HeartStart MRx Lithium Ion Battery ModuleRecalls & reported problems

Philips Medical Systems Hsg · Lithium Ion Battery Module · Class 3

All models of HeartStart (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00884838002791), received between Oct 14, 2019 and Oct 4, 2021 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Battery Problem” (12 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Nov 23, 2018Class IITerminated

    Affected lithium-ion batteries may contain a defective component (Thermal Cut-off or TCO). Should the component fail, the battery will no longer charge or deliver power to the Monitor/Defibrillator, which may prevent operation of the device if it is not connected to AC or DC power. Should a second, unaffected battery be present in the Monitor/Defibrillator, the battery failure may go unnoticed.

    Z-0603-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Battery Problem12
  • Failure of Device to Self-Test1
  • Failure to Charge1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • No Patient Involvement1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 4, 2021Battery Problem; Battery Problem

    IT WAS REPORTED TO PHILIPS THAT THE BATTERY FAILS TO CHARGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12569038 ↗
  • MalfunctionOct 4, 2021Battery Problem; Battery Problem

    IT WAS REPORTED TO PHILIPS THAT THE BATTERY FAILS TO CHARGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12568805 ↗
  • MalfunctionAug 31, 2021Battery Problem; Battery Problem

    IT WAS REPORTED TO PHILIPS THAT THE BATTERY MALFUNCTIONED. THE CUSTOMER WORKED WITH THE TECHNICAL SUPPORT REPRESENTATIVE. THE CUSTOMER CONFIRMED THE NEED FOR A BATTERY EXCHANGE. THE REPORTED PROBLEM WAS VERIFIED IN THE LAB. THE BATTERY PACK WAS UNABLE TO CHARGE EITHER IN HEARTSTART MRX DEVICE OR A CADEX CHARGER. TROUBLESHOOTING CONDUCTED REVEALED FUSE F1 OPEN. THE FUSE WAS REPLACED AND THE BATTERY OPERATED AS NORMAL. THE CUSTOMER WAS SENT A BATTERY REPLACEMENT. THE BATTERY IS CUSTOMER REPLACEABLE. NO FURTHER ACTION AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12390417 ↗
  • MalfunctionJul 8, 2021Battery Problem; Failure of Device to Self-Test; Battery Problem; Failure of Device to Self-Test

    IT WAS REPORTED TO PHILIPS THAT THE RFU INDICATOR DISPLAYED A RED X AND THE ALARM SOUNDED. THERE WAS NO REPORTED PATIENT INVOLVEMENT,

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12138031 ↗
  • MalfunctionJun 28, 2021Battery Problem; Battery Problem

    IT WAS REPORTED TO PHILIPS THAT THE BATTERY FAILS TO CALIBRATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12081660 ↗
  • MalfunctionJun 28, 2021Battery Problem; Battery Problem

    IT WAS REPORTED TO PHILIPS THAT THE BATTERY FAILS TO CALIBRATE. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12080975 ↗
  • MalfunctionJun 16, 2021Battery Problem; Battery Problem

    IT WAS REPORTED TO PHILIPS THAT THE BATTERY FAILS TO CHARGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12010519 ↗
  • MalfunctionJun 16, 2021Battery Problem; Battery Problem

    IT WAS REPORTED TO PHILIPS THAT THE BATTERY FAILS TO CHARGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12006274 ↗
  • MalfunctionJun 16, 2021Failure to Charge; Battery Problem; Failure to Charge; Battery Problem

    IT WAS REPORTED TO PHILIPS THAT THE BATTERY FAILS TO CHARGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12006270 ↗
  • MalfunctionMay 22, 2021Battery Problem; Battery Problem

    IT WAS REPORTED TO PHILIPS THAT THE DEVICE BATTERY IS NOT RECOGNIZED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11868434 ↗
  • MalfunctionMay 21, 2021Battery Problem; Battery Problem

    IT WAS REPORTED TO PHILIPS THAT THE BATTERY FAILS CALIBRATION BY ONLY BEING ABLE TO CHARGE TO 53%. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11864499 ↗
  • MalfunctionOct 14, 2019Battery Problem

    IT WAS REPORTED TO PHILIPS THAT THE BATTERY IS DEFECTIVE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 9186578 ↗

Other Automated external defibrillators (non-wearable) devices

All Automated external defibrillators (non-wearable) devices →