Skip to content

LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001588)Recalls & reported problems

Physio-Control, Inc. · 15AAEBABBABBBAAAAAAAAAAFLP15 MONITOR/DEFIB · Class 3

All models of LIFEPAK 15 monitor/defibrillator (93)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Physio-Control LIFEPAK 15 monitor/defibrillator has 108 adverse event reports, with 2 deaths and no injuries. Many reports describe the unit shutting down unexpectedly, failing to power on, or not responding to buttons, and some describe energy output problems that could delay or prevent shock therapy. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
108
Deaths reported
2
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 108 adverse event reports that list this exact device (DI 00883873877081), received between May 18, 2018 and Aug 6, 2025 — classified as 2 deaths, 106 malfunctions.

The most frequently reported specific device problem is “Unexpected Shutdown” (46 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 5 reportsJan 2024: 1 reportsFeb 2024: 4 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 3 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unexpected Shutdown46
  • Defibrillation/Stimulation Problem21
  • Failure to Power Up14
  • Electrical /Electronic Property Problem11
  • Failure to Deliver Shock/Stimulation5
  • Failure to Sense4
  • Device Sensing Problem3
  • Failure To Run On AC/DC2

Patient problems

  • No Clinical Signs, Symptoms or Conditions86
  • No Known Impact Or Consequence To Patient12
  • No Patient Involvement7
  • Arrhythmia1
  • Death1
  • Ventricular Fibrillation1

Recent adverse event reports

The 25 most recent of 108 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2025Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    DURING ROUTINE PREVENTATIVE MAINTENANCE TESTING BY STRYKER, A LOANER DEVICE SHOWED AN ERROR CODE IN THE SERVICE LOG. THIS ERROR CODE IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22724758 ↗
  • MalfunctionOct 11, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT RESPOND TO BUTTON PRESSES WHEN ATTEMPTED. AS A RESULT, DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE, IF NEEDED. THERE WERE NO REPORTS OF ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20432355 ↗
  • MalfunctionOct 1, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT COMPLETE ITS BOOT UP CYCLE. AS A RESULT, DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20340994 ↗
  • MalfunctionSep 30, 2024Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION, STRYKER OBSERVED THAT THE DEVICE HAD AN EVENT CODE LOGGED IN THE DEVICE'S MEMORY THAT IS INDICATIVE OF A DEVICE FAILURE THAT COULD RESULT IN A PARTIAL LOSS OF DEFIBRILLATOR OUTPUT ENERGY DUE TO A LOSS OF THE NEGATIVE PORTION OF THE BIPHASIC OUTPUT WAVEFORM, RESULTING IN A MONOPHASIC SHOCK. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20338110 ↗
  • MalfunctionSep 9, 2024Failure to Power Up; Failure to Power Up

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. AS A RESULT, DEFIBRILLATION WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20168264 ↗
  • MalfunctionSep 6, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION OF THE CUSTOMER¿S DEVICE, STRYKER OBSERVED THAT THE DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20155061 ↗
  • MalfunctionJun 13, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY POWERED OFF MULTIPLE TIMES. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19529337 ↗
  • MalfunctionJun 3, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WILL NOT POWER ON. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19454675 ↗
  • MalfunctionMay 14, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION OF THE CUSTOMER¿S DEVICE, STRYKER OBSERVED THAT THE DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19308145 ↗
  • MalfunctionMay 2, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DOES NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19228050 ↗
  • MalfunctionMay 1, 2024Defibrillation/Stimulation Problem; Unexpected Shutdown; Defibrillation/Stimulation Problem; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY LOST POWER AND HAS LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE DELIVER ENERGY. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19220721 ↗
  • MalfunctionApr 30, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE LOGGED A CRITICAL EVENT CODE. THE EVENT CODE IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19211263 ↗
  • MalfunctionApr 22, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD AN EVENT CODE LOGGED IN ITS MEMORY. THE EVENT CODE LOGGED IN THE DEVICE'S MEMORY IS INDICATIVE OF A FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. AS A RESULT, DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19152742 ↗
  • MalfunctionApr 2, 2024Failure to Power Up; Unexpected Shutdown; Failure to Power Up; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEY WERE USING THEIR DEVICE TO MONITOR A PATIENT AND THE DEVICE UNEXPECTEDLY POWERED OFF AND WOULDN'T POWER BACK ON. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19020091 ↗
  • MalfunctionMar 19, 2024Failure to Sense; Failure to Sense

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE LOST CONNECTION WITH THE PADS DURING A PATIENT INVOLVED EVENT. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18932045 ↗
  • MalfunctionFeb 26, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION OF THE CUSTOMER¿S DEVICE, STRYKER OBSERVED THAT THE DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18784508 ↗
  • MalfunctionFeb 14, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    A STRYKER SERVICE REPRESENTATIVE WAS PERFORMING ROUTINE PREVENTATIVE MAINTENANCE ON A STRYKER OWNED LOANER DEVICE AND OBSERVED AN EVENT CODE LOGGED IN THE DEVICE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18701726 ↗
  • MalfunctionFeb 14, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON INSPECTION, STRYKER OBSERVED THAT THE DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18701002 ↗
  • MalfunctionFeb 6, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE POWERED OFF BY ITSELF. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18649186 ↗
  • MalfunctionJan 8, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DOES NOT COMPLETING BOOT-UP CYCLE DURING INITIAL POWER ON. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18466467 ↗
  • MalfunctionDec 19, 2023Defibrillation/Stimulation Problem; Unexpected Shutdown; Defibrillation/Stimulation Problem; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY SHUT DOWN AND HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18357201 ↗
  • MalfunctionDec 19, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THE BATTERY PIN OF THEIR DEVICE WAS MISSING. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18354031 ↗
  • MalfunctionDec 14, 2023Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED TO POWER ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18323910 ↗
  • MalfunctionDec 6, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD A FROZEN SCREEN. THIS CAN BE INDICATIVE OF A DEVICE LOCK-UP. AS A RESULT, DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18272924 ↗
  • MalfunctionDec 6, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S BATTERY PIN BROKEN. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18272556 ↗

FDA clearance and classification

The FDA regulates product code MKJ (automated external defibrillators (non-wearable)) under 21 CFR 870.5310 as a Class III device. 21 CFR 870.5310 ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

FDA definition of this device type: This device is a non-wearable prescription use only aed. These are devices that include automated external defibrillation. Automated external defibrillators use external pad-type electrodes to sense, detect, classify and treat (with an electrical shock) ventricular fibrillation. These devices are intended to be used on suspected victims of sudden cardiac arrest. A person in cardiac arrest is unresponsive and is not breathing normally. The device can be sold with prescription only.

Common questions about the LIFEPAK® 15 monitor/defibrillator 99577

Has the LIFEPAK® 15 monitor/defibrillator 99577 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00883873877081), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the LIFEPAK® 15 monitor/defibrillator 99577?

The FDA's MAUDE database lists 108 adverse event reports for this device, received between May 18, 2018 and Aug 6, 2025. In those reports, the most common specific device problems are “Unexpected Shutdown” (46), “Defibrillation/Stimulation Problem” (21) and “Failure to Power Up” (14); the most common patient problems are “Arrhythmia” (1), “Death” (1) and “Ventricular Fibrillation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the LIFEPAK® 15 monitor/defibrillator 99577?

2 reports are classified as deaths, 0 as injuries and 106 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the LIFEPAK® 15 monitor/defibrillator 99577 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the LIFEPAK® 15 monitor/defibrillator 99577, and is it still sold?

The FDA's device database lists Physio-Control, Inc. as the company responsible for it (brand LIFEPAK® 15 monitor/defibrillator, model 99577). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page