CorVocet recallZ-0137-2023 · Merit Medical Systems, Inc. · initiated Aug 3, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient needle stick injury.
FDA root cause category: Process change control
What the company is doing
On 8/3/2022, recall notices were mailed to customers who were asked to do the following: 1) Identify affected devices, quarantine, discontinue use and distribution, and return to the recalling firm. 2) If product was further distributed, share the recall notice with these customers/users. 3) Complete and return the customer response form. Customers with questions can contact Customer Service at 801-208-4381, 6AM-6PM, Mon-Fri.
Recall details
- Recall number
- Z-0137-2023
- Status
- Open, Classified
- Initiated by the company
- Aug 3, 2022
- Posted by the FDA
- Oct 27, 2022
- Recalling firm
- Merit Medical Systems, Inc.
- Quantity
- 1822
- Product code
- KNW
- FDA submissions
- K180450
- Distribution
- Worldwide distribution - US National distribution including in the states of FL, IL, OH, PA, CA, WI, VA, AZ, AR, MS, TN, MO, MA, MI, NY, NE, TX, WA, MN, LA, DE, IN, AL, NJ, OR, MD, GA, CT, OK, UT, KS, NC, CO, KY, WV, AK, NV, RI, SC, ND, NH, MT and the countries of Finland, Denmark, United Arab Emirates, Spain, Sweden, Norway, Jordan, Australia, Germany, Austria, Italy, United Kingdom, France, Canada, Mexico, Puerto Rico, Chile, Reunion, Belgium, Switzerland, Japan, South Africa, Qatar, Russian Federation, Hungary, Thailand, Namibia, Colombia, New Zealand, Netherlands, Saudi Arabia, Costa Rica, Oman, Cyprus, Slovenia, Croatia, China.
Product description
CorVocet Biopsy System/Catalog Code: CORA1610/B, CORA1610S/B, CORA1615/B, CORA1810/A, CORA1810/B, CORA1810S/B, CORA1815/A, CORA1815/B, CORA1815S/A, CORA1815S/B, CORA1820/A, CORA1820/B, CORA1820S/A, CORA1825/A, CORA1825/B, CORA1825S/A, CORA2010/A, CORA2010/B, CORA2010S/A, CORA2015/A, CORA2015/B, CORA2015S/A, CORA2020/A, CORA2020/B, CORCA1410/A, CORCA1410/B, CORCA1415/A, CORCA1415/B, CORCA1610/A, CORCA1610/B, CORCA1610S/B, CORCA1615/A, CORCA1615/B, CORCA1615S/B, CORCA1810B/A, CORCA1810B/B, CORCA1810SB/A, CORCA1810SB/B, CORCA1815B/A, CORCA1815B/B, CORCA1815SB/A, CORCA1815SB/B, CORCA1820B/A, CORCA1820B/B, CORCA1820SB/A, CORCA1820SB/B, CORCA1825B/A, CORCA1825SB/B, CORCA2010B/A, CORCA2010B/B, CORCA2010SB/A, CORCA2010SB/B, CORCA2015B/A, CORCA2015B/B, CORCA2015SB/A, CORCA2015SB/B, CORCA2020B/A, CORCA2020B/B
Code information (models, lots, serial numbers)
Show the full code information (7k characters)
Catalog Code-Lot/UDI: CORA1610/B-I2075805, I2093969, I2101330, I2124107, I2233501, I2286703, I2325899/00884450476192; CORA1610S/B-I2138037, I2352845/00884450476222; CORA1615/B-H1408433, H1462686, I1863888, I2127448, I2170247, I2259080/00884450476253; CORA1810/A-H1715532, H1756745, H1767115, H1771762, H1776067, H1784861, H1800979, H1808121, H1825693/00884450316405; CORA1810/B-I1781878, I1818817, I1837234, I1843704, I1855418, I1866359, I1952354, I1958525, I1971876, I1990075, I1998862, I2043630, I2070249, I2075810, I2093993, I2114550, I2116110, I2124108, I2127451, I2137798, I2155289, I2206046, I2243065, I2263876, I2263877, I2282032, I2291981, I2297372, I2385808/00884450476116; CORA1810S/B-I1877912, I1877916, I1902603, I1971867, I1998851, I2033178, I2070341, I2075809, I2124109, I2127455, I2170248, I2370120/00884450477625; CORA1815/A-H1733631, H1734003, H1739133, H1756746, H1771770, H1795481, H1800980, H1807701/00884450316412; CORA1815/B-I1821113, I1838444, I1855417, I1920169, I1920191, I1946564, I1952362, I1989173, I1990074, I2033176, I2070349, I2075808, I2093995, I2114555, I2116111, I2124110, I2127456, I2137805, I2170249, I2201366, I2233502, I2263885, I2269827, I2325901, I2349440/00884450477618; CORA1815S/A-H1756747, H1784863/00884450316528; CORA1815S/B-I1869477, I1874063, I2075806, I2114557, I2127463, I2325902/00884450477601; CORA1820/A-H1375632, H1382959, H1718587, H1734004, H1739135, H1756748, H1762731, H1771773, H1771881, H1784864, H1815597/00884450316429; CORA1820/B-I1823060, I1971866, I2114564, I2137809, I2170250, I2303053, I2385809/00884450477595; CORA1820S/A-H1739136, H1790841/00884450316535; CORA1825/A-H1749790, H1762724, H1808022, H1815598/00884450316436; CORA1825/B-I1843706, I1958504, I2234738, I2239042/00884450477571; CORA1825S/A-H1790845/00884450316542; CORA2010/A-H1784872, H1815600/00884450316443; CORA2010/B-I1863885, I2201378, I2206047, I2237417/00884450477557; CORA2010S/A-H1692834, H1776070/00884450316559; CORA2015/A-H1795483/00884450316450; CORA2015/B-I1823512, I1860367, I1962715, I2140223, I2243072, I2283710, I2303054, I2330668/00884450477526; CORA2015S/A-H1697709, H1771823, H1784884, H1795484/00884450316566; CORA2020/A-H1815606/00884450316467; CORA2020/B-I1781880, I1971845, I2007291, I2228365/00884450477496; CORCA1410/A-H1639707/00884450316580; CORCA1410/B-I2244480, I2384730/00884450476284; CORCA1415/A-, H1776075/00884450316597; CORCA1415/B-I2191846, I2286700, I2318838, I2325906/00884450476314; CORCA1610/A-H1428538, H1714401/00884450316603; CORCA1610/B-I1971855, I1998714, I2114620, I2170251, I2201382, I2233505, I2318858, I2343154/00884450476338; CORCA1610S/B-I1874064, I2010416/00884450476345; CORCA1615/A-H1703701/00884450316610; CORCA1615/B-I1845903, I1888547, I1920175, I1962716, I1998860, I2103685, I2137221, I2137223, I2170252, I2237426, I2286696, I2297374, I2303056, I2330680, I2385810/00884450476352; CORCA1615S/B-I2114626, I2246748/00884450476369; CORCA1810B/A-H1718595, H1734012, H1749798, H1756750, H1771846, H1776080, H1782158, H1790850, H1800985, H1808042/00884450316627; CORCA1810B/B-I1814729, I1838323, I1843705, I1845904, I1869474, I1877911, I1888548, I1902611, I1920161, I1946561, I1946566, I1958530, I1962714, I1990098, I1990104, I1998856, I2025072, I2070266, I2070374, I2093853, I2101369, I2104884, I2114628, I2114714, I2114715, I2116109, I2124195, I2127511, I2150947, I2150948, I2150949, I2159264, I2165513, I2165514, I2201386, I2221444, I2286695, I2303061, I2330685, I2343157, I2385811/00884450476376; CORCA1810SB/A-H1733634, H1800986/00884450316801; CORCA1810SB/B-I1862927, I1920180, I2033140, I2114634, I2233507, I2303065, I2330686/00884450476383; CORCA1815B/A-H1476238, H1608916, H1714408, H1734014, H1734016, H1743880, H1749800, H1749802, H1762717, H1771848, H1776082, H1782159, H1790847, H1795487, H1807720/00884450316634; CORCA1815B/B-I1827256, I1838326, I1843709, I1845905, I1860329, I1860365, I1877914, I1888549, I1902608, I1902609, I1920165, I1920186, I1946560, I1946567, I1952368, I1958533, I1962702, I1971860, I1971861, I1990097, I1990103, I2007293, I2058398, I2061959, I2062021, I2065393, I2075813, I2075839, I2094063, I2098053, I2102664, I2105077, I2114651, I2114652, I2124197, I2127514, I2127515, I2150950, I2150952, I2159275, I2159276, I2170253, I2201393, I2217616, I2281931, I2318869, I2330533, I2330692, I2336279, I2343163, I2349439, I2357134, I2385813, I2385814/00884450476390; CORCA1815SB/A-H1714409, H1743881, H1756752, H1762715/00884450316740; CORCA1815SB/B-I1838328, I1845906, I1877918, I1958536, I2061956, I2075811, I2093850, I2094072, I2233509, I2325909, I2330694, I2399997/00884450476406; CORCA1820B/A-H1718598, H1733635, H1734019, H1749803, H1756754, H1762714, H1771883, H1776084, H1790851, H1807721/00884450316641; CORCA1820B/B-I1827257, I1845907, I1902605, I1902607, I1920166, I1946562, I1971929, I1990094, I1998850, I1998858, I2007294, I2094075, I2103689, I2114708, I2124200, I2127516, I2150980, I2159282, I2159283, I2233510, I2286679, I2343166, I2352856/00884450476413; CORCA1820SB/A-H1733637/00884450316757; CORCA1820SB/B-I2137856, I2325911/00884450476437; CORCA1825B/A-H1718603/00884450316658; CORCA1825B/B-I1946563, I2137863, I2201410, I2265757, I2325913/00884450476451; CORCA1825SB/B-I2156607/00884450476468; CORCA2010B/A-H1445217, H1703705, H1708690, H1718606, H1718607, H1749806, H1762738, H1771884, H1807725, H1815614/00884450316665; CORCA2010B/B-I18210937/I1845908, I1862929, I1874055, I1877913, I1958538, I1971927, I1990070, I2075817, I2094080, I2114709, I2127519, I2170256, I2217622, I2325922/00884450476475; CORCA2010SB/A-H1749807/00884450316771; CORCA2010SB/B-I1874058, I1902604, I1998852, I2122957, I2233511/00884450477632; CORCA2015B/A-H1624841, H1639711, H1657959, H1697713, H1708691, H1714416, H1718609, H1733639, H1734020, H1743882, H1749810, H1771851, H1776086, H1784915, H1790853, H1807727/00884450316672; CORCA2015B/B-I1827258, I1855420, I1862900, I1873128, I1874060, I1888545, I1902614, I1920156, I1952374, I1990072, I1998863, I2070355, I2080473, I2094083, I2114711, I2116112, I2124201, I2127521, I2137868, I2159284, I2191857, I2246760, I2263980, I2287299, I2303068, I2352880, I2370145/00884450477649; CORCA2015SB/A-H1662694, H1733642, H1739148, H1771853, H1776087, H1807729, H1815615/00884450316788; CORCA2015SB/B-I1827270, I1845912, I1862899, I2080472, I2170257, I2201414/00884450477656; CORCA2020B/A-H1662697, H1692861, H1708693, H1718613, H1756755, H1771854, H1776088, H1790855, H1807731/00884450316689; CORCA2020B/B, I1827272, I1845913, I1869475, I1874065, I1877915, I1971882, I1990071, I2070361, I2075814, I2114712, I2151307, I2170259, I2201416, I2330716/00884450477663; CORCA2020SB/B-I2263985, I2297377/00884450477670;
Devices named in this recall
60 devices in the FDA device registry whose identifier appears in the recall's code information.
- CorVocet™ 00884450476390Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall8 reports
- CorVocet™ 00884450316672Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall3 reports
- CorVocet™ 00884450476475Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall3 reports
- CorVocet™ 00884450316412Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall2 reports
- CorVocet™ 00884450476352Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall2 reports
- CorVocet™ 00884450476406Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall2 reports
- CorVocet™ 00884450316634Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall1 report
- CorVocet™ 00884450316641Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall1 report
- CorVocet™ 00884450476383Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall1 report
- CorVocet™ 00884450477656Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall1 report
- CorVocet™ 00884450477663Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall1 report
- CorVocet™ 00884450316405Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316429Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316436Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316443Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316450Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316467Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316528Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316535Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316542Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316559Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316566Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316580Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316597Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316603Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316610Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316627Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316658Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316665Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316689Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316740Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316757Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316771Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316788Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450316801Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476116Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476192Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476222Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476253Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476284Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476314Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476338Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476345Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476369Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476376Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476413Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476437Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476451Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450476468Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450477496Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450477526Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450477557Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450477571Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450477595Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450477601Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450477618Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450477625Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450477632Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450477649Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
- CorVocet™ 00884450477670Merit Medical Systems, Inc. · INSTRUMENT, BIOPSY1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page