LIFEPAK® 12 defibrillator/monitor 97400 (REF 97400-001371)Recalls & reported problems

Physio-Control, Inc. · LP12AABCBBBACBBBAABAAAAALP12 MONITOR-DEFIB · Class 3

All models of LIFEPAK 12 defibrillator/monitor (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00883873950982), received between Aug 28, 2018 and Apr 8, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Failure to Power Up” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Power Up2
  • Connection Problem1
  • Electrical /Electronic Property Problem1
  • Unexpected Shutdown1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 8, 2024Failure to Power Up; Failure to Power Up

    A CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON. AS A RESULT, DEFIBRILLATION WOULD NOT BE AVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19057627 ↗
  • MalfunctionAug 15, 2019Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE UNEXPECTEDLY LOST POWER TWICE DURING PATIENT CARE. PHYSIO-CONTROL CONTACTED THE CUSTOMER TO REQUEST ADDITIONAL INFORMATION ON THE PATIENT. NO RESPONSE HAS BEEN RECEIVED FROM THE CUSTOMER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8896312 ↗
  • MalfunctionJun 19, 2019Failure to Power Up; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEY ATTEMPTED TO USE THIS DEVICE ON A PATIENT AND THE DEVICE WOULD NOT POWER ON. THE CUSTOMER SWITCHED TO A BACK UP DEVICE AFTER A 2 MINUTE DELAY. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8712653 ↗
  • MalfunctionNov 12, 2018Connection Problem; Connection Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT DURING AN EVENT, A SERVICE INDICATOR APPEARED. MEDICAL PERSONNEL WERE MONITORING A PATIENT AND THE UNIT CONTINUED TO OPERATE WITHOUT ANY DISCREPANCIES DURING THE EVENT. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE. UPON EVALUATION OF THE CUSTOMER'S DEVICE, PHYSIO-CONTROL OBSERVED THAT IT WOULD ONLY INTERMITTENTLY DETECT THAT A TEST PATIENT LOAD WAS CONNECTED VIA A THERAPY CABLE AND, AS A RESULT, DEFIBRILLATION WAS ONLY POSSIBLE INTERMITTENTLY. PHYSIO-CONTROL HAS ATTEMPTED TO CONTACT THE CUSTOMER IN ORDER TO OBTAIN ADDITIONAL DETAILS ABOUT THE PATIENT; HOWEVER, NO RESPONSE HAS BEEN RECEIVED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8062442 ↗
  • MalfunctionAug 28, 2018Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE DISPLAYED THE ¿SERVICE NEEDED¿ ACROSS THE DISPLAY DURING A PATIENT EVENT. THIS IS INDICATIVE OF A POTENTIAL DEVICE LOCK-UP. THE CUSTOMER ADVISED THAT DURING A TRANSPORT OF A PATIENT, AFTER HITTING A BUMP IN THE ROAD, THEIR DEVICE DISPLAY WENT BLANK AND THEN ¿SERVICE NEEDED¿ APPEARED ON THE SCREEN. WHEN THIS OCCURRED ALL LEDS BECAME ILLUMINATED AND REMAINED LIT. AFTER APPROXIMATELY 5 MINUTES THEY POWERED THE DEVICE OFF AND THEN BACK ON AGAIN, THEREAFTER NO FURTHER ISSUES WERE OBSERVED. THE CUSTOMER ADVISED THAT THERE WERE NO ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE. NO FURTHER DETAILS WERE PROVIDED. PHYSIO-CONTROL HAS MADE SEVERAL ATTEMPTS TO CONTACT THE CUSTOMER FOR FURTHER INFORMATION ON THE EVENT AND PATIENT, BUT HAS BEEN UNSUCCESSFUL.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7824772 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →

Sources and updates