LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001216)Recalls & reported problems
Physio-Control, Inc. · 15AABAABBABBBABAAAAAAAABLP15 MONITOR/DEFIB · Class 3
All models of LIFEPAK 15 monitor/defibrillator (93)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker LIFEPAK 15 monitor/defibrillator is a portable resuscitation device used in hospitals and by emergency services. The 43 reports mostly describe units failing to power on, shutting down unexpectedly, or failing to deliver the shock when needed, along with error codes logged in the memory and one report of shock being delivered at an unexpected level. One report mentions a death and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 44
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 44 adverse event reports that list this exact device (DI 00883873912805), received between Aug 15, 2018 and Sep 14, 2026 — classified as 1 death, 43 malfunctions.
The most frequently reported specific device problem is “Failure to Deliver Shock/Stimulation” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Deliver Shock/Stimulation15
- Failure to Power Up8
- Failure to Deliver Energy6
- Unexpected Shutdown6
- Electrical /Electronic Property Problem4
- Failure to Charge3
- Defibrillation/Stimulation Problem2
- Device Sensing Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions23
- No Patient Involvement11
- No Known Impact Or Consequence To Patient9
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 25 most recent of 44 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2026Failure to Charge
The customer contacted Stryker to report that their device would not charge during a patient case. In this state the device may not be able to deliver defibrillation therapy if needed. This issue is patient related; however there was no adverse event reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50756462 ↗MalfunctionAug 18, 2026Electrical /Electronic Property Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50299071 ↗MalfunctionMay 28, 2026Failure to Deliver Shock/Stimulation
A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER ENERGY. IN THIS STATE, THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY, IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25318109 ↗MalfunctionOct 28, 2025Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WILL NOT POWER ON. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23406194 ↗MalfunctionJul 25, 2025Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22609911 ↗MalfunctionApr 3, 2025Failure to Deliver Energy; Failure to Deliver Energy
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS UNABLE TO DELIVER DEFIBRILLATION ENERGY DURING A PATIENT EVENT. THE DEVICE WAS ABLE TO CHARGE TO 200J SUCCESSFULLY. THIS ISSUE IS PATIENT RELATED; HOWEVER, THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21769849 ↗MalfunctionJan 31, 2025Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21282382 ↗MalfunctionAug 28, 2024Failure to Analyze Signal; Device Sensing Problem; Failure to Analyze Signal; Device Sensing Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S SERVICE INDICATOR IS PRESENT AND THERE IS AN EVENT CODE LOGGED IN THE DEVICE'S MEMORY. THE EVENT CODE LOGGED IN THE DEVICE'S MEMORY CAN INDICATE THAT THE AED FUNCTION OF THE DEVICE MAY BE INOPERABLE. AS A RESULT, THE UNIT MAY NOT BE ABLE TO ANALYZE A PATIENT'S RHYTHM AND PROVIDE DEFIBRILLATION THERAPY WHILE IN AED MODE, IF IT WAS NECESSARY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20083846 ↗MalfunctionAug 22, 2023Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
THE CUSTOMER CONTACTED STRYKER TO REPORT THEIR DEVICE DELIVERED DEFIBRILLATION AT AN UNEXPECTED LEVEL. IN THIS STATE THE DEVICE MAY DELIVER INAPPROPRIATE DEFIBRILLATION THERAPY. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17593083 ↗MalfunctionMar 13, 2023Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS UNABLE TO DELIVER A SHOCK WHEN USING INTERNAL PADDLES DURING A PATIENT EVENT. THE CUSTOMER USED EXTERNAL PADDLES AS THE BACKUP PROCEDURE. IN THIS STATE, DEFIBRILLATION THERAPY MAY BE DELAYED OR PREVENTED FROM BEING DELIVERED TO THE PATIENT IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER, THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16530708 ↗MalfunctionJan 10, 2023Failure to Deliver Energy; Failure to Deliver Energy
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S SERVICE INDICATOR IS ILLUMINATED AND THERE WAS AN EVENT CODE LOGGED IN THE DEVICE'S MEMORY THAT IS RELATED TO A POTENTIAL FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16132497 ↗MalfunctionNov 30, 2022Failure to Deliver Shock/Stimulation; Key or Button Unresponsive/not Working; Failure to Deliver Shock/Stimulation; Key or Button Unresponsive/not Working
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE DID NOT PROVIDE A SHOCK WHEN THE SHOCK BUTTON WAS PRESSED. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15890650 ↗MalfunctionJul 15, 2022Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WAS NOT POWERING ON. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY MAY NOT BE AVAILABLE IF NEEDED. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15026690 ↗DeathNov 2, 2021Failure to Deliver Shock/Stimulation; Unexpected Shutdown; Failure to Deliver Shock/Stimulation; Unexpected Shutdown
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEY WERE USING THIS DEVICE ON A PATIENT AND WHEN THEY PRESSED THE DEVICE'S SHOCK BUTTON TO DELIVER A SHOCK TO THE PATIENT, THE DEVICE UNEXPECTEDLY POWERED OFF AND THEN POWER CYCLED. AS A RESULT, DEFIBRILLATION THERAPY WAS DELAYED. THE CUSTOMER INFORMED PHYSIO-CONTROL THAT THE PATIENT INVOLVED IN THE REPORTED EVENT IS DECEASED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12739139 ↗MalfunctionJun 14, 2021Power Problem; Power Problem
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THE POWER PORT ON THE BACK OF THEIR DEVICE IS BROKEN AND THE DEVICE WOULD NOT POWER ON. THERE WAS NO PATIENT INVOLVEMENT REPORTED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11995452 ↗MalfunctionMay 28, 2021Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE POWERED OFF BY ITSELF DURING TESTING. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11901046 ↗MalfunctionMay 4, 2021Failure to Power Up; Failure to Power Up
PHYSIO-CONTROL RECEIVED A REPORT THAT "ON/OFF BUTTON NOT RESPONDING". IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED. THERE IS NO PATIENT INVOLVEMENT IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11771298 ↗MalfunctionMar 8, 2021Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11441094 ↗MalfunctionMar 3, 2021Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE FAILED TO DELIVER A SHOCK TO A PATIENT WHEN THE USER PRESSED THE SHOCK BUTTON. THE SHOCK BUTTON WAS PRESSED A SECOND TIME AND THEN A SHOCK WAS SUCCESSFULLY DELIVERED TO THE PATIENT. AS A RESULT, DEFIBRILLATION THERAPY MAY BE DELAYED OR UNAVAILABLE, IF NEEDED. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11408355 ↗MalfunctionFeb 17, 2021Failure to Deliver Shock/Stimulation; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE HAD LOGGED AN EVENT CODE IN THE MEMORY WHICH IS RELATED TO A POTENTIAL ISSUE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11338394 ↗MalfunctionDec 8, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE UNEXPECTEDLY SHUTDOWN, AND DISPLAYED A SERVICE CODE. THE SERVICE CODE MAY BE INDICATIVE OF A DEVICE LOCKUP OR RESET. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED, HOWEVER, THERE WAS NO ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10970088 ↗MalfunctionNov 6, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE'S BUTTONS ARE NOT FUNCTIONING. AS A RESULT, DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10801914 ↗MalfunctionJun 22, 2020Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10182578 ↗MalfunctionJun 16, 2020Electrical Power Problem; Electrical Power Problem
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THE ON/OFF BUTTON OF THEIR DEVICE WAS DIFFICULT TO ACTIVATE. THERE WAS NO PATIENT USE REPORTED WITH THE EVENT. PHYSIO-CONTROL EVALUATED THE DEVICE AND FOUND THE ON/OFF BUTTON WAS INOPERABLE. IN THIS STATE THE DEVICE WOULD BE INOPERABLE AND DEFIBRILLATION THERAPY WOULD NOT BE AVAILABLE IF NEEDED
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10156372 ↗MalfunctionMay 13, 2020Failure to Deliver Shock/Stimulation; Display or Visual Feedback Problem; Failure to Deliver Shock/Stimulation; Display or Visual Feedback Problem
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE FAILED TO DELIVER SHOCK AND THE ECG WAS NOT VISIBLE ON THE SCREEN. DEFIBRILLATION WAS AGAIN ATTEMPTED AND WAS SUCCESSFUL, AND ROSC WAS OBTAINED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE PATIENT OUTCOME REPORTED. THE REPORTER STATED AN INTERNAL INVESTIGATION IS ONGOING AS USE ERROR IS THE SUSPECTED CAUSE OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10053835 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page