LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001373)Recalls & reported problems

Physio-Control, Inc. · 15AAEBABBABBBBAAAAAAAAAFLP15 MONITOR/DEFIB · Class 3

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Physio-Control LIFEPAK 15 monitor/defibrillator has 109 adverse event reports, with 3 deaths and no injuries. Many reports describe the unit shutting down unexpectedly, failing to power on, or keypad buttons not working, and some say this could delay or prevent shock therapy. 1 recall is open, started in September 2023, over battery pins that may loosen and cause power problems.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
109
Deaths reported
3
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 109 adverse event reports that list this exact device (DI 00883873884331), received between Apr 4, 2018 and Aug 22, 2024 — classified as 3 deaths, 106 malfunctions.

The most frequently reported specific device problem is “Unexpected Shutdown” (50 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 4 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 29, 2023Open, Classified

    Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.

    Z-0300-2024 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Unexpected Shutdown50
  • Failure to Power Up11
  • Defibrillation/Stimulation Problem10
  • Electrical /Electronic Property Problem8
  • Device Sensing Problem6
  • Failure to Charge5
  • Failure to Deliver Shock/Stimulation4
  • Failure to Sense4

Patient problems

  • No Clinical Signs, Symptoms or Conditions72
  • No Known Impact Or Consequence To Patient19
  • No Patient Involvement12
  • Cardiac Arrest4
  • Death1
  • No Consequences Or Impact To Patient1
  • Respiratory Failure1
  • Ventricular Fibrillation1

Recent adverse event reports

The 25 most recent of 109 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 22, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT NONE OF THEIR DEVICE'S KEYPAD BUTTONS WERE WORKING DURING A PATIENT EVENT. AS A RESULT, DEFIBRILLATION THERAPY MAY HAVE BEEN DELAYED OR UNAVAILABLE, IF NEEDED. THEY WERE ABLE TO SWITCH TO A BACK UP DEVICE. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20025140 ↗
  • MalfunctionAug 13, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE MAIN KEYPAD OF THEIR DEVICE WOULD NOT WORK. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19969707 ↗
  • MalfunctionAug 9, 2024Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19947057 ↗
  • MalfunctionJul 24, 2024Device Sensing Problem; Device Sensing Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE PADDLES LEAD ECG WENT TO DASHED LINE DURING TRANSPORT. THIS ISSUE HAS THE POTENTIAL TO EITHER DELAY, OR PREVENT, DEFIBRILLATION FROM BEING DELIVERED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19819618 ↗
  • MalfunctionJul 19, 2024Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE'S BATTERY COULD NOT BE INSERTED PROPERLY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19784648 ↗
  • MalfunctionJun 26, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE DEVICE WOULD INTERMITTENTLY TURN OFF WHEN CONNECTED TO AC POWER. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19614824 ↗
  • MalfunctionJun 12, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY LOST POWER DURING INTERVENTION ON A PATIENT. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19522727 ↗
  • MalfunctionJun 10, 2024Device Sensing Problem; Device Sensing Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DOES NOT RECOGNIZE PADDLES. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19506273 ↗
  • MalfunctionMay 19, 2024Unexpected Shutdown; Unexpected Shutdown

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD POWER CYCLED DURING USE. AS A RESULT, DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19348580 ↗
  • MalfunctionApr 23, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DOES NOT COMPLETING BOOT-UP CYCLE DURING INITIAL POWER ON AND THAT THE DEVICE HAS BLANK SCREEN. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19162389 ↗
  • MalfunctionApr 8, 2024Failure to Power Up; Failure to Power Up

    A CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON INSPECTION, IT WAS OBSERVED THAT THE DEVICE WOULD NOT POWER ON. AS A RESULT, DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19067461 ↗
  • MalfunctionApr 3, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY LOST POWER AND REBOOTED ITSELF SEVERAL TIMES. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO PATIENT INVOLVEMENT IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19030531 ↗
  • MalfunctionMar 22, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD UNEXPECTEDLY LOST POWER TWICE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18961194 ↗
  • MalfunctionFeb 7, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING EVALUATION STRYKER OBSERVED THAT THEIR DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE DELIVER ENERGY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18657024 ↗
  • MalfunctionJan 22, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY LOST POWER. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18557121 ↗
  • MalfunctionJan 19, 2024Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY LOST POWER DURING INTERVENTION ON A PATIENT. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18547804 ↗
  • MalfunctionJan 11, 2024Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    A STRYKER SERVICE TECHNICIAN PERFORMED PREVENTIVE MAINTENANCE OF DEVICE. DURING EVALUATION IT WAS OBSERVED THAT THE DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE DELIVER ENERGY. THIS ISSUE HAS THE POTENTIAL TO EITHER DELAY, OR PREVENT, DEFIBRILLATION FROM BEING DELIVERED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18494264 ↗
  • MalfunctionJan 10, 2024Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR WOULD NOT DELIVER DEFRIBRILLATION ENERGY. AS A RESULT, APPROPRIATE DEFIBRILLATION THERAPY WOULD NOT BE DELIVERED, IF NEEDED. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18484502 ↗
  • DeathDec 20, 2023Device Sensing Problem; Device Sensing Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT DURING A PATIENT EVENT THEIR DEVICE WAS UNABLE TO DETECT THE PATIENT THROUGH THE DEFIBRILLATION ELECTRODES. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED. THE CUSTOMER ADVISED THAT THE DEVICE PROMPTED THEM TO ¿CONNECT ELECTRODES¿ WHEN THE DEFIBRILLATION ELECTRODES WERE CONNECTED TO THE PATIENT. THE CREW UNPLUGGED THE CABLES FROM THE DEVICE AND THE DEFIBRILLATION ELECTRODES AND RECONNECTED THEM. THE DEFIBRILLATION ELECTRODES WERE REPLACED. HOWEVER, THE ISSUE REMAINED. THE CABLES WERE SWITCHED WITH ANOTHER DEVICE TO TREAT THE PATIENT. THE PATIENT WAS PRONOUNCED DEAD AT THE HOSPITAL AFTER ONGOING RESUSCITATION AT THE EMERGENCY DEPARTMENT.

    Patient problems: Cardiac Arrest

    MAUDE report 18364846 ↗
  • MalfunctionNov 14, 2023Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE DEVICE HAD LOGGED AN EVENT CODE INTO ITS MEMORY THAT'S INDICATIVE OF A POTENTIAL FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18130224 ↗
  • MalfunctionOct 6, 2023Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DOES NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17886086 ↗
  • MalfunctionSep 15, 2023Failure to Power Up; Failure to Power Up

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. AS A RESULT, DEFIBRILLATION WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17759483 ↗
  • MalfunctionSep 7, 2023Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT COMPLETE ITS INITIAL BOOT-UP CYCLE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17702525 ↗
  • MalfunctionAug 18, 2023Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY LOST POWER DURING INTERVENTION ON A PATIENT. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17570281 ↗
  • MalfunctionJun 28, 2023Material Separation; Material Separation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT A BATTERY PIN OF THEIR DEVICE WAS PUSHED IN. IN THIS STATE THE DEVICE HAS THE POTENTIAL TO POWER OFF BY ITSELF. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17215028 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →