CorVocet™ 00884450476475Recalls & reported problems

Merit Medical Systems, Inc. · Class 2

All models of CorVocet (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00884450476475), received between Nov 8, 2021 and Jan 20, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 3, 2022Class IIOpen, Classified

    Biopsy Systems have components that are misaligned due to an issue with the manufacturing assembly equipment, which may lead to the needle ejecting when triggered outside the body, which may result in user or patient needle stick injury.

    Z-0137-2023 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Failure to Fire1
  • Fracture1
  • Material Fragmentation1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 20, 2026Break; Fracture; Material Fragmentation; Material Separation

    THE ACCOUNT ALLEGES THAT THE TOP PART OF THE BIOPSY NEEDLE IS LOOSE. NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24119186 ↗
  • MalfunctionJan 12, 2026

    THE TOP PART OF THE BIOPSY NEEDLE IS LOOSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24032019 ↗
  • MalfunctionNov 8, 2021Failure to Fire

    PATIENT WITH HISTORY OF RIGHT LUNG NODULE WAS ADMITTED FOR A CT SCAN-GUIDED BIOPSY OF RIGHT LUNG. THE NEEDLE OF THE CORVOCET BIOPSY SYSTEM WAS CHECKED PRIOR TO USING THE DEVICE. THE NEEDLE DID NOT FIRE AS IT SHOULD HAVE. NEEDLE WAS NOT USED ON PATIENT. A NEW NEEDLE WAS USED FOR THE CASE. NO KNOWN HARM TO PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 12770132 ↗

Other INSTRUMENT, BIOPSY devices

All INSTRUMENT, BIOPSY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page