Skip to content

COMPREHENSIVE SHOULDER SYSTEM 113134Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

All models of Comprehensive Shoulder System (68)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00880304553972), received between Sep 3, 2024 and May 15, 2026 — classified as 2 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Erosion1
  • Metal Related Pathology1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 15, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL IMPLANTATION APPROXIMATELY FOUR (4) YEARS AND SIX (6) MONTHS AGO. SUBSEQUENTLY, IT WAS NOTED POST-OPERATIVELY THAT THE TM MODULAR PEG OF THE GLENOID PENETRATED THE POLY LINER AND PROTRUDED. THE PATIENT UNDERWENT A REVISION APPROXIMATELY FOUR (4) YEARS AND FIVE (5) MONTHS POST-IMPLANTATION WHERE THE INITIAL PLAN WAS TO CONVERT FROM A TSA TO RTSA; HOWEVER, DUE TO METALLOSIS AND SEVERE BONE RESORPTION, IMPLANTING A GLENOID IMPLANT WAS UNABLE TO OCCUR. THEREFORE, BONE WAS HARVESTED FROM THE ILIUM AND USED FOR A BONE GRAFT AT THE SITE OF THE RESORPTION. IT IS NOTED THE PATIENT WAS VERY ACTIVE, IN WHICH THIS EVENT HAS CAUSED POST-OP MANAGEMENT DIFFICULT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Erosion; Metal Related Pathology

    MAUDE report 25184065 ↗
  • InjurySep 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO PAIN. DUE DILIGENCE IS COMPLETE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Pain

    MAUDE report 20125527 ↗

Other PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices

All PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices →

Sources and updates