LIFEPAK 15 monitor/defibrillator 97577 (REF 99577-001957U)Recalls & reported problems
Physio-Control, Inc. · 15AACBABBABBBAAAAAAAAABFLP15 MONITOR/DEFIB · Class 3
All models of LIFEPAK 15 monitor/defibrillator (93)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00883873793749), received between Jun 7, 2021 and Mar 15, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Unexpected Shutdown” (2 reports).
17 other recalls cover devices cleared under the same FDA submission (P160026); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unexpected Shutdown2
- Device Sensing Problem1
- Failure to Power Up1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 15, 2024Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY SHUT DOWN WHEN POWERED ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18910578 ↗MalfunctionOct 11, 2023Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT COMPLETE ITS INITIAL BOOT-UP CYCLE. THIS ISSUE HAS THE POTENTIAL TO EITHER DELAY, OR PREVENT, DEFIBRILLATION FROM BEING DELIVERED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17910136 ↗MalfunctionSep 18, 2023Device Sensing Problem; Device Sensing Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT SHOW THE THERAPY CABLE ON THE SCREEN. IN THIS STATE DEFIBRILLATION THERAPY WOULD BE DELAYED OR UNAVAILABLE IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER, THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17764023 ↗MalfunctionJun 7, 2021Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE UNEXPECTEDLY SHUTS DOWN. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11946313 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160026. One submission often covers many models, so check the product and lot details in each record.
- Z-2065-2026 · Apr 13, 2026 · Open, ClassifiedLIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001
- Z-2498-2025 · Jul 22, 2025 · Open, ClassifiedLIFEPAK 1000 Part / Catalog Numbers / UDI-DI code: 320371500023 / 99425-000025 / 885074698953 320371500229 / 99425-000025 / 885074698953 331706400001 / 99425-000025 / 885074698953
- Z-2497-2025 · Jul 22, 2025 · Open, ClassifiedLIFEPAK 15 Part / Catalog Numbers / UDI-DI code: V15-7-000022 / 99577-001957 / 883873871690 V15-2-000051 / 99577-000046 / 883873988022 V15-2-000099 / 99577-000093 / 883873979020 V
- Z-1232-2025 · Feb 3, 2025 · Open, ClassifiedLIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.
- Z-1184-2025 · Jan 21, 2025 · Open, ClassifiedLIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sen
- Z-0230-2025 · Aug 30, 2024 · Open, ClassifiedLIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o 99577-001255 o 99577-001256 o 99577-001368 o 99
- Z-0567-2024 · Nov 6, 2023 · Open, ClassifiedAC Power Adapter for use with the LIFEPAK 15 Monitor/Defibrillator, REF: 11140-000098
- Z-0300-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-001373, Part Number V15-7-000002
- Z-0297-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-000055, Part Number V15-2-000061
- Z-0304-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-001958, Part Number: V15-7-000023
- Z-0302-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-001956, Part Number V15-7-000021
- Z-0305-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-001960, Part Number: V15-7-000026
- Z-0298-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-000061, Part Number V15-2-000067
- Z-0296-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-000046, Part Number V15-2-000051
- Z-0299-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-001255, Part Number V15-2-001602
…and 2 more.
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page