Skip to content

LIFEPAK® 15 monitor/defibrillator 99577 (REF 99577-000049)Recalls & reported problems

Physio-Control, Inc. · 15AABBABBBBABAAAAAAAAAAALP15 MONITOR/DEFIB · Class 3

All models of LIFEPAK 15 monitor/defibrillator (93)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00883873987995), received between Dec 1, 2021 and Oct 5, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Failure to Deliver Shock/Stimulation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Failure to Deliver Shock/Stimulation1
  • Failure to Power Up1

Patient problems

  • Arrhythmia1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 5, 2022Failure to Power Up; Failure to Power Up

    STRYKER WAS PERFORMING PREVENTATIVE MAINTENANCE ON A CUSTOMER'S DEVICE AND OBSERVED THAT THE DEVICE IS NOT COMPLETING THE BOOT UP PROCESS. THE DEVICE IS LOCKING UP AT THE SELF TEST SCREEN. AS A RESULT, DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF NEEDED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15546403 ↗
  • MalfunctionDec 1, 2021Failure to Deliver Shock/Stimulation; Failure to Deliver Shock/Stimulation

    THE CUSTOMER CONTACTED STRYKER TO REPORT THAT THEIR DEVICE'S DISPLAY WENT BLANK WHILE TREATING A PATIENT. IN THIS STATE THE DEVICE MAY NOT BE ABLE TO DELIVER DEFIBRILLATION THERAPY IF NEEDED. THIS ISSUE IS PATIENT RELATED; HOWEVER THERE WAS NO ADVERSE EVENT REPORTED. THE CUSTOMER REPORTED THE USER WAS ABLE TO USE A BACKUP MONITOR WITHOUT FURTHER INCIDENT DURING THE PATIENT EVENT.

    Patient problems: Arrhythmia

    MAUDE report 12910843 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page