Skip to content

CorVocetRecalls & reported problems across all models

Merit Medical Systems, Inc. · 62 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one CorVocet model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
27
Deaths reported
0
Injuries reported
9
Recalls
1

We found 62 models of CorVocet by Merit Medical Systems, Inc. in the FDA's device registry (GUDID). Together they are named in 27 adverse event reports in the FDA's MAUDE database, received between Jun 22, 2020 and Sep 29, 2026 — classified as 9 injuries, 18 malfunctions.

The most frequently reported specific device problem is “Fracture” (10 reports).

1 FDA recall record name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 4 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

CorVocet recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 3, 2022Open, Classified

    Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient needle stick injury.

    Z-0137-2023 · Root cause: Process change control · names 60 of these models · Recall details · FDA record ↗

Most reported CorVocet problems

Device problems

  • Fracture10
  • Material Separation6
  • Break4
  • Failure to Fire3
  • Fluid/Blood Leak2
  • Material Fragmentation2
  • Activation, Positioning or Separation Problem1
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions17
  • No Consequences Or Impact To Patient4
  • Insufficient Information2
  • Device Embedded In Tissue or Plaque1
  • Exposure to Body Fluids1
  • Hemoptysis1
  • Hemorrhage/Blood Loss/Bleeding1
  • Needle Stick/Puncture1

CorVocet models

62 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent CorVocet adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Merit Medical Systems, Inc. products

All Merit Medical Systems, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page