LIFEPAK 1000 defibrillator 99425 (REF 99425-000025)Recalls & reported problems
Physio-Control, Inc. · LP1000 CONFIGURED UNIT1000AAAABAAACBBAAAAAAAAA · Class 3
All models of LIFEPAK 1000 defibrillator (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This LIFEPAK 1000 defibrillator has 76 adverse event reports mentioning it. They describe units failing to power on or deliver defibrillation therapy, unresponsive keys, and other electronic problems, with no deaths and no injuries. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 76
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 76 adverse event reports that list this exact device (DI 00883873815465), received between Jan 3, 2019 and Jun 5, 2026 — classified as 76 malfunctions.
The most frequently reported specific device problem is “Failure to Power Up” (25 reports).
17 other recalls cover devices cleared under the same FDA submission (P160026); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Power Up25
- Electrical /Electronic Property Problem10
- Failure to Sense6
- Device Sensing Problem5
- Failure to Deliver Shock/Stimulation5
- Key or Button Unresponsive/not Working5
- Failure to Charge4
- Connection Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions52
- No Patient Involvement21
- No Known Impact Or Consequence To Patient2
- Cardiac Arrest1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 76 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 5, 2026Loose or Intermittent Connection
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE LOCKING OF THE BATTERY OF THEIR DEVICE WAS NOT WORKING PROPERLY AND THE BATTERY WAS FREQUENTLY DISCONNECTED FROM THE DEVICE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25408964 ↗MalfunctionJan 7, 2026Failure to Power Up
DURING PREVENTIVE MAINTENANCE, STRYKER OBSERVED THAT THE CUSTOMER'S DEVICE DOES NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED.THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23987642 ↗MalfunctionSep 3, 2025Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
A CUSTOMER RETURNED THEIR DEVICE FOR FUNCTIONAL AND PERFORMANCE TESTING. DURING THE DEVICE'S PERFORMANCE INSPECTION PROCEDURE (PIP), STRYKER OBSERVED THE DEVICE DID NOT DELIVER DEFIBRILLATION ENERGY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22952624 ↗MalfunctionFeb 26, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE'S KEYPAD WAS UNRESPONSIVE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21468175 ↗MalfunctionDec 12, 2024Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20919681 ↗MalfunctionNov 14, 2024Electrical /Electronic Property Problem; Device Sensing Problem; Electrical /Electronic Property Problem; Device Sensing Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT SWITCH TO MANUAL MODE. ADDITIONALLY, THE CUSTOMER REPORTED THAT THE DEVICE DOES NOT RECOGNIZE ECG CABLE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20680571 ↗MalfunctionNov 12, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THE DEVICE WOULD NOT GO INTO MANUAL MODE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20666374 ↗MalfunctionJul 9, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE SOFTKEYS NOT FUNCTIONAL. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19699795 ↗MalfunctionApr 30, 2024Failure to Power Up; Failure to Power Up
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. AS A RESULT, DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19212579 ↗MalfunctionApr 22, 2024Failure to Deliver Shock/Stimulation; Device Sensing Problem; Failure to Deliver Shock/Stimulation; Device Sensing Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED TO PROVIDE DEFIBRILLATION THERAPY DURING INTERVENTION ON A PATIENT. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19151540 ↗MalfunctionApr 18, 2024Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT PROVIDE SHOCK. AS A RESULT, DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WERE NO REPORTS OF ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19137823 ↗MalfunctionApr 16, 2024Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation; Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE GAVE INCORRECT SHOCK ADVISORY IN AED MODE DURING INTERVENTION ON A PATIENT. THE DEVICE INDICATED NORMAL SINUS RHYTHM BUT A SHOCK WAS ADVISED. IN THIS STATE, WRONG DEFIBRILLATION THERAPY MAY BE DELIVERED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19114029 ↗MalfunctionDec 21, 2023Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18377075 ↗MalfunctionDec 21, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WILL NOT COMPLETE BOOT-UP CYCLE DURING POWER UP. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18374648 ↗MalfunctionDec 15, 2023Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE LEFT SOFTKEY WAS NOT FUNCTIONAL. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR NOT AVAILABLE IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18335218 ↗MalfunctionOct 12, 2023Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17926313 ↗MalfunctionSep 5, 2023Failure to Sense; Failure to Sense
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED TO RECOGNIZE ELECTRODES. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17683674 ↗MalfunctionJun 22, 2023Incorrect Interpretation of Signal; Incorrect Interpretation of Signal
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE INAPPROPRIATELY PROVIDED A SHOCK ADVISORY. IN THIS STATE, WRONG DEFIBRILLATION THERAPY MAY BE DELIVERED. THIS ISSUE IS PATIENT RELATED; HOWEVER, NO ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17180303 ↗MalfunctionMay 9, 2023Failure to Power Up; Electrical Power Problem; Failure to Power Up; Electrical Power Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, STRYKER OBSERVED THAT THE DEVICE DOESN'T COMPLETE BOOT UP CYCLE AND DOES NOT POWER ON. AS A RESULT, DEFIBRILLATION THERAPY MAY BE DELAYED OR UNAVAILABLE, IF NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16898813 ↗MalfunctionMay 5, 2023Device Sensing Problem; Device Sensing Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE WOULD NOT RECOGNIZE THE DEFIBRILLATION ELECTRODES. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16874434 ↗MalfunctionMay 4, 2023Charging Problem; Connection Problem; Charging Problem; Connection Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE IS NOT RECOGNIZING THE CONNECTED ELECTRODES AND IS TAKING TOO LONG TO CHARGE DEFIBRILLATION ENERGY. AS A RESULT, DEFIBRILLATION THERAPY MAY BE DELAYED OR UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16868481 ↗MalfunctionFeb 28, 2023Failure to Sense; Failure to Sense
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS NOT DETECTING THEIR DEFIBRILLATION ELECTRODES DURING A PATIENT EVENT. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16456844 ↗MalfunctionFeb 27, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, STRYKER OBSERVED THAT THE DEVICE'S DISPLAY IS BLANK. AS A RESULT, THE LABELING FOR BOTH THE ANALYZE AND CHARGE SOFT KEY BUTTONS AT THE BOTTOM OF THE DISPLAY WOULD NOT BE VISIBLE TO THE DEVICE USER. THIS ISSUE HAS THE POTENTIAL TO DELAY OR PREVENT DEFIBRILLATION FROM BEING DELIVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16443752 ↗MalfunctionFeb 21, 2023Connection Problem; Connection Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DOES NOT RECOGNIZE WHEN THE DEFIBRILLATION ELECTRODES ARE CONNECTED TO IT. AS A RESULT, DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16414142 ↗MalfunctionFeb 8, 2023Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16326320 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160026. One submission often covers many models, so check the product and lot details in each record.
- Z-2065-2026 · Apr 13, 2026 · Open, ClassifiedLIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001
- Z-2497-2025 · Jul 22, 2025 · Open, ClassifiedLIFEPAK 15 Part / Catalog Numbers / UDI-DI code: V15-7-000022 / 99577-001957 / 883873871690 V15-2-000051 / 99577-000046 / 883873988022 V15-2-000099 / 99577-000093 / 883873979020 V
- Z-2498-2025 · Jul 22, 2025 · Open, ClassifiedLIFEPAK 1000 Part / Catalog Numbers / UDI-DI code: 320371500023 / 99425-000025 / 885074698953 320371500229 / 99425-000025 / 885074698953 331706400001 / 99425-000025 / 885074698953
- Z-1232-2025 · Feb 3, 2025 · Open, ClassifiedLIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.
- Z-1184-2025 · Jan 21, 2025 · Open, ClassifiedLIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sen
- Z-0230-2025 · Aug 30, 2024 · Open, ClassifiedLIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-000076 o 99577-001250 o 99577-001255 o 99577-001256 o 99577-001368 o 99
- Z-0567-2024 · Nov 6, 2023 · Open, ClassifiedAC Power Adapter for use with the LIFEPAK 15 Monitor/Defibrillator, REF: 11140-000098
- Z-0300-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-001373, Part Number V15-7-000002
- Z-0299-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-001255, Part Number V15-2-001602
- Z-0297-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-000055, Part Number V15-2-000061
- Z-0298-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-000061, Part Number V15-2-000067
- Z-0301-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-001955, Part Number V15-5-000584 and V15-7-000020
- Z-0296-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-000046, Part Number V15-2-000051
- Z-0302-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-001956, Part Number V15-7-000021
- Z-0304-2024 · Sep 29, 2023 · Open, ClassifiedLIFEPAK 15, REF: 99577-001958, Part Number: V15-7-000023
…and 2 more.
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page