LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001217)Recalls & reported problems

Physio-Control, Inc. · 15AABBABBABBBBAAAAAAAAABLP15 MONITOR/DEFIB · Class 3

All models of LIFEPAK 15 monitor/defibrillator (93)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a Physio-Control LIFEPAK 15 monitor and defibrillator. Most of the 84 reports describe the device shutting down unexpectedly, not powering on, buttons not responding and failing to deliver a shock, some of them during patient use. There are 2 deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
84
Deaths reported
2
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 84 adverse event reports that list this exact device (DI 00883873912799), received between Aug 6, 2018 and Jul 19, 2026 — classified as 2 deaths, 82 malfunctions.

The most frequently reported specific device problem is “Unexpected Shutdown” (46 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unexpected Shutdown46
  • Electrical /Electronic Property Problem17
  • Failure to Deliver Energy5
  • Failure to Deliver Shock/Stimulation5
  • Failure to Power Up4
  • Key or Button Unresponsive/not Working3
  • Defibrillation/Stimulation Problem2
  • Device Sensing Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions44
  • No Known Impact Or Consequence To Patient21
  • No Patient Involvement15
  • Death3
  • No Consequences Or Impact To Patient3
  • Angina1
  • Ventricular Fibrillation1

Recent adverse event reports

The 25 most recent of 84 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 19, 2026Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, STRYKER OBSERVED AN EVENT CODE LOGGED IN THE DEVICE'S MEMORY. THE EVENT CODE LOGGED IS RELATED TO A POTENTIAL FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25867977 ↗
  • MalfunctionMay 28, 2026Device Sensing Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE WOULD NOT LET THE USER SELECT DEFIBRILLATION PADDLES WHEN CONNECTED. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25317233 ↗
  • MalfunctionFeb 4, 2026Fracture

    A CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON INSPECTION, IT WAS OBSERVED THAT THE POWER CONNECTOR WAS BROKEN. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24256823 ↗
  • MalfunctionMay 22, 2025Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FREEZES WHEN DUMPING THE DEFIBRILLATION ENERGY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22072462 ↗
  • MalfunctionNov 17, 2023Electrical /Electronic Property Problem; Key or Button Unresponsive/not Working; Electrical /Electronic Property Problem; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE ANALYZED AND PROMPTED THEM TO DELIVER A SHOCK EVEN THOUGH THE ANALYZE BUTTON WASN'T PRESSED. IT WAS ALSO REPORTED THAT THE DEVICE'S KEYPAD IS INTERMITTENTLY NOT WORKING. AS A RESULT, DEFIBRILLATION THERAPY MAY BE DELAYED OR UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18161913 ↗
  • MalfunctionOct 6, 2022Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION OF THE CUSTOMER¿S DEVICE, STRYKER OBSERVED THAT THE DEVICE'S MAIN KEYPAD WAS DEFECTIVE. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15554015 ↗
  • MalfunctionSep 14, 2022Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15422231 ↗
  • MalfunctionAug 31, 2022Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY SHUT OFF DURING PATIENT USE. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15329144 ↗
  • MalfunctionAug 8, 2022Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY POWERED OFF. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15183137 ↗
  • MalfunctionJul 14, 2022Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE FAILED TO DELIVER A SHOCK AS EXPECTED DURING TRAINING. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15018655 ↗
  • MalfunctionJun 13, 2022Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S MAIN KEYPAD DOES NOT FUNCTION. IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER, THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14678086 ↗
  • MalfunctionApr 12, 2022Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD AN UNEXPECTED LOSS OF POWER. IN THIS STATE, THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14088025 ↗
  • MalfunctionMar 20, 2022Electrical /Electronic Property Problem; Unexpected Shutdown; Electrical /Electronic Property Problem; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD A BLANK SCREEN AND WAS NOT RESPONSIVE. AS A RESULT, DEFIBRILLATION THERAPY MAY HAVE BEEN UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13831094 ↗
  • MalfunctionMar 16, 2022Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE POWERED OFF BY ITSELF DURING A PATIENT EVENT. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED BY THE CUSTOMER.

    Patient problems: Angina

    MAUDE report 13779730 ↗
  • MalfunctionMar 8, 2022Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE UNEXPECTEDLY SHUT DOWN. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13694607 ↗
  • MalfunctionJan 28, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS UNABLE TO DELIVER DEFIBRILLATION THERAPY DURING A PATIENT EVENT. THE CUSTOMER ADVISED THAT THE DEVICE WAS CHARGED TO 100J, WHEN "SYNCH" WAS PRESSED THE MONITOR FLICKERED, THEN THE SCREEN TURNED BLACK. AFTER A FEW SECONDS THE SCREEN RETURNED. THE DEVICE WAS CHARGED TO 100J, THEN DUMPED, AND CHARGED TO 360J WHEN MONITOR FLICKERED AGAIN AND POWER CYCLED. THE SERVICE INDICATOR WAS THEN ILLUMINATED. IN ADDITION THE DEVICE LOGGED A POTENTIALLY CRITICAL EVENT CODE THAT IS INDICATIVE OF A DEVICE LOCKUP. THIS ISSUE IS PATIENT RELATED; HOWEVER THE PATIENT WAS SAVED BY THE MEDICS.

    Patient problems: Ventricular Fibrillation

    MAUDE report 13389380 ↗
  • MalfunctionJan 4, 2022Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY POWERED OFF DURING A PATIENT EVENT. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13158955 ↗
  • MalfunctionDec 22, 2021Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. UPON EVALUATION OF THE CUSTOMER'S DEVICE, STRYKER OBSERVED AN ERROR MESSAGE INDICATING ABNORMAL ENERGY DELIVERED. IN THIS STATE DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF IT IS NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13067184 ↗
  • MalfunctionDec 17, 2021Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE UNEXPECTEDLY POWERED OFF. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13029270 ↗
  • MalfunctionSep 17, 2021Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE UNEXPECTEDLY POWERED OFF DURING A PATIENT EVENT. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12492866 ↗
  • MalfunctionSep 17, 2021Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE UNEXPECTEDLY POWERED OFF DURING A PATIENT EVENT. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12492823 ↗
  • MalfunctionAug 31, 2021Failure to Deliver Energy; Failure to Deliver Energy

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, PHYSIO-CONTROL OBSERVED THAT THE DEVICE WOULD NOT DELIVER DEFIBRILLATION ENERGY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12395294 ↗
  • MalfunctionJul 7, 2021Energy Output Problem; Energy Output Problem

    PHYSIO-CONTROL WAS PERFORMING PREVENTATIVE MAINTENANCE ON A CUSTOMER'S DEVICE AND OBSERVED THE DEVICE GAVE AN "ABNORMAL ENERGY DELIVERED" MESSAGE WHEN PERFORMING A TEST SHOCK AND THE ENERGY DELIVERED WAS 240 JOULES AT A SELECTED ENERGY LEVEL OF 200 JOULES. AS A RESULT, THE WRONG DEFIBRILLATION THERAPY MAY BE DELIVERED TO A PATIENT. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12129278 ↗
  • MalfunctionJun 21, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE'S SCREEN WENT BLANK, THE CREW NEEDED TO USE THE POWER BUTTON TO GET THE SCREEN BACK ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER NO HEALTH CONSEQUENCES OR IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12037071 ↗
  • MalfunctionJun 14, 2021Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE POWERED OFF BY ITSELF TWICE DURING A PATIENT EVENT. THE CUSTOMER ADVISED THIS OCCURRED WHILE MONITORING THE PATIENT'S ECG RHYTHM. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11990650 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →