Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00884838031746), received between Jan 24, 2018 and May 30, 2024 — classified as 3 deaths, 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Defective Alarm” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Alarm2
  • Delayed Alarm2
  • Adverse Event Without Identified Device or Use Problem1
  • Communication or Transmission Problem1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Asystole2
  • Death2
  • Cardiac Arrest1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 30, 2024Defective Alarm; Communication or Transmission Problem; Defective Alarm; Communication or Transmission Problem

    THE CUSTOMER REPORTED THAT THERE WAS NO RED ALARM ON CENTRAL MONITORING. IT IS UNKNOWN IF THE DEVICE WAS IN USE AT TIME OF EVENT, AND THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19425278 ↗
  • InjurySep 12, 2022Delayed Alarm; Delayed Alarm

    IT WAS REPORTED THAT THERE WAS A DELAY IN RECEIVING THE ALERT FOR AN ASYSTOLE ALARM FROM THE INTELLISPACE EVENT MANAGEMENT (IEM) SYSTEM AND, AS A RESULT, THE PATIENT HAD EXPERIENCED AN ADVERSE EVENT AND IT WAS NECESSARY TO TRANSFER THE PATIENT TO THE INTENSIVE CARE UNIT (ICU).

    Patient problems: Asystole

    MAUDE report 15403156 ↗
  • DeathFeb 24, 2022Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    HOLD SMP 2.24IT WAS REPORTED THAT ALARMS WERE NOT PAGED FOR BED 09B ON (B)(6) 2022 BETWEEN 06:45 AND 19:00 RESULTING IN A PATIENT DEATH.

    Patient problems: Asystole

    MAUDE report 13591204 ↗
  • DeathSep 26, 2019Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED THAT THEY NEED ACTIVITY LOG FOR (B)(6) 2019 AT 10:06AM FOR BED RM 334 TELE-V-FIB/TACH, DUE TO A RECENT PAGING ISSUE RELATED TO TELEMETRY ALERTS WITH THE EMERGIN (IEM) PAGING SYSTEM (PART 865206). THE SITE RELIES ON THE PHILIPS IEM PAGING AND VOCERA PHONE SYSTEM FOR PRIMARY ALERTS. THE PATIENT HAD A CARDIAC ARREST AND DIED.

    Patient problems: Cardiac Arrest; Death

    MAUDE report 9125090 ↗
  • InjuryJun 11, 2019Delayed Alarm

    THE CUSTOMER REPORTED THAT ROOM 306 ALERTED A VTACH ON (B)(6) 2019 AT 9:47:03. THERE WAS A DELAY OF ALARMS VIA THE PAGING SYSTEM. NO INFORMATION WAS RECEIVED INDICATING THE DEVICE WAS NOT A CONTRIBUTING FACTOR IN THE EVENT. THE TYPE OF ADVERSE EVENT WAS NOT PROVIDED AT THE TIME OF THIS REPORT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8686332 ↗
  • DeathJan 24, 2018Defective Alarm

    THE CUSTOMER REPORTED THAT ON (B)(6) 2018, AS ALARMING STOPPED THROUGH THE PAGING (IEM) SYSTEM AT 10:17AM. THE PATIENT LEADS OFF OF TELEMETRY STARTING HAPPENING AT 17:30PM, AND NO IEM LEADS OFF ALERTS WERE RECEIVED. AT 19:00, THEY FOUND THE PATIENT IN ROOM 335 IN 3 SOUTH EXPIRED. THE PATIENT DIED.

    Patient problems: Death

    MAUDE report 7215509 ↗

Other SYSTEM,NETWORK AND COMMUNICATION,PHYSIOLOGICAL MONITORS devices

All SYSTEM,NETWORK AND COMMUNICATION,PHYSIOLOGICAL MONITORS devices →