LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001950)Recalls & reported problems
Physio-Control, Inc. · 15AABAABBABBBAAAAAAAAAAFLP15 MONITOR/DEFIB · Class 3
All models of LIFEPAK 15 monitor/defibrillator (93)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker LIFEPAK 15 is a portable monitor/defibrillator used in emergency care. The 43 reports mostly describe units failing to power up or failing to deliver a shock through the paddles when needed, along with unresponsive buttons and one report of an electric shock to a patient. There are 43 reports, no deaths and 1 injury. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 43
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 00883873871768), received between Jan 8, 2019 and Oct 4, 2024 — classified as 1 injury, 42 malfunctions.
The most frequently reported specific device problem is “Defibrillation/Stimulation Problem” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Defibrillation/Stimulation Problem9
- Electrical /Electronic Property Problem9
- Failure to Power Up9
- Key or Button Unresponsive/not Working5
- Failure to Deliver Shock/Stimulation3
- Failure to Sense2
- Inappropriate/Inadequate Shock/Stimulation2
- Unexpected Shutdown2
Patient problems
- No Clinical Signs, Symptoms or Conditions38
- No Known Impact Or Consequence To Patient3
- No Patient Involvement3
- Electric Shock1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 4, 2024Failure to Deliver Shock/Stimulation; Failure to Sense; Failure to Deliver Shock/Stimulation; Failure to Sense
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED TO PROVIDE DEFIBRILLATION THERAPY. DURING EVALUATION OF THE DEVICE ELECTRONIC RECORDS IT WAS OBSERVED THAT 'PADDLE LEAD OFF' WAS DETECTED DURING INTERVENTION. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20368981 ↗MalfunctionSep 16, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD AN EVENT CODE LOGGED IN ITS MEMORY. THE EVENT CODE LOGGED IN THE DEVICE'S MEMORY IS INDICATIVE OF A FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. AS A RESULT, DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20225869 ↗MalfunctionAug 27, 2024Failure to Deliver Energy; Failure to Deliver Energy
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WILL NOT DELIVER DEFIBRILLATION ENERGY THROUGH THE PADDLES. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20074139 ↗MalfunctionJul 9, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED TO THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19699250 ↗MalfunctionJul 1, 2024Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WILL NOT DELIVER DEFIBRILLATION ENERGY THROUGH THE PADDLES. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19645672 ↗MalfunctionJun 12, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE LOGGED A CRITICAL EVENT CODE. THE EVENT CODE IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19517010 ↗MalfunctionApr 30, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAIN KEYPAD WAS NOT FUNCTIONING. IN THIS STATE DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE TO A PATIENT IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19211262 ↗MalfunctionApr 11, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD AN EVENT CODE LOGGED IN ITS MEMORY. THE EVENT CODE LOGGED IN THE DEVICE'S MEMORY IS INDICATIVE OF A FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. AS A RESULT, DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19087237 ↗MalfunctionMar 23, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE GAVE BLANK SCREEN AND THE LEDS ILLUMINATED. THE DEVICE MAY HAVE BEEN IN A LOCK UP CONDITION AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18965550 ↗MalfunctionFeb 12, 2024Failure to Analyze Signal; Failure to Analyze Signal
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DID NOT ANALYZE A SHOCKABLE RHYTHM. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18692134 ↗MalfunctionJan 30, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE LOGGED A CRITICAL EVENT CODE. THE EVENT CODE IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18610689 ↗MalfunctionJan 22, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT CONTROLS ON THEIR DEVICE ARE UNRESPONSIVE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550960 ↗MalfunctionJan 22, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE KEYPAD IS INOPERATABLE. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18550105 ↗MalfunctionNov 14, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT WHEN THEY DELIVERED A SHOCK FOR SYNCHRONIZED CARDIOVERSION TO A PATIENT, THEIR DEVICE FROZE. THE USERS HAD TO POWER CYCLE THE DEVICE TO CONTINUE MONITORING THE PATIENT. IN THIS STATE, THE DEVICE COULD DELAY OR PROHIBIT THE DELIVERING OF DEFIBRILLATION THERAPY TO A PATIENT IF NEEDED. THE CUSTOMER ADVISED THAT THE WAS NO HARM TO THE PATIENT AS CARDIOVERSION WAS ACHIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18129937 ↗MalfunctionOct 11, 2023Unexpected Shutdown; Unexpected Shutdown
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD POWERERED OFF UNEXPECTEDLY. AS A RESULT, DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WERE NO REPORTS OF ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17917276 ↗MalfunctionSep 27, 2023Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING EVALUATION STRYKER OBSERVED THAT THEIR DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE DELIVER ENERGY. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17825193 ↗MalfunctionSep 18, 2023Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED THE USER TEST AND LOGGED AN EVENT CODE IN THE DEVICE¿S MEMORY THAT IS INDICATIVE OF A DEVICE FAILURE THAT COULD RESULT IN A PARTIAL LOSS OF DEFIBRILLATOR OUTPUT ENERGY DUE TO A LOSS OF THE NEGATIVE PORTION OF THE BIPHASIC OUTPUT WAVEFORM, RESULTING IN A MONOPHASIC SHOCK. ADDITIONALLY, THE DEFIBRILLATOR OUTPUT ENERGY MAY BE REDUCED BY UP TO APPROXIMATELY 20% FROM THE SELECTED ENERGY LEVEL. AS A RESULT, THE WRONG DEFIBRILLATION THERAPY WOULD BE DELIVERED, IF IT WERE NECESSARY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17764026 ↗MalfunctionSep 15, 2023Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE SHOCK BUTTON OF THEIR DEVICE WAS NOT RESPONDING PROPERLY AND WAS BEING DIFFICULT TO PRESS, WHEN ATTEMPTED. AS A RESULT, DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE, IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17753768 ↗MalfunctionAug 29, 2023Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO TO CHARGE FOR DEFIBRILLATION ENERGY. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17639575 ↗MalfunctionAug 16, 2023Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED. AFTER A SHORT DELAY ANOTHER DEVICE WAS LOCATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17559906 ↗MalfunctionJul 5, 2023Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT DEFIBRILLATION TOLERANCE IS OUT OF RANGE ON THEIR DEVICE. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17258999 ↗MalfunctionJun 30, 2023Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17238192 ↗MalfunctionApr 26, 2023Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT CONTROLS ON THEIR DEVICE ARE INOPERATIVE. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16814013 ↗MalfunctionFeb 9, 2023Failure to Sense; Failure to Sense
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEY WERE USING THEIR DEVICE DURING A PATIENT EVENT AND WERE NOT ABLE TO GET A GOOD ECG RHYTHM WHEN CONNECTING THE DEFIBRILLATION ELECTRODES TO THE PATIENT. AS A RESULT, DEFIBRILLATION THERAPY MAY HAVE BEEN DELAYED OR UNAVAILABLE, IF NEEDED. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16336993 ↗MalfunctionFeb 8, 2023Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD AN EVENT CODE LOGGED IN ITS MEMORY. THE EVENT CODE LOGGED IN THE DEVICE'S MEMORY IS INDICATIVE OF A FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. AS A RESULT, DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE, IF NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16328900 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page