LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-001213)Recalls & reported problems
Physio-Control, Inc. · 15AABAAAAAABBABAAAAAAAABLP15 MONITOR/DEFIB · Class 3
All models of LIFEPAK 15 monitor/defibrillator (93)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00883873912836), received between May 1, 2018 and Jun 14, 2024 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Failure to Deliver Energy” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Deliver Energy3
- Failure to Power Up2
- Defibrillation/Stimulation Problem1
- Electrical /Electronic Property Problem1
- Failure to Deliver Shock/Stimulation1
- Unexpected Shutdown1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Known Impact Or Consequence To Patient4
- No Patient Involvement1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 14, 2024Failure to Deliver Energy; Failure to Deliver Energy
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THERE WAS AN EVENT CODE LOGGED IN THEIR DEVICE'S MEMORY. THE EVENT CODE LOGGED IS RELATED TO A POTENTIAL FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19536914 ↗MalfunctionDec 5, 2023Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE HAS LOGGED AN EVENT IN THE MEMORY WHICH CAN BE RELATED TO A FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO PROVIDE DEFIBRILLATION THERAPY IF IT WERE NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18268040 ↗MalfunctionFeb 17, 2021Failure to Deliver Energy
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THERE WAS AN EVENT CODE LOGGED IN THE DEVICE'S MEMORY. THE EVENT CODE LOGGED IS INDICATIVE OF A FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. AS A RESULT, DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11339257 ↗MalfunctionFeb 15, 2021Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE'S POWER BUTTON WAS NOT FUNCTIONING. AS A RESULT, DEFIBRILLATION THERAPY MAY HAVE BEEN DELAYED OR UNAVAILABLE, IF NEEDED. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11327825 ↗MalfunctionDec 16, 2019Failure to Power Up; Failure to Power Up
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE "WOULD NOT POWER ON". THERE WAS NO PATIENT USE REPORTED WITH THE EVENT.
Patient problems: No Patient Involvement
MAUDE report 9476541 ↗MalfunctionJul 9, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE LOCKED UP AFTER SUCCESSFULLY DELIVERING FOR DEFIBRILLATION ENERGY TO A PATIENT. THE CUSTOMER ADVISED THAT AFTER THEIR DEVICE SUCCESSFULLY SHOCKED THE PATIENT, THEIR DEVICE LOCKED UP. THE USER HAD TO REMOVE THE BATTERIES IN ORDER TO TURN THE DEVICE OFF. AFTER RE-INSTALLING THE BATTERIES THEY WERE ABLE TO POWER THE DEVICE BACK ON. THE PATIENT WAS SUCCESSFULLY CONVERTED DURING THE DEFIBRILLATION SHOCK. THE PATIENT SURVIVED THE EVENT, HOWEVER THE CUSTOMER WAS UNABLE TO PROVIDE ANY FURTHER INFORMATION ON THE EVENT OR THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8774338 ↗MalfunctionDec 1, 2018Unexpected Shutdown; Unexpected Shutdown
THE CUSTOMER, A BIOMEDICAL ENGINEER, CONTACTED PHYSIO-CONTROL TO REPORT THAT DURING AN EVENT THEIR DEVICE POWERED OFF BY ITSELF MULTIPLE TIMES. THE CUSTOMER ADVISED THAT MEDICAL PERSONNEL WERE IN THE PROCESS OF IMPLANTING AND PROGRAMMING A PACEMAKER WHEN THE REPORTED ISSUE OCCURRED. MEDICAL PERSONNEL STATED THAT THE DEVICE WAS PLUGGED INTO AC POWER AT THE TIME AND HAD TWO FULLY CHARGED BATTERIES INSTALLED. THE DEVICE WAS BEING USED TO MONITOR THE PATIENT'S ECG WAVEFORM DURING THE PROCEDURE. THE CUSTOMER STATED THAT EACH TIME THE DEVICE LOST POWER, MEDICAL PERSONNEL WOULD POWER IT BACK ON, ONLY TO HAVE IT POWER OFF AGAIN. THEY BELIEVED THIS OCCURRED THREE TIMES DURING THE EVENT. NO FURTHER DETAILS WERE PROVIDED. PHYSIO-CONTROL HAS MADE MULTIPLE ATTEMPTS TO CONTACT THE CUSTOMER IN ORDER TO OBTAIN ADDITIONAL INFORMATION ABOUT BOTH THE PATIENT AND THE EVENT; HOWEVER, NO RESPONSE HAS BEEN RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8122349 ↗MalfunctionNov 29, 2018Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEY ATTEMPTED TO SHOCK A PATIENT AND THE DEVICE WOULD NOT SHOCK. THIS OCCURRED TWICE DURING THE EVENT. THE CUSTOMER OBTAINED A BACK-UP DEVICE AND REPORTED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE. THE CUSTOMER ADVISED PHYSIO-CONTROL THAT NO FURTHER PATIENT INFORMATION IS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8117132 ↗MalfunctionMay 1, 2018Failure to Deliver Energy; Failure to Deliver Energy
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WAS USED DURING A PATIENT EVENT AND WOULD NOT SHOCK WHEN THEY WERE ATTEMPTING TO PERFORM A SYNCHRONIZED CARDIOVERSION PROCEDURE THE DEVICE ALSO DISPLAYED THE SERVICE INDICATOR. THE USER POWERED THE DEVICE OFF AND BACK ON AND WAS THEN ABLE TO DELIVER A SHOCK TO THE PATIENT. THERE WERE NO REPORTS OF ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE. PHYSIO-CONTROL HAS MADE MULTIPLE ATTEMPTS TO CONTACT THE CUSTOMER IN ORDER TO OBTAIN ADDITIONAL INFORMATION ON THE PATIENT; HOWEVER, NO RESPONSE HAS BEEN RECEIVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7478675 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page