LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001964)Recalls & reported problems

Physio-Control, Inc. · 15AABBABBABBBABAAAAAAAAFLP15 MONITOR/DEFIB · Class 3

All models of LIFEPAK 15 monitor/defibrillator (93)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker LIFEPAK 15 is a portable monitor/defibrillator used to check heart rhythms and deliver shocks when needed. The 43 reports describe units failing to finish starting up, powering off on their own, or not delivering a shock, along with unresponsive buttons and a battery that would not seat properly. All were malfunctions with no patient involved, so no deaths and no injuries are reported. There are no recalls naming this device, and no reports were filed in the past12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
43
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 00883873871621), received between Apr 5, 2018 and May 29, 2024 — classified as 43 malfunctions.

The most frequently reported specific device problem is “Defibrillation/Stimulation Problem” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 3 reportsJan 2024: 3 reportsFeb 2024: 8 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defibrillation/Stimulation Problem14
  • Electrical /Electronic Property Problem8
  • Unexpected Shutdown8
  • Failure to Power Up6
  • Failure to Deliver Shock/Stimulation3
  • Key or Button Unresponsive/not Working3
  • Failure to Deliver Energy2
  • Charging Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions39
  • No Known Impact Or Consequence To Patient3
  • No Patient Involvement2

Recent adverse event reports

The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 29, 2024Electrical /Electronic Property Problem; Unexpected Shutdown; Electrical /Electronic Property Problem; Unexpected Shutdown

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT COMPLETE ITS BOOT UP CYCLE DURING INITIAL POWER ON AND HAD POWERED OFF UNEXPECTEDLY. AS A RESULT, DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19421352 ↗
  • MalfunctionMay 19, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT COMPLETE ITS BOOT UP CYCLE DURING INITIAL POWER ON. AS A RESULT, DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19348570 ↗
  • MalfunctionApr 29, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAIN KEYPAD WAS UNRESPONSIVE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY IF IT WERE NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19199981 ↗
  • MalfunctionApr 25, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT A BATTERY PIN OF THEIR DEVICE HAD BEEN PUSHED IN, WITH THE RESULT THAT THE BATTERY COULD NOT BE INSERTED INTO THE WELL. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19179954 ↗
  • MalfunctionApr 3, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER REPORTED TO THE STRYKER THAT POWER BUTTON ON THEIR DEVICE IS DIFFICULT TO PUSH IN ORDER TO POWER ON THE DEVICE. THIS ISSUE HAS THE POTENTIAL TO EITHER DELAY, OR PREVENT, DEFIBRILLATION FROM BEING DELIVERED.  THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19030530 ↗
  • MalfunctionFeb 20, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE DELIVER ENERGY. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18736891 ↗
  • MalfunctionFeb 7, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18664361 ↗
  • MalfunctionFeb 6, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18652749 ↗
  • MalfunctionFeb 6, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE DELIVER ENERGY. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18648756 ↗
  • MalfunctionFeb 5, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY, WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18646961 ↗
  • MalfunctionFeb 5, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY, WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18641165 ↗
  • MalfunctionFeb 2, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18636437 ↗
  • MalfunctionFeb 1, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER REPORTED THAT AN EVENT CODE IS LOGGED IN THE MEMORY OF THE DEVICE. THE EVENT CODE LOGGED IN ITS MEMORY IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. AS A RESULT, DEFIBRILLATION WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18623712 ↗
  • MalfunctionJan 31, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE LOGGED A CRITICAL EVENT CODE. THE EVENT CODE IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18620022 ↗
  • MalfunctionJan 31, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE DELIVER ENERGY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18612841 ↗
  • MalfunctionJan 30, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE LOGGED A CRITICAL EVENT CODE. THE EVENT CODE IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18610603 ↗
  • MalfunctionDec 21, 2023Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE DID NOT DELIVER A SHOCK AS EXPECTED WHILE IN USE WITH A PATIENT. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18378222 ↗
  • MalfunctionDec 15, 2023Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE POWERED OFF BY ITSELF MULTIPLE TIMES DURING A PATIENT EVENT. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER, THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18335189 ↗
  • MalfunctionDec 8, 2023Unexpected Shutdown; Unexpected Shutdown

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT HIS DEVICE UNEXPECTEDLY LOSES POWER WHEN TILTED BACK. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF NECESSARY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18289948 ↗
  • MalfunctionOct 13, 2023Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED TO PROVIDE DEFIBRILLATION THERAPY. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17929369 ↗
  • MalfunctionOct 11, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAIN KEYPAD ON THEIR DEVICE WAS UNRESPONSIVE. AS A RESULT, DEFIBRILLATION WOULD NOT BE AVAILABLE IF NECESSARY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17917203 ↗
  • MalfunctionOct 3, 2023Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17863069 ↗
  • MalfunctionOct 3, 2023Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE EXPERIENCED AN UNEXPECTED LOSS OF POWER. THIS ISSUE HAS THE POTENTIAL TO EITHER DELAY, OR PREVENT, DEFIBRILLATION FROM BEING DELIVERED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17863064 ↗
  • MalfunctionOct 3, 2023Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR SMALL AND MAIN KEYPAD WORKED INTERMITTENTLY. THIS ISSUE HAS THE POTENTIAL TO EITHER DELAY, OR PREVENT, DEFIBRILLATION FROM BEING DELIVERED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17860903 ↗
  • MalfunctionAug 25, 2023Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WILL NOT COMPLETE THE BOOT-UP CYCLE DURING POWER-UP. IN THIS STATE, THE DEVICE MAY BE UNABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED DURING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17624399 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →