LIFEPAK CR Plus and LIFEPAK Express defibrillators 97403 (REF 97403-000018)Recalls & reported problems
Physio-Control, Inc. · ASSY-LPCR PLUS,AUTO,ENGLISH,AHA2005 · Class 3
All models of LIFEPAK CR Plus and LIFEPAK Express defibrillators (25)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00883873820674), received between Dec 5, 2018 and Dec 21, 2023 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Power Problem” (6 reports).
1 other recall cover devices cleared under the same FDA submission (P160012); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Power Problem6
- Failure to Power Up1
- No Audible Prompt/Feedback1
Patient problems
- No Patient Involvement5
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 21, 2023Power Problem; Power Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, CHARGE-PAK AND SERVICE WRENCH). WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18376409 ↗MalfunctionNov 9, 2023Power Problem; Power Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, CHARGE-PAK AND SERVICE WRENCH). WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF IT WAS NEEDED. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18099806 ↗MalfunctionOct 27, 2022Power Problem; Power Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, CHARGE-PAK AND SERVICE WRENCH). WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF IT WAS NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15686498 ↗MalfunctionMay 23, 2019Power Problem; Power Problem
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, CHARGE-PAK AND SERVICE WRENCH). WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF IT WAS NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 8638746 ↗MalfunctionFeb 26, 2019Power Problem; Power Problem
A CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, CHARGE-PAK AND SERVICE WRENCH). WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF IT WAS NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 8372475 ↗MalfunctionFeb 18, 2019No Audible Prompt/Feedback; No Audible Prompt/Feedback
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE DID NOT PROVIDE ANY VOICE PROMPTS WHEN THE DEVICE WAS POWERED ON. WITHOUT VOICE PROMPTS, A DEVICE USER WOULD NOT RECEIVE THE AUDIBLE INSTRUCTIONS ON HOW TO PROPERLY USE THE DEVICE ON A PATIENT WHICH COULD DELAY, OR PREVENT, DEFIBRILLATION IF NECESSARY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 8348170 ↗MalfunctionJan 24, 2019Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE DISPLAYED ALL THREE ICONS (CHARGE-PAK, ATTENTION, AND SERVICE WRENCH) AND WOULD NO LONGER POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 8273733 ↗MalfunctionDec 5, 2018Power Problem; Power Problem
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE HAD ALL THREE ICONS (CHARGE PAK, ATTENTION AND SERVICE WRENCH) PRESENT ON ITS READINESS DISPLAY. THE THREE ICONS BEING ILLUMINATED INDICATES THAT THE DEVICE MAY NOT BE ABLE TO PROVIDE SUFFICIENT DEFIBRILLATION THERAPY, IF NECESSARY. THE CUSTOMER ADVISED THAT THEY HAD OBTAINED THE DEVICE FOR POSSIBLE USE WITH A PATIENT WHEN THE REPORTED ISSUE WAS OBSERVED; HOWEVER, THE DEVICE WAS NOT NEEDED AND WAS NEVER CONNECTED TO THE PATIENT FOR USE.
Patient problems: No Patient Involvement
MAUDE report 8133199 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P160012. One submission often covers many models, so check the product and lot details in each record.
- Z-1544-2026 · Jan 21, 2026 · Open, ClassifiedPHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page