Skip to content

LIFEPAK CR Plus and LIFEPAK Express defibrillators 99403 (REF 99403-000007)Recalls & reported problems

Physio-Control, Inc. · ASSY-LP CR PLUS.AUTO,CIRCULATION · Class 3

All models of LIFEPAK CR Plus and LIFEPAK Express defibrillators (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00883873820285), received between Jan 22, 2019 and Jan 22, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure to Power Up” (1 report).

1 other recall cover devices cleared under the same FDA submission (P160012); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure to Power Up1

Patient problems

  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 22, 2019Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, CHARGE-PAK AND SERVICE WRENCH). THE CUSTOMER ADVISED PHYSIO THE DEVICE WOULD NOT POWER ON WHEN PROMPTED. WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF IT WAS NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 8269141 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160012. One submission often covers many models, so check the product and lot details in each record.

  • Z-1544-2026 · Jan 21, 2026 · Open, Classified
    PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →

Sources and updates