LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001219)Recalls & reported problems

Physio-Control, Inc. · 15AAEBABBABBBBAAAAAAAAABLP15 MONITOR/DEFIB · Class 3

All models of LIFEPAK 15 monitor/defibrillator (93)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Physio-Control LIFEPAK 15 monitor/defibrillator has 77 adverse event reports, all malfunctions. Reports describe unexpected shutdowns, failure to power up, trouble delivering shocks, and slow or failed charging, with no patient involvement. No deaths; no injuries; no recalls. Reports dropped from 4 in the prior 12 months to 0 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
77
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 77 adverse event reports that list this exact device (DI 00883873912775), received between Jun 8, 2018 and Sep 12, 2025 — classified as 77 malfunctions.

The most frequently reported specific device problem is “Unexpected Shutdown” (29 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unexpected Shutdown29
  • Electrical /Electronic Property Problem13
  • Failure to Power Up11
  • Failure to Deliver Shock/Stimulation6
  • Inappropriate/Inadequate Shock/Stimulation4
  • Connection Problem3
  • Device Sensing Problem3
  • Failure to Charge3

Patient problems

  • No Clinical Signs, Symptoms or Conditions39
  • No Patient Involvement25
  • No Known Impact Or Consequence To Patient17

Recent adverse event reports

The 25 most recent of 77 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 12, 2025Device Sensing Problem; Device Sensing Problem

    A CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION, STRYKER OBSERVED THE DEVICE WOULD NOT RECOGNIZE THE DEFIBRILLATION ELECTRODES. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF IT WAS NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23044287 ↗
  • MalfunctionApr 29, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE SHUTS DOWN UNEXPECTEDLY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21930317 ↗
  • MalfunctionMar 3, 2025Failure to Charge; Charging Problem; Failure to Charge; Charging Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE IS NOT CHARGING TO THE SELECTED 300 JOULES ENERGY LEVEL AND IS ALSO VERY SLOW TO COMPLETE A CHARGE AT OTHER ENERGY LEVELS. AS A RESULT, DEFIBRILLATION THERAPY MAY BE DELAYED OR UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21510982 ↗
  • MalfunctionJan 15, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE SHUT DOWN UNEXPECTEDLY. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21164498 ↗
  • MalfunctionOct 16, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT CONTROLS ON THEIR DEVICE ARE UNRESPONSIVE. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20459620 ↗
  • MalfunctionSep 11, 2023Failure to Analyze Signal; Failure to Analyze Signal

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S SERVICE INDICATOR IS PRESENT AND THERE IS AN EVENT CODE LOGGED IN THE DEVICE'S MEMORY. THE EVENT CODE LOGGED IN THE DEVICE'S MEMORY CAN INDICATE THAT THE AED FUNCTION OF THE DEVICE MAY BE INOPERABLE. AS A RESULT, THE UNIT MAY NOT BE ABLE TO ANALYZE A PATIENT'S RHYTHM AND PROVIDE DEFIBRILLATION THERAPY WHILE IN AED MODE, IF IT WAS NECESSARY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17726496 ↗
  • MalfunctionJul 27, 2023Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17404823 ↗
  • MalfunctionJul 21, 2023Electrical /Electronic Property Problem; Failure to Power Up; Electrical /Electronic Property Problem; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE SCREEN FROZE WHEN POWERED UP. IN THIS STATE, THE DEVICE MAY BE UNABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED DURING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17368800 ↗
  • MalfunctionMar 9, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE IS NOT COMPLETING ITS BOOT UP CYCLE. AS A RESULT, DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16512066 ↗
  • MalfunctionMar 3, 2023Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY POWERED OFF. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED DURING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16482637 ↗
  • MalfunctionAug 22, 2022Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY SHUT DOWN. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15271088 ↗
  • MalfunctionJul 15, 2022Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED TO POWER ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15026688 ↗
  • MalfunctionJul 14, 2022Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT FULLY BOOT UP. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15018480 ↗
  • MalfunctionJun 29, 2022Failure to Deliver Shock/Stimulation; Connection Problem; Failure to Deliver Shock/Stimulation; Connection Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE IS FAILING TO DELIVER A SHOCK AND WON'T DETECT THE ATTACHED PADDLES OR THERAPY CABLE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14853625 ↗
  • MalfunctionApr 1, 2022Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED TO DELIVER A SHOCK AS EXPECTED. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13978303 ↗
  • MalfunctionMar 26, 2022Connection Problem; Connection Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT RECOGNIZE THE CONNECTED DEFIBRILLATION ELECTRODES DURING A PATIENT EVENT. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13923522 ↗
  • MalfunctionMar 20, 2022Failure to Deliver Shock/Stimulation; Failure to Sense; Failure to Deliver Shock/Stimulation; Failure to Sense

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEY WERE USING THEIR DEVICE ON A PATIENT AND IT WOULD NOT SHOW THE PATIENT'S ECG THROUGH THE PADDLES LEAD. ALSO, THE DEVICE FAILED TO DELIVER A SHOCK TO THE PATIENT WHEN PROMPTED. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13831396 ↗
  • MalfunctionFeb 1, 2022Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY POWERED OFF DURING A PATIENT EVENT. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13415924 ↗
  • MalfunctionJan 28, 2022Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE POWERED OFF ITSELF DURING A PATIENT EVENT. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13389374 ↗
  • MalfunctionDec 6, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THE DEVICE'S KEYPAD BUTTONS WERE NOT FUNCTIONING. AS A RESULT, DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12935309 ↗
  • MalfunctionSep 15, 2021Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE ON/OFF BUTTON IS NOT FUNCTIONING OPTIMALLY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO POWER ON AND DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12478763 ↗
  • MalfunctionAug 28, 2021Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12381274 ↗
  • MalfunctionAug 20, 2021Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE SHUTDOWN ON IT'S OWN. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12351615 ↗
  • MalfunctionAug 20, 2021Failure to Charge; Failure to Charge

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, PHYSIO-CONTROL OBSERVED THAT THE DEVICE HAD AN EVENT CODE LOGGED IN ITS MEMORY. THE EVENT CODE LOGGED IN THE DEVICE'S MEMORY IS INDICATIVE OF A FAILURE OF THE DEVICE TO CHARGE DEFIBRILLATION ENERGY. AS A RESULT, DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE, IF NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12351113 ↗
  • MalfunctionJul 27, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WAS STUCK ON THE SET UP MODE SCREEN. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12229538 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →