Skip to content

CorVocet™ 00884450476406Recalls & reported problems

Merit Medical Systems, Inc. · Class 2

All models of CorVocet (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00884450476406), received between Jun 29, 2021 and Jul 16, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 3, 2022Class IIOpen, Classified

    Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient needle stick injury.

    Z-0137-2023 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation, Positioning or Separation Problem1
  • Fluid/Blood Leak1

Patient problems

  • Exposure to Body Fluids1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 16, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THE ACCOUNT ALLEGES THAT WHEN DOING BIOPSIES WITH THE INTRODUCER, THEY NOTICED REFLUX OF BLOOD THROUGH THE HUB OF THE INTRODUCER NEEDLE. A PATIENT HAD FLUID SPATTER IN HIS EYES FOLLOWING BIOPSY OF A PELVIC MASS.

    Patient problems: Exposure to Body Fluids

    MAUDE report 22522718 ↗
  • MalfunctionJun 29, 2021Activation, Positioning or Separation Problem

    AS REPORTED BY STAFF, "BIOPSY DEVICE APPEARED TO FIRE BUT ONLY BLOOD WAS ON THE NEEDLE INSTEAD OF A 20 MM TISSUE SPECIMEN." ASSOCIATE REPORTING INDICATED TWO DEVICES FAILED BOTH WITH THE SAME LOT NUMBER (BEING REPORTED) REQUIRING A THIRD DEVICE TO BE OPENED AND USED TO COMPLETE THE PROCEDURE (DEVICE LOT NUMBER THAT WAS SUCCESSFULLY USED IS UNKNOWN).

    Patient problems: Insufficient Information

    MAUDE report 12083492 ↗

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page