LIFEPAK CR2 99512 (REF 99512-001268)Recalls & reported problems

Physio-Control, Inc. · CR2BBAAADDHBBBCDBCDBAAAALPCR2 CONFIGURED UNIT · Class 3

All models of LIFEPAK CR2 (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a Physio-Control LIFEPAK CR2 automated external defibrillator. Many of the 83 reports describe a missing lid magnet and the unit failing to power on when the lid is opened, and some describe the unit not delivering defibrillation energy. Most reports note no patient involvement. There are no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
83
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 83 adverse event reports that list this exact device (DI 00883873824061), received between Jan 9, 2020 and Mar 18, 2026 — classified as 83 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (48 reports).

2 other recalls cover devices cleared under the same FDA submission (P170018); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 3 reportsJun 2024: 2 reportsJul 2024: 3 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component48
  • Failure to Power Up22
  • Electrical /Electronic Property Problem4
  • Failure to Deliver Energy3
  • Failure to Sense3
  • Failure to Deliver Shock/Stimulation2
  • No Audible Prompt/Feedback2
  • Audible Prompt/Feedback Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions82
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 83 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2026Failure to Deliver Energy

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING EVALUATION IT WAS FOUND THAT THE DEVICE WOULD NOT DELIVER DEFIBRILLATION ENERGY.THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24626897 ↗
  • MalfunctionOct 4, 2024Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    A CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING INSPECTION, STRYKER OBSERVED THAT THE DEVICE WOULD NOT PROVIDE DEFIBRILLATION ENERGY. AS A RESULT, DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20377946 ↗
  • MalfunctionSep 13, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS MISSING ITS LID MAGNET. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USER HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20210857 ↗
  • MalfunctionAug 20, 2024Failure to Power Up; Failure to Power Up

    A CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON INSPECTION, IT WAS REPORTED THAT THE DEVICE WOULD NOT POWER ON WHEN THE LID IS OPENED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USER HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20004793 ↗
  • MalfunctionJul 26, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION OF THE CUSTOMER¿S DEVICE, STRYKER OBSERVED THAT THE DEVICE WOULD NOT POWER ON WHEN THE DEVICE LID WAS ACTIVATED. IN THIS STATE DEFIBRILLATION THERAPY MAY BE DELAYED IF NEEDED OR MAY LEAD TO USE ERROR RESULTING IN A FAILURE TO DELIVER DEFIBRILLATION. THERE WAS NO PATIENT USE REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19838118 ↗
  • MalfunctionJul 25, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING EVALUATION IT WAS FOUND THAT THE DEVICE WASN'T POWERING ON WHEN LID IS OPENED. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19824777 ↗
  • MalfunctionJul 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE¿S LID MAGNET WAS MISSING FROM THE LID. AS A RESULT, THE DEVICE MAY NOT AUTOMATICALLY POWER ON WHEN THE LID IS OPENED. THIS MAY LEAD TO DEFIBRILLATION THERAPY BEING DELAYED OR UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19721843 ↗
  • MalfunctionJun 7, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON BY OPENING THE LID. THIS ISSUE HAS THE POTENTIAL TO EITHER DELAY, OR PREVENT, DEFIBRILLATION FROM BEING DELIVERED.  THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19484362 ↗
  • MalfunctionJun 5, 2024Failure to Deliver Energy; Failure to Deliver Energy

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING EVALUATION IT WAS FOUND THAT THE DEVICE DID NOT DELIVER ENERGY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19468156 ↗
  • MalfunctionMay 14, 2024Audible Prompt/Feedback Problem; Audible Prompt/Feedback Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DOES NOT PROVIDE AUDIO PROMPTS AS EXPECTED UPON OPENING THE LID. THIS COULD LEAD TO A USE ERROR RESULTING IN A FAILURE TO DELIVER DEFIBRILLATION. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19315790 ↗
  • MalfunctionMay 14, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING EVALUATION STRYKER OBSERVED THAT THE DEVICE INTERMITTENTLY DOESN'T TURN ON WHEN LID OPENED. THIS ISSUE HAS THE POTENTIAL TO EITHER DELAY, OR PREVENT, DEFIBRILLATION FROM BEING DELIVERED.  THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19310844 ↗
  • MalfunctionMay 2, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USER HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON.  THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19230581 ↗
  • MalfunctionApr 17, 2024Power Problem; Power Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE WOULD NOT RECOGNIZE THE BATTERY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19127788 ↗
  • MalfunctionMar 18, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, STRYKER OBSERVED THAT THE DEVICE DOESN'T POWER ON WHEN THE LID IS OPENED. AS A RESULT, DEFIBRILLATION THERAPY MAY BE DELAYED OR UNAVAILABLE, IF NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18926725 ↗
  • MalfunctionMar 5, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS MISSING ITS LID MAGNET. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USER HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON.  THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18834918 ↗
  • MalfunctionFeb 20, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18740869 ↗
  • MalfunctionJan 12, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USERS HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WERE NO REPORTS OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18501300 ↗
  • MalfunctionDec 21, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USER HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WAS NO PATIENT INVOLVEMENT REPORTED DURING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18375069 ↗
  • MalfunctionNov 20, 2023Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING EVALUATION STRYKER OBSERVED THAT THE DEVICE WILL NOT TURN ON WHEN THE LID IS OPENED. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED.   THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18172552 ↗
  • MalfunctionNov 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THAT THEIR DEVICE WAS MISSING ITS LID MAGNET. IN THIS STATE, THE DEVICE WOULD NOT DETECT THE LID BEING OPENED OR CLOSED, AND THEREFORE WOULD NOT AUTOMATICALLY POWER ON OR OFF WHICH CAN LEAD TO A USE ERROR RESULTING IN A FAILURE TO DELIVER DEFIBRILLATION. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18172421 ↗
  • MalfunctionNov 15, 2023Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION OF THE CUSTOMER¿S DEVICE, STRYKER OBSERVED THAT THE DEVICE WOULD NOT POWER ON WHEN THE DEVICE LID WAS ACTIVATED. IN THIS STATE DEFIBRILLATION THERAPY MAY BE DELAYED IF NEEDED OR MAY LEAD TO USE ERROR RESULTING IN A FAILURE TO DELIVER DEFIBRILLATION. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18139435 ↗
  • MalfunctionNov 9, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAGNET WAS MISSING FROM THE LID OF THEIR DEVICE. THIS WILL RESULT IN AN INABILITY OF THE DEVICE TO DETECT THE LID BEING OPENED OR CLOSED. IN THIS STATE, A DELAY IN DEFIBRILLATION MAY OCCUR, AS THE USER HAS TO PUSH THE POWER BUTTON UNDERNEATH THE LID TO TURN THE DEVICE ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18107264 ↗
  • MalfunctionOct 11, 2023Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17912472 ↗
  • MalfunctionAug 24, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE¿S LID MAGNET WAS MISSING FROM THE LID. AS A RESULT, THE DEVICE MAY NOT AUTOMATICALLY POWER ON WHEN THE LID IS OPENED. THIS MAY LEAD TO DEFIBRILLATION THERAPY BEING DELAYED OR UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17612413 ↗
  • MalfunctionJul 19, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS MISSING THE MAGNET FROM THE LID OF THE DEVICE. IN THIS STATE THE DEVICE WILL NOT TURN ON AUTOMATICALLY WHEN THE LID IS ACTIVATED, WHICH CAN CAUSE A DELAY IN DEFIBRILLATION THERAPY IF NEEDED OR MAY LEAD TO USE ERROR RESULTING IN A FAILURE TO DELIVER DEFIBRILLATION. THERE WAS NO PATIENT USE REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17350675 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P170018. One submission often covers many models, so check the product and lot details in each record.

  • Z-2499-2025 · Jul 22, 2025 · Open, Classified
    LIFEPAK CR2 Part / Catalog Numbers / UDI-DI code: CR2-2-001451 99512-001434 883873785911 LIFEPAK CR2: The LIFEPAK CR2 defibrillator is indicated for use on patients 1 year of age
  • Z-1189-2021 · Jan 14, 2021 · Open, Classified
    PHYSIO CONTROL LIFEPAK CR2 Defibrillator 3321605-xxx The LIFEPAK CR2 defibrillator is designed for indoor and outdoor use in a stationary environment. After the electrode pads a