Skip to content

HeRO® 00884450310151Recalls & reported problems

Merit Medical Systems, Inc. · Class 2 · Implant

All models of HeRO (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00884450310151), received between Mar 19, 2020 and Nov 4, 2020 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Separation2
  • Detachment of Device or Device Component1
  • Migration1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 4, 2020Material Separation; Migration; Material Separation; Migration

    THE ACCOUNT ALLEGES THAT DURING A LONG-TERM HEMODIALYSIS DE-CLOTTING PROCEDURE, THE DISTAL OUTFLOW COMPONENTS RADIOPAQUE RING DETACHED WHILE DEPLOYING A 7MM X 40MM BALLOON WITHIN A CLOT LOCATED WITHIN THE HEMODIALYSIS OUTFLOW COMPONENT. THE ACCOUNT STATES THAT ALL BALLOON INFLATIONS WERE CLOSE TO THE OUTFLOW COMPONENT RADIOPAQUE MARKER BAND. POST BALLOON INFLATIONS, THE MARKER BAND DETACHED FROM THE COMPONENT BUT REMAINED ON THE GUIDEWIRE. A VASCULAR SNARE DEVICE WAS QUICKLY ADVANCED TO SUCCESSFULLY CAPTURE THE FOREIGN BODY FOR SAFE REMOVAL FROM THE PATIENT. NO PATIENT INJURY TO REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10785308 ↗
  • MalfunctionMar 19, 2020Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    THE ACCOUNT ALLEGES THAT DURING A THROMBECTOMY PROCEDURE USING A BALLOON TO CLEAR THE EXCESSIVE THROMBUS FROM THE VENOUS OUTPUT COMPONENT [VOC], THE DISTAL RING ON THE VOC DEVICE DETACHED AND MIGRATED TO THE PATIENT'S LUNG. A NEW VOC WAS PLACED FOR CONTINUED TREATMENT. THE PATIENT TOLERATED THE PROCEDURE WELL AND REMAINED ASYMPTOMATIC. NO ADDITIONAL PROCEDURES HAVE BEEN SCHEDULED.

    Patient problems: Foreign Body In Patient

    MAUDE report 9857530 ↗

Other PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices

All PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page