LIFEPAK CR Plus and LIFEPAK Express defibrillators 99427 (REF 99427-000129)Recalls & reported problems

Physio-Control, Inc. · ASSY,LP EXPRESS, SEMI, ENG,AHA2005, FDA · Class 3

All models of LIFEPAK CR Plus and LIFEPAK Express defibrillators (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This LIFEPAK CR Plus and LIFEPAK Express automated external defibrillator is a portable emergency device. The 26 reports mostly describe power problems, such as failing to turn on, shutting down unexpectedly, or showing warning lights, meaning the shock might not be available if needed, with no patient use mentioned. There were no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
26
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 00883873823644), received between Mar 29, 2019 and Mar 10, 2025 — classified as 26 malfunctions.

The most frequently reported specific device problem is “Failure to Power Up” (9 reports).

1 other recall cover devices cleared under the same FDA submission (P160012); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Power Up9
  • Power Problem8
  • Unexpected Shutdown5
  • Audible Prompt/Feedback Problem2
  • Connection Problem1
  • Electrical /Electronic Property Problem1
  • Electrical Power Problem1
  • Failure to Analyze Signal1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • No Patient Involvement12

Recent adverse event reports

The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 10, 2025Power Problem; Power Problem

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, CHARGE-PAK AND SERVICE WRENCH). WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF IT WAS NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21561892 ↗
  • MalfunctionDec 8, 2023Power Problem; Power Problem

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, CHARGE-PAK AND SERVICE WRENCH). WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF IT WAS NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18289473 ↗
  • MalfunctionNov 8, 2023Electrical Power Problem; Electrical Power Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION OF THE CUSTOMER¿S DEVICE, STRYKER OBSERVED THAT THE DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18089944 ↗
  • MalfunctionOct 20, 2023Power Problem; Power Problem

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, CHARGE-PAK AND SERVICE WRENCH). WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF IT WAS NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17976246 ↗
  • MalfunctionMay 9, 2023Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16898815 ↗
  • MalfunctionJul 5, 2022Failure to Power Up; Unexpected Shutdown; Audible Prompt/Feedback Problem; Failure to Power Up; Unexpected Shutdown; Audible Prompt/Feedback Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE POWERS OFF, HAS NO AUDIBLE PROMPTS AND WILL SUBSEQUENTLY NOT POWER ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14915684 ↗
  • MalfunctionJun 23, 2022Power Problem; Power Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DISPLAYED ALL THREE ICONS (CHARGE-PAK, ATTENTION, AND SERVICE WRENCH). THE ILLUMINATION OF ALL THREE ICONS INDICATES THAT THE DEVICE MAY NOT BE ABLE TO PROVIDE SUFFICIENT DEFIBRILLATION THERAPY, IF NECESSARY. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14789313 ↗
  • MalfunctionOct 20, 2021Unexpected Shutdown; Unexpected Shutdown

    DURING ROUTINE PREVENTATIVE MAINTENANCE TESTING BY PHYSIO-CONTROL, THE DEVICE'S BATTERY WAS FOUND NOT CHARGED. IN THIS STATE THE DEVICE WOULD NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT USE REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12669556 ↗
  • MalfunctionMay 12, 2021Failure to Power Up; Failure to Power Up

    PHYSIO-CONTROL RECEIVED A REPORT FROM THE CUSTOMER THAT THE DEVICE CANNOT BE TURNED ON; THEY REPORTED, "WHEN ON BUTTON IS PRESSED, IT SAYS "CALL FOR HELP NOW", THEN SHUTS OFF." IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED.THERE WAS NO REPORT OF PATIENT INVOLVEMENT ASSOCIATED TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11815449 ↗
  • MalfunctionFeb 10, 2021Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11311546 ↗
  • MalfunctionFeb 10, 2021Visual Prompts will not Clear; Visual Prompts will not Clear

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, PHYSIO-CONTROL OBSERVED AN EVENT CODE LOGGED IN THE DEVICE'S MEMORY. THE EVENT CODE LOGGED IS RELATED TO A POTENTIAL FAILURE OF THE DEVICE TO FAIL TO RECOGNIZE A SHOCKABLE RHYTHM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11311453 ↗
  • MalfunctionJan 6, 2021Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, PHYSIO-CONTROL OBSERVED THAT THE DEVICE WOULD FAILED TO POWER UP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11127799 ↗
  • MalfunctionDec 16, 2020Power Problem; Power Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, CHARGE-PAK AND SERVICE WRENCH). WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF IT WAS NEEDED. THERE WAS NO PATIENT USE REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11017180 ↗
  • MalfunctionSep 18, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR AED WOULD VERBALIZE WHEN THEY OPEN THE FRONT LID AND IT WOULD NOT STOP VERBALIZING WHEN THEY CLOSE THE LID. IN THIS STATE, THE DEVICE HLC INTERNAL BATTERY COULD EXPERIENCE PREMATURE DEPLETION AND IF NEEDED, DEFIBRILLATION (AUTOMATED, MANUAL OR SYNCHRONIZED CARDIOVERSION) THERAPY MAY NOT BE ABLE TO BE DELIVERED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10557086 ↗
  • MalfunctionAug 12, 2020Failure to Analyze Signal; Failure to Analyze Signal

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE PHYSIO-CONTROL OBSERVED AN EVENT CODE LOGGED IN THE DEVICE'S MEMORY. THE EVENT CODE LOGGED IS RELATED TO A POTENTIAL FAILURE OF THE DEVICE TO RECOGNIZE A SHOCKABLE RHYTHM. AS A RESULT, DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF NEEDED.

    Patient problems: No Patient Involvement

    MAUDE report 10396139 ↗
  • MalfunctionJun 1, 2020Connection Problem; Connection Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE DOES NOT RECOGNIZE THE ATTACHED QUICKCOMBO PADS/CABLE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT USE REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 10105554 ↗
  • MalfunctionMay 21, 2020Power Problem; Audible Prompt/Feedback Problem; Power Problem; Audible Prompt/Feedback Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NOT PROVIDE ANY VOICE PROMPTS. THE CUSTOMER ALSO ADVISED THAT THEIR DEVICE ALL THREE ICONS (CHARGE-PAK, ATTENTION, AND SERVICE WRENCH). IN THIS STATE, THE USER WOULD NOT BE ABLE TO USE THE DEVICE PROPERLY ON A PATIENT IF NEEDED AND THE DEVICE MAY NOT BE ABLE TO PROVIDE SUFFICIENT DEFIBRILLATION THERAPY, IF NECESSARY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 10080358 ↗
  • MalfunctionFeb 28, 2020Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE ON BUTTON NO LONGER WORKED. IN THIS STATE THE DEVICE WOULD NOT TURN AND BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 9771949 ↗
  • MalfunctionOct 1, 2019Power Problem; Power Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS ATTENTION, CHARGE-PAK AND SERVICE WRENCH. WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF IT WAS NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 9142550 ↗
  • MalfunctionJul 22, 2019Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORTED OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, PHYSIO-CONTROL OBSERVED THAT THE DEVICE UNEXPECTEDLY POWERED OFF DURING TESTING.

    Patient problems: No Patient Involvement

    MAUDE report 8814079 ↗
  • MalfunctionMay 15, 2019Power Problem; Power Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WAS SHOWING ALL THREE ICONS (ATTENTION, CHARGE-PAK AND SERVICE WRENCH). WITH ALL THREE ICONS ILLUMINATED, THE DEVICE MAY NOT HAVE SUFFICIENT POWER AVAILABLE TO DELIVER EFFECTIVE DEFIBRILLATION THERAPY TO A PATIENT, IF IT WAS NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 8612075 ↗
  • MalfunctionMay 9, 2019Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, PHYSIO-CONTROL OBSERVED THAT THE DEVICE UNEXPECTEDLY POWERED OFF WHILE IT WAS BEING TESTED.

    Patient problems: No Patient Involvement

    MAUDE report 8597805 ↗
  • MalfunctionApr 9, 2019Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE IS NOT POWERING ON. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 8497846 ↗
  • MalfunctionApr 8, 2019Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT WHEN THE ON/OFF BUTTON ON THEIR DEVICE IS PRESSED THE DEVICE DOES NOT APPEAR TO TURN ON. THERE IS NO RESPONSE FROM THE DEVICE AND NO AUDIO PROMPTS CAN BE HEARD, DESPITE ITS READINESS INDICATOR SHOWING THE OK ICON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 8491722 ↗
  • MalfunctionApr 4, 2019Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NO LONGER POWER ON. WHEN THE POWER BUTTON WAS PRESSED, THE DEVICE LID WOULD OPEN BUT NO VOICE PROMPTS WERE HEARD. THE DEVICE READINESS INDICATOR DISPLAYED ¿OK¿. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 8485552 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160012. One submission often covers many models, so check the product and lot details in each record.

  • Z-1544-2026 · Jan 21, 2026 · Open, Classified
    PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code: 00721902629013 Catalog Number: 11101-000017 UDI-DI code: 00721902682