Skip to content

LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-000051)Recalls & reported problems

Physio-Control, Inc. · 15AACBABBBBABAAAAAAAAAAALP15 MONITOR/DEFIB · Class 3

All models of LIFEPAK 15 monitor/defibrillator (93)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00883873987971), received between Jul 3, 2019 and Jun 9, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem2
  • Failure to Power Up1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 9, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, PHYSIO-CONTROL OBSERVED THAT THE DEVICE WOULD NOT COMPLETE THE BOOT UP CYCLE. AS A RESULT, DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF NEEDED.

    Patient problems: No Patient Involvement

    MAUDE report 10135847 ↗
  • MalfunctionJul 3, 2019Electrical /Electronic Property Problem; Failure to Power Up; Electrical /Electronic Property Problem; Failure to Power Up

    THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORTED THAT THEY WERE USING THIS DEVICE ON A PATIENT AND WERE HAVING DISPLAY ISSUES SO THEY ATTEMPTED TO RESTART THE DEVICE. WHEN THE DEVICE POWERED BACK ON IT LOCKED UP ON THE SELF-TEST SCREEN. AS A RESULT DEFIBRILLATION THERAPY MAY HAVE BEEN UNAVAILABLE, IF IT WAS NEEDED. THE CUSTOMER OBTAINED A BACK UP DEVICE AND USED THAT TO CONTINUE PATIENT CARE. THERE WERE NO REPORTS OF ANY ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 8757835 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page