LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001930)Recalls & reported problems
Physio-Control, Inc. · 15AAAAAAAAABAAAAAAAAAAAFLP15 MONITOR/DEFIB · Class 3
All models of LIFEPAK 15 monitor/defibrillator (93)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00883873872093), received between Oct 23, 2018 and Aug 12, 2024 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Defibrillation/Stimulation Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Defibrillation/Stimulation Problem6
- Electrical /Electronic Property Problem2
- Failure to Deliver Shock/Stimulation2
- Key or Button Unresponsive/not Working2
- Display or Visual Feedback Problem1
- Failure to Power Up1
- Inappropriate/Inadequate Shock/Stimulation1
- Unexpected Shutdown1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 12, 2024Failure to Deliver Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation; Failure to Deliver Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT A SERVICE INDICATOR IS PRESENT AND EVENT CODES ARE LOGGED IN THE MEMORY OF THEIR DEVICE. THE EVENT CODES INDICATE A DEVICE FAILURE THAT COULD RESULT IN A PARTIAL LOSS OF DEFIBRILLATOR OUTPUT ENERGY DUE TO A LOSS OF THE NEGATIVE PORTION OF THE BIPHASIC OUTPUT WAVEFORM, RESULTING IN A MONOPHASIC SHOCK, AS WELL AS A POTENTIAL FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. UPON EVALUATION OF THE CUSTOMER'S DEVICE, STRYKER OBSERVED THAT THE DEVICE WOULD NOT DELIVER DEFIBRILLATION ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19959087 ↗MalfunctionJul 1, 2024Defibrillation/Stimulation Problem; Key or Button Unresponsive/not Working; Defibrillation/Stimulation Problem; Key or Button Unresponsive/not Working
THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE'S POWER BUTTON WAS FUNCTIONING INTERMITTENTLY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19651990 ↗MalfunctionJun 14, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19538274 ↗MalfunctionJun 6, 2024Display or Visual Feedback Problem; Electrical /Electronic Property Problem; Display or Visual Feedback Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT COMPLETE THE BOOT UP CYCLE AND THE SCREEN REMAINED BLANK. IN THIS STATE THE DEVICE WOULD NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19475848 ↗MalfunctionFeb 14, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE LOGGED A CRITICAL EVENT CODE. THE EVENT CODE IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18702273 ↗MalfunctionDec 15, 2023Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY, WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE DELIVERING ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18333644 ↗MalfunctionOct 31, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING EVALUATION STRYKER OBSERVED THAT THE DEVICE WILL NOT RECOGNIZE THERAPY CABLE. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18041067 ↗MalfunctionOct 17, 2023Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S MAIN KEYPAD IS NOT FUNCTIONING. IN THIS STATE, THE DEVICE MAY BE UNABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED DURING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17947708 ↗MalfunctionOct 9, 2023Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE WOULD NOT ESCALATE TO APPROPRIATE DEFIBRILLATION ENERGY. IN THIS STATE, WRONG DEFIBRILLATION THERAPY MAY BE DELIVERED. THIS ISSUE IS PATIENT-RELATED; HOWEVER, THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17895656 ↗MalfunctionJan 4, 2023Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR SERVICE INDICATOR IS ILLUMINATED. UPON EVALUATION OF THE CUSTOMER'S DEVICE, STRYKER OBSERVED AN EVENT CODE LOGGED IN THE DEVICE'S MEMORY THAT IS RELATED TO A POTENTIAL FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. STRYKER ALSO OBSERVED ANOTHER EVENT CODE LOGGED IN THE DEVICE'S MEMORY THAT IS INDICATIVE OF A DEVICE FAILURE THAT COULD RESULT IN A PARTIAL LOSS OF DEFIBRILLATOR OUTPUT ENERGY DUE TO A LOSS OF THE NEGATIVE PORTION OF THE BIPHASIC OUTPUT WAVEFORM, RESULTING IN A MONOPHASIC SHOCK. ADDITIONALLY, THE DEFIBRILLATOR OUTPUT ENERGY MAY BE REDUCED BY UP TO APPROXIMATELY 20% FROM THE SELECTED ENERGY LEVEL. AS A RESULT, THE WRONG DEFIBRILLATION THERAPY WOULD BE DELIVERED, IF IT WERE NECESSARY. THERE WAS NO PATIENT INVOLVEMENT REPORTED DURING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16099686 ↗MalfunctionOct 5, 2022Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15544510 ↗MalfunctionOct 28, 2020Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE POWERED OFF BY ITSELF DURING A PATIENT EVENT. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10753105 ↗MalfunctionOct 23, 2018Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WAS NOT SHOCKING PROPERLY DURING AN EVENT. THE CUSTOMER ADVISED THAT PERSONNEL ATTEMPTED TO CHANGE THE JOULES BUT THAT IT CONTINUED TO NOT WORK. THE CUSTOMER ADVISED THAT HE HAD NO FURTHER DETAILS ABOUT THE REPORTED ISSUE. THE CUSTOMER FURTHER ADVISED THAT THERE WERE NO ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE. A BACKUP DEVICE, ALSO A LIFEPAK 15, WAS AVAILABLE AND USED TO CONTINUE PATIENT CARE. THE CUSTOMER ADVISED THAT HE HAD NO FURTHER INFORMATION REGARDING THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7994870 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page