Skip to content

LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001242)Recalls & reported problems

Physio-Control, Inc. · 15AABAABAABBBAAAAAAAAAABLP15 MONITOR/DEFIB · Class 3

All models of LIFEPAK 15 monitor/defibrillator (93)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00883873911778), received between Jan 24, 2019 and Mar 2, 2026 — classified as 13 malfunctions.

The most frequently reported specific device problem is “Failure to Deliver Shock/Stimulation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Deliver Shock/Stimulation3
  • Electrical /Electronic Property Problem2
  • Energy Output Problem2
  • Failure to Deliver Energy2
  • Failure to Power Up2
  • Inappropriate/Inadequate Shock/Stimulation2
  • Failure to Charge1
  • Key or Button Unresponsive/not Working1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • No Patient Involvement3

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 2, 2026Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE POWER BUTTON OF THEIR DEVICE WAS STUCK. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED.THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24478685 ↗
  • MalfunctionSep 3, 2025Failure to Charge; Failure to Power Up; Inappropriate/Inadequate Shock/Stimulation; Failure to Charge; Failure to Power Up; Inappropriate/Inadequate Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DOES NOT POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22950526 ↗
  • MalfunctionJun 17, 2025Failure to Deliver Energy; Failure to Deliver Energy

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DID NOT DELIVER A SHOCK INTO A DEFIBRILLATOR ANALYZER. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22236656 ↗
  • MalfunctionDec 8, 2023Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT DEFIBRILLATE. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18289761 ↗
  • MalfunctionApr 10, 2023Energy Output Problem; Energy Output Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, STRYKER OBSERVED THAT THE DEVICE GAVE AN "ABNORMAL ENERGY DELIVERED" MESSAGE WHEN DELIVERING DEFIBRILLATION. AS A RESULT, THE WRONG DEFIBRILLATION THERAPY MAY BE DELIVERED TO A PATIENT, IF NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16708685 ↗
  • MalfunctionMar 31, 2023Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. DURING EVALUATION IT WAS FOUND THAT THE DEVICE'S DEFIBRILLATION ENERGY DELIVERY LEVEL WAS NOT AS EXPECTED. IN THIS STATE, THE DEVICE MAY DELIVER THE WRONG DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED DURING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16657208 ↗
  • MalfunctionFeb 8, 2023Energy Output Problem; Energy Output Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD AN ABNORMAL ENERGY DELIVERY. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO DELIVER APPROPRIATE DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED DURING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16319963 ↗
  • MalfunctionSep 26, 2022Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON AFTER BEING DROPPED. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15485642 ↗
  • MalfunctionMar 14, 2022Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED TO SHOCK. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. CUSTOMER PROCEEDED WITH ANOTHER DEVICE. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13760030 ↗
  • MalfunctionMar 4, 2022Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS NOT DELIVERING DEFIBRILLATION ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13668955 ↗
  • MalfunctionJul 13, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION OF THE CUSTOMER¿S DEVICE, PHYSIO-CONTROL OBSERVED THAT THE CUSTOMER'S DEVICE WAS ABLE TURN ON WITH THE POWER BUTTON; HOWEVER NO OTHER BUTTONS WERE FUNCTIONING. IN THIS STATE DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE TO A PATIENT IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 10263115 ↗
  • MalfunctionApr 2, 2019Failure to Deliver Energy; Failure to Deliver Energy

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT WHEN THEY ATTEMPT TO DELIVER A TEST SHOCK WITH THE DEVICE THEY HAVE TO PRESS THE SHOCK BUTTON SEVERAL TIMES. THEY ALSO REPORTED THAT THE DEVICE HAS AN EVENT CODE LOGGED IN ITS MEMORY. THE EVENT CODE LOGGED IS RELATED TO A POTENTIAL FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 8476518 ↗
  • MalfunctionJan 24, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 8273687 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page