Skip to content

Arcos Modular Revision Hip System 11-300817Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

All models of Arcos Modular Revision Hip System (71)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00880304474253), received between Jun 29, 2021 and Apr 3, 2026 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert2
  • Adverse Event Without Identified Device or Use Problem1
  • Material Deformation1
  • Migration1
  • Positioning Failure1
  • Separation Failure1

Patient problems

  • Failure of Implant2
  • No Clinical Signs, Symptoms or Conditions2
  • Ambulation or Postural Difficulties1
  • Bone Fracture(s)1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 3, 2026Migration

    IT WAS REPORTED THAT THE PATIENT HAD A FAILED HIP RESURFACING SURGERY WHICH WAS TREATED WITH CONVERSION TO TOTAL HIP. DUE TO BEARING COMPATIBILITY AND ACETABULAR DEFECT THE ACETABULAR COMPONENT WAS LEFT INSITU AND AN ARCOS HIP STEM WITH LDH DM HEAD WAS IMPLANTED. THE STEM SUBSIDED NECESSITATING ITS REMOVAL AND SUBSEQUENT REVISION OF THE FEMORAL SIDE. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 24778098 ↗
  • InjuryMar 31, 2025Positioning Failure; Material Deformation; Positioning Failure; Material Deformation

    IT WAS REPORTED BY THE SURGEON IMPLANTED AN ARCOS STS 17 X 150 STEM. THE SURGEON REMOVED THE STEM AS IT WAS NOT IMPLANTED DEEP ENOUGH AND WANTED TO REAM FURTHER. THE SURGEON USED A SLAP HAMMER FROM ANOTHER COMPANY AND THREADED THE SLAP HAMMER INTO THE ARCOS STS STEM, BUT THE THREAD STRIPPED OUT OF THE ARCOS IMPLANT. SURGEON THEN ATTACHED THE BLACK HANDLE IMPACTOR FROM THE ARCOS INSTRUMENTS. THE SURGEON TRIED TO KNOCK THE STEM OUT, THE IMPACTOR BROKE LEAVING THE THREAD INSIDE THE TOP OF THE STEM.

    Patient problems: Failure of Implant

    MAUDE report 21730377 ↗
  • MalfunctionJan 24, 2025Difficult to Insert; Difficult to Insert

    THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21231207 ↗
  • MalfunctionOct 28, 2024Difficult to Insert; Difficult to Insert

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20546841 ↗
  • InjuryJun 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO FURTHER INFORMATION AT THE TIME OF THIS REPORT.

    Patient problems: Bone Fracture(s)

    MAUDE report 19518615 ↗
  • InjuryJun 29, 2021Separation Failure; Separation Failure

    IT WAS REPORTED THAT DURING A HIP PROCEDURE, THE TAPER DISASSEMBLY TOOL WAS FOUND TO BE CROSS-THREADED ON THE TAPER CONE AND THE SURGEON WAS UNABLE TO REMOVE THE TAPER FROM THE STEM. THE SURGEON THEN OPTED TO PERFORM AN ETO TO REMOVE THE STEM, TAPER AND DISASSEMBLY TOOL ALL IN ONE PIECE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant

    MAUDE report 12087822 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish devices →

Sources and updates