HeartStart Pads 989803125411Recalls & reported problems
Philips Medical Systems Hsg · HS Pedi Radiotransparent Pads · Class 3
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00884838002883), received between Nov 24, 2021 and Mar 8, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Electrical /Electronic Property Problem2
- Degraded1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information2
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 8, 2023Degraded; Degraded
THE PADS CONNECTOR PROJECT CHANGED THE ALS PADS CONNECTOR ASSEMBLY SUPPLIER (TIER 2) FOR PART NUMBER M3719A. AS PART OF THE PROJECT, THE HAZARDOUS SUBSTANCE COMPLIANCE EVIDENCE WAS UPDATED. ON (B)(6) 2023, IT WAS DISCOVERED THAT THE NEW SUPPLIER'S WIRING INSULATION OF THE CONNECTOR ASSEMBLIES USED TO MANUFACTURE M3718A AND M3719A PADS WIRING INSULATION CONTAINS A PVC COMPOUND MATERIAL COMPRISED OF 20-30% DEHP MATERIAL. THIS MATERIAL HAS BEEN USED SINCE INITIAL PRODUCTION RELEASE. THIS LEVEL OF DEHP MATERIAL DOES NOT COMPLY WITH THE FOLLOWING REGULATIONS: ROHS2; ROHS3; PROPOSITION 65; REACH. PHTHALATES HAVE BEEN IN THE M3719A WIRE INSULATION MATERIAL SINCE INITIAL PRODUCTION RELEASE. THERE WAS NO PATIENT INVOLVEMENT. PHILIPS IS IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION CONCERNING THIS EVENT AND THE COMPLAINT IS STILL UNDER INVESTIGATION. A FINAL REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16507793 ↗MalfunctionNov 29, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED TO PHILIPS THAT, WHEN THE CUSTOMER TRIED TO USE THE X-RAY TRANSMISSION TYPE PEDIATRIC DEFIBRILLATION PAD M3719A, THE GEL PEELED OFF FROM THE PAD SIDE AND REMAINED ON THE MOUNT. THE NEXT M3719A PREPARED IN SUCCESSION WAS IN THE SAME STATE. THE FIRST PAD IS BEING ADDRESSED IN (B)(4). THE SECOND PAD IS ADDRESSED HERE IN (B)(4). THE PADS WERE BEING USED TO MONITOR A PATIENT. THE DEVICE WAS IN USE ON A PATIENT AND THERE WAS NO ADVERSE EVENT TO PATIENT OR USER. THE INFANT RADIOTRANSPARENT MULTIFUNCTION ELECTRODE PADS WERE RETURNED FOR FAILURE ANALYSIS. THE INFANT RADIOTRANSPARENT MULTIFUNCTION ELECTRODE PADS WERE RECEIVED WITH SIGNIFICANT UNUNIFORM GEL SURFACE AND SIGNS OF AIR/BUBBLES BETWEEN THE GEL AND THE CLEAR PLASTIC LINER. THE REPORTED PROBLEM WAS VERIFIED IN THE LAB. THE STERNUM PAD PRESENTED SIGNIFICANT UNUNIFORM GEL SURFACE AND THE APEX PAD SHOWED SIGNS OF AIR/BUBBLES BETWEEN THE GEL AND THE CLEAR PLASTIC LINER RESULTING IN PADS FAILING TESTS. THE REPORTED ISSUE WAS CONFIRMED. THE INFANT RADIOTRANSPARENT MULTIFUNCTION ELECTRODE PADS WERE SCRAPPED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12885868 ↗MalfunctionNov 24, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED TO PHILIPS THAT, WHEN THE CUSTOMER TRIED TO USE THE X-RAY TRANSMISSION TYPE PEDIATRIC DEFIBRILLATION PAD M3719A, THE GEL PEELED OFF FROM THE PAD SIDE AND REMAINED ON THE MOUNT. THE NEXT M3719A PREPARED IN SUCCESSION WAS IN THE SAME STATE. THE FIRST PAD IS BEING ADDRESSED IN PR (B)(4). IT WAS INITIALLY REPORTED THAT THERE MAY HAVE BEEN A DELAY IN LIFE SAVING THERAPY, HOWEVER, IT WAS LATER CLARIFIED BY EMAIL FROM TECHNICAL SUPPORT THAT THE CUSTOMER WAS MONITORING A PATIENT AT THE TIME OF THE ISSUE. THEREFORE, THIS REPORT HAS BEEN UPDATED FROM A SERIOUS INJURY TO A PRODUCT PROBLEM. THE SECOND PAD IS ADDRESSED IN PR (B)(4). THE CUSTOMER CONTACTED THE PHILIPS RESPONSE CENTER. THE CUSTOMER SHIPPED THE INVOLVED PAD SET TO PHILIPS FOR EVALUATION. ONE DEFIBRILLATION PAD SET WAS RETURNED FOR FAILURE ANALYSIS. THE REPORTED PROBLEM WAS VERIFIED. THE STERNUM PADS PRESENTED SIGNIFICANT UN-UNIFORM GEL SURFACE AND THE APEX PADS SHOWED SIGNS OF AIR/BUBBLES BETWEEN THE GEL AND THE CLEAR PLASTIC LINER RESULTING IN PADS FAILING TESTS. THE REPORTED PROBLEM WAS VERIFIED. THE PAD SET WILL BE SCRAPPED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12871559 ↗
Other Automated External Defibrillators (Non-Wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page