LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 99402 (REF 99402-000033)Recalls & reported problems

Physio-Control, Inc. · LP20-CONFIGURED,W/PACING,SPO2,ENGLISH · Class 3

All models of LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors (23)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Physio-Control LIFEPAK 20 is a defibrillator and monitor used in emergency care. Of its 78 reports, all are malfunctions, with no injuries and no deaths. Many describe the unit failing to power on or shutting down, and some describe weak or missing energy output. Most reports say no patient was involved. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
77
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 77 adverse event reports that list this exact device (DI 00883873813799), received between Dec 26, 2018 and Apr 17, 2025 — classified as 77 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (18 reports).

17 other recalls cover devices cleared under the same FDA submission (P160026); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem18
  • Failure to Power Up16
  • Unexpected Shutdown10
  • Inappropriate/Inadequate Shock/Stimulation8
  • Connection Problem4
  • Device Sensing Problem3
  • Energy Output Problem3
  • Failure to Deliver Energy3

Patient problems

  • No Patient Involvement50
  • No Clinical Signs, Symptoms or Conditions21
  • No Known Impact Or Consequence To Patient11
  • Death1

Recent adverse event reports

The 25 most recent of 77 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE ON/OFF BUTTON OF THEIR DEVICE IS UNRESPONSIVE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21856313 ↗
  • MalfunctionApr 22, 2024Erratic or Intermittent Display; Erratic or Intermittent Display

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE DISPLAY OF THEIR DEVICE WOULD COME ON INTERMITTENTLY. THIS ISSUE HAS THE POTENTIAL TO EITHER DELAY, OR PREVENT, DEFIBRILLATION FROM BEING DELIVERED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19152418 ↗
  • MalfunctionDec 21, 2023Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS NOT WORKING CORRECTLY WHEN ATTEMPTING TO SHOCK WITH THE QUICK COMBO THERAPY CABLE. AS A RESULT, DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18378821 ↗
  • MalfunctionNov 10, 2023Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE DEFIBRILLATION ENERGY LEVEL DELIVERED FROM THEIR DEVICE WAS NOT AS EXPECTED, WHEN ATTEMPTED. AS A RESULT, WRONG DEFIBRILLATION THERAPY WOULD BE DELIVERED, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18112618 ↗
  • MalfunctionSep 15, 2023Failure to Analyze Signal; Failure to Analyze Signal

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT PROVIDE ANY READINGS THROUGH THE DEFIBRILLATION ELECTRODE. IN THIS STATE DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE TO A PATIENT IF NEEDED. THE CUSTOMER ADVISED THAT A THIRD PARTY DEFIBRILLATION ELECTRODES WERE USED DURING THE EVENT. THIS ISSUE IS PATIENT RELATED; HOWEVER, THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17758709 ↗
  • MalfunctionNov 10, 2022Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S ON BUTTON WAS NOT FUNCTIONING. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15776905 ↗
  • MalfunctionSep 19, 2022Failure to Deliver Energy; Failure to Deliver Energy

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD NO ENERGY OUTPUT. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15447716 ↗
  • MalfunctionAug 3, 2022Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED TO DELIVER ENERGY DURING DEFIBRILLATION. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15156694 ↗
  • MalfunctionMay 3, 2021Failure to Deliver Shock/Stimulation; Device Sensing Problem; Failure to Deliver Shock/Stimulation; Device Sensing Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WAS UNABLE TO DELIVER A SHOCK DURING A PATIENT EVENT. THE CUSTOMER ADVISED THAT THE DEVICE WAS ABLE TO CHARGE FOR DEFIBRILLATION BUT NOT DELIVER A SHOCK. THE CUSTOMER ALSO STATED THE DEVICE WAS GIVING A CONNECT ELECTRODES MESSAGE. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED.THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11761710 ↗
  • MalfunctionJan 26, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE'S DISPLAY WAS COMPLETELY WHITE. A WHITE DISPLAY CAN BE INDICATIVE OF A LOCK UP CONDITION. AS A RESULT, DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11234184 ↗
  • MalfunctionDec 23, 2020Failure to Sense; Failure to Sense

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE DOES NOT RECOGNIZE THE QUICK COMBO ELECTRODES. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11067557 ↗
  • MalfunctionDec 9, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    A PHYSIO-CONTROL SERVICE REPRESENTATIVE WAS PERFORMING ROUTINE MAINTENANCE ON A LOANER DEVICE AND OBSERVED NON-CRITICAL EVENT CODES LOGGED IN THE MEMORY OF THE DEVICE. UPON PERFORMING REPAIRS IT WAS OBSERVED THAT THE DEVICE WAS UNABLE TO BOOT COMPLETE A BOOT CYCLE. IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10979671 ↗
  • MalfunctionDec 1, 2020Failure to Power Up; Unexpected Shutdown; Failure to Power Up; Unexpected Shutdown

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE UNEXPECTEDLY POWERED OFF AND WOULD NOT POWER BACK ON DURING A PATIENT EVENT. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10926016 ↗
  • MalfunctionNov 19, 2020Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE UNEXPECTEDLY POWERED DOWN PRIOR TO BEING ABLE TO DELIVER ENERGY TO THE PATIENT. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10872097 ↗
  • MalfunctionOct 21, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE HAD A NON-CRITICAL ISSUE AND MAY HAVE A DAMAGED THERAPY CABLE ACCESSORY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT USE REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10714982 ↗
  • MalfunctionOct 8, 2020Electrical /Electronic Property Problem; Failure to Power Up; Electrical /Electronic Property Problem; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, PHYSIO-CONTROL OBSERVED THAT THE DEVICE'S DISPLAY WAS COMPLETELY WHITE. A WHITE DISPLAY CAN BE INDICATIVE OF A DEVICE LOCK UP CONDITION. AS A RESULT DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10650128 ↗
  • MalfunctionSep 28, 2020Energy Output Problem; Failure to Power Up; Energy Output Problem; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEY CANNOT TURN ON THEIR DEVICE. IN THIS STATE THE DEVICE WOULD NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. ADDITIONALLY, AFTER REPAIRS IT WAS FOUND THAT THE UNIT WAS NOT DELIVERING ENERGY PROPERLY, WHICH COULD ALSO PREVENT PROPER DEFIBRILLATION THERAPY. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10597683 ↗
  • MalfunctionAug 31, 2020Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE POWERED OFF TWICE DURING TESTING. IT WAS REPORTED THAT THE DEVICE POWERED OFF ONCE DURING A USER TEST, THEN DURING A TEST DEFIB SCENARIO POWERED ITSELF OFF WITHOUT CYCLING BACK ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 10475089 ↗
  • MalfunctionAug 24, 2020Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE UNEXPECTEDLY POWERED DOWN MULTIPLE TIMES. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 10446255 ↗
  • MalfunctionAug 10, 2020Power Problem; Power Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NOT COMPLETE THE BOOT UP PROCESS. IN THIS STATE THE DEVICE WOULD NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT USE REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 10388257 ↗
  • MalfunctionJul 9, 2020Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 10253740 ↗
  • MalfunctionMay 28, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION OF THE CUSTOMER¿S DEVICE, PHYSIO-CONTROL OBSERVED THAT THE DEVICE WAS UNABLE TO COMPLETE A BOOT A CYCLE. IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 10094817 ↗
  • MalfunctionMay 5, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING EVALUATION, PHYSIO-CONTROL OBSERVED THAT THE DEVICE WOULD NOT COMPLETE BOOT UP DURING INITIAL POWER ON. THE DEVICE WOULD SHOW A SPLASH SCREEN AND WOULD NOT ADVANCE TO THE BOOT CYCLE FROM THERE. IN THIS STATE THE DEVICE WOULD NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 10024958 ↗
  • MalfunctionMay 1, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WAS ABLE TO COMPLETE A BOOT UP CYCLE. IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 10018675 ↗
  • MalfunctionMar 26, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE STOPPED SHOCKING AFTER SUCCESSFULLY SHOCKING MULTIPLE TIMES. THE CUSTOMER ADVISED THAT THIS OCCURRED DURING TRAINING. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 9882499 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160026. One submission often covers many models, so check the product and lot details in each record.

  • Z-2065-2026 · Apr 13, 2026 · Open, Classified
    LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001
  • Z-2498-2025 · Jul 22, 2025 · Open, Classified
    LIFEPAK 1000 Part / Catalog Numbers / UDI-DI code: 320371500023 / 99425-000025 / 885074698953 320371500229 / 99425-000025 / 885074698953 331706400001 / 99425-000025 / 885074698953
  • Z-2497-2025 · Jul 22, 2025 · Open, Classified
    LIFEPAK 15 Part / Catalog Numbers / UDI-DI code: V15-7-000022 / 99577-001957 / 883873871690 V15-2-000051 / 99577-000046 / 883873988022 V15-2-000099 / 99577-000093 / 883873979020 V
  • Z-1232-2025 · Feb 3, 2025 · Open, Classified
    LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.
  • Z-1184-2025 · Jan 21, 2025 · Open, Classified
    LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sen
  • Z-0230-2025 · Aug 30, 2024 · Open, Classified
    LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o 99577-001255 o 99577-001256 o 99577-001368 o 99
  • Z-0567-2024 · Nov 6, 2023 · Open, Classified
    AC Power Adapter for use with the LIFEPAK 15 Monitor/Defibrillator, REF: 11140-000098
  • Z-0300-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001373, Part Number V15-7-000002
  • Z-0304-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001958, Part Number: V15-7-000023
  • Z-0299-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001255, Part Number V15-2-001602
  • Z-0302-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001956, Part Number V15-7-000021
  • Z-0296-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-000046, Part Number V15-2-000051
  • Z-0297-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-000055, Part Number V15-2-000061
  • Z-0303-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001957, Part Number V15-5-000586 and V15-7-000022
  • Z-0301-2024 · Sep 29, 2023 · Open, Classified
    LIFEPAK 15, REF: 99577-001955, Part Number V15-5-000584 and V15-7-000020

…and 2 more.