Skip to content

G7 OSSEOTI ACETABULAR SHELL 110010249Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00880304544246), received between May 29, 2025 and May 29, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 29, 2025Positioning Failure; Device-Device Incompatibility; Positioning Failure; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE CUP WAS IMPLANTED WITH THREE SCREWS. A LINER WAS ATTEMPTED TO BE IMPLANTED AND WOULD NOT MATE WITH THE CUP. A SECOND LINER WAS OPENED AND ALSO WOULD NOT MATE WITH THE CUP. TRYING TO PLACE THE LINER RESULTED IN THE CUP LOSING FIXATION. THE SHELL AND SCREWS WERE REMOVED AND A SECOND, LARGER CUP WAS IMPLANTED. THE LINER WOULD STILL NOT MATE WITH THE CUP. A THIRD LINER WAS ABLE TO BE IMPLANTED INTO THE SECOND CUP. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22109043 ↗

Other prosthesis, hip, constrained, cemented or uncemented, metal/polymer, + additive devices

All prosthesis, hip, constrained, cemented or uncemented, metal/polymer, + additive devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.