LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001934)Recalls & reported problems

Physio-Control, Inc. · 15AABBABBABBAAAAAAAAAAAFLP15 MONITOR/DEFIB · Class 3

All models of LIFEPAK 15 monitor/defibrillator (93)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00883873872055), received between Nov 8, 2018 and Oct 27, 2023 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Defibrillation/Stimulation Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defibrillation/Stimulation Problem2
  • Failure to Power Up2
  • Failure to Deliver Energy1
  • Failure to Deliver Shock/Stimulation1
  • Inappropriate or Unexpected Reset1
  • Unexpected Shutdown1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 27, 2023Failure to Power Up; Unexpected Shutdown; Failure to Power Up; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS NOT ABLE TO POWER ON OCCASIONALLY AND THAT THE DEVICE UNEXPECTEDLY LOST POWER. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY IF IT WAS NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18025914 ↗
  • MalfunctionOct 20, 2023Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE DELIVER ENERGY. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17976248 ↗
  • MalfunctionMay 20, 2022Failure to Deliver Energy; Failure to Deliver Energy

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD AN EVENT CODE LOGGED IN ITS MEMORY. THE EVENT CODE LOGGED IN THE DEVICE'S MEMORY IS INDICATIVE OF A FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. AS A RESULT, DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14461876 ↗
  • MalfunctionMay 18, 2022Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE LOGGED A CRITICAL EVENT CODE. THE EVENT CODE IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14431409 ↗
  • MalfunctionMay 13, 2021Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON DESPITE CHANGING THE BATTERIES. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11822792 ↗
  • MalfunctionJan 6, 2020Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NOT SHOCK A PATIENT WHEN ATTEMPTED. A BACKUP DEVICE WAS AVAILABLE AND USED TO PROVIDE DEFIBRILLATION THE PATIENT. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9556112 ↗
  • MalfunctionNov 8, 2018Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE RE-BOOTS ITSELF APPROXIMATELY EVERY 30 SECONDS AND WILL NOT DISPLAY ECG. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 8052250 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →