalls & reported problems
Biomet Orthopedics, LLC · Class 1
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00880304456808), received between Oct 10, 2019 and Mar 25, 2022 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fracture4
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 25, 2022Fracture; Fracture
IT WAS REPORTED THAT THE SURGEON WAS TRYING TO REMOVE THE COMPREHENSIVE BROACH FROM THE PATIENT'S HUMERUS USING THE EXTRACTION TOOL AND WHILE USING THE SLAP HAMMER THE EXTRACTION TOOL FRACTURED. ANOTHER INSTRUMENT WAS BROUGHT INTO THE ROOM TO USE AND THAT ONE BROKE AS WELL. NO FOREIGN BODY WAS RETAIN NO REPORT OF ANY HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13901264 ↗MalfunctionMar 25, 2022Fracture; Fracture
IT WAS REPORTED THAT THE SURGEON WAS TRYING TO REMOVE THE COMPREHENSIVE BROACH FROM THE PATIENT'S HUMERUS USING THE EXTRACTION TOOL AND WHILE USING THE SLAP HAMMER THE EXTRACTION TOOL FRACTURED. ANOTHER INSTRUMENT WAS BROUGHT INTO THE ROOM TO USE AND THAT ONE BROKE AS WELL. NO FOREIGN BODY WAS RETAIN NO REPORT OF ANY HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13901255 ↗MalfunctionNov 6, 2019Fracture; Fracture
NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORTING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9284484 ↗MalfunctionOct 10, 2019Fracture; Fracture
NO FURTHER INFORMATION AVAILABLE AT THE TIME OF THIS REPORTING.
Patient problems: No Information
MAUDE report 9175367 ↗
Other EXTRACTOR devices
- ATTUNE 2545-00-138Depuy (Ireland) · EXTRACTOR392 reports
- PINNACLE 2217-50-001Depuy (Ireland) · EXTRACTOR243 reports
- RECLAIM 2975-00-635Depuy (Ireland) · EXTRACTOR157 reports
- ATTUNE 2545-00-060Depuy (Ireland) · EXTRACTOR153 reports
- ESC PINNACLE 2218-00-010Depuy (Ireland) · EXTRACTOR151 reports
- NA 2125-00-600Depuy Orthopaedics, Inc. · EXTRACTOR130 reports
- P.F.C. 86-5226Depuy (Ireland) · EXTRACTOR80 reports
- NA 03.010.000Synthes GmbH · EXTRACTOR74 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.