BV PulseraRecalls & reported problems

Philips Medical Systems Nederland B.V. · Mobile, diagnostic X-ray imaging and viewing system. Used for radiological guidance and visualization during diagnostic, interventional, and surgical procedures. · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00884838055506), received between Sep 8, 2022 and Jan 6, 2025 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Smoking” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 12, 2021Open, Classified

    Instructions for Use do not specify the maximum surface temperature of the X-ray tank (assembly housing the X-ray tube, beam filter & cooling oil) and the image intensifier/detector of the system as required an result in burns

    Z-0356-2022 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Smoking4
  • Positioning Failure2
  • Radiation Output Failure2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Handling Problem1
  • Difficult or Delayed Positioning1
  • Unexpected Shutdown1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Insufficient Information1
  • Laceration(s)1
  • Skin Tears1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 6, 2025Smoking; Smoking

    IT HAS BEEN REPORTED TO PHILIPS THAT THE LEFT LCD MONITOR STARTED TO SMOKE DURING A CASE. THE CUSTOMER UNPLUGGED THE SYSTEM AND ABORTED THE PROCEDURE. THERE WAS NO REPORTED PATIENT OR USER HARM. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21081911 ↗
  • MalfunctionOct 16, 2024Difficult or Delayed Positioning; Difficult or Delayed Positioning

    IT HAS BEEN REPORTED TO PHILIPS THAT DURING A PROCEDURE IN (B)(6) 2024, THE C-ARM WAS BEING ROTATED IN THE OBLIQUE DIRECTION BY THE USER DURING A SURGERY. THEY STATED THAT THE C-ARM WAS MORE DIFFICULT THAN USUAL TO PUSH. THEY BENT DOWN TO USE THEIR LEGS AND FELT A PULL IN THEIR RIGHT-SIDE FLANK AND INTO THE LOWER BACK. THE USER WAS ADVISED TO TAKE IBUPROFEN AND REST. THE USER REPORTED THAT IN (B)(6) 2024 THEY REINJURED THEIR BACK IN THE SAME WAY. THEY WERE ADVISED BY TWO DIFFERENT NURSE PRACTITIONERS TO REST AND USE THE MEDICATIONS PROVIDED ALONG WITH ICE AND HEAT. IT WAS RECOMMENDED TO HAVE THE INJURY FURTHER EVALUATED BY A PHYSICIAN. THE USER IS CURRENTLY UNABLE TO WORK AND WAITING FOR PHYSICAL THERAPY, AN MRI, AND X-RAYS TO BE APPROVED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20460105 ↗
  • MalfunctionAug 13, 2024Smoking; Smoking

    IT HAS BEEN REPORTED TO PHILIPS THAT THE LEFT MONITOR MADE A POPPING SOUND AND SMOKE CAME OUT OF IT BEFORE TURNING OFF. THERE WAS NO REPORTED PATIENT OR USER HARM. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19977432 ↗
  • MalfunctionDec 5, 2023Radiation Output Failure; Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE DEVICE'S WIRED FOOTSWITCH WAS UNABLE TO TRIGGER FLUOROSCOPY. THE DEVICE WAS IN CLINICAL USE. NO PATIENT OR USER HARM WAS REPORTED. A PHILIPS FIELD SERVICE ENGINEER (FSE) INSPECTED THE SYSTEM ONSITE AND REPLACED THE FOOTSWITCH WHICH RESOLVED THE ISSUE. THE DEVICE WAS RETURNED TO USE IN GOOD WORKING ORDER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18265905 ↗
  • MalfunctionAug 4, 2023Positioning Failure; Adverse Event Without Identified Device or Use Problem; Positioning Failure; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO PHILIPS THAT THE USER LOWERED THE TABLE OF A BV PULSERA SYSTEM WITHOUT THE C-ARM BEING FULLY REMOVED FROM UNDERNEATH THE TABLE. THIS CAUSED THE C-ARM TO BE BUMPED AND TIP OVER. AS A RESULT, THE TECHNOLOGIST¿S LEFT ANKLE WAS REPORTED TO BE DEGLOVED AND REQUIRE 22 STICHES TO REPAIR THE SKIN. STICHES WERE REMOVED AFTER APPROXIMATELY ONE WEEK AND THERE WAS NO LASTING EFFECT OR IMPACT TO THE TECHNOLOGIST¿S HEALTH. FURTHERMORE, IT WAS CONFIRMED THAT THERE WAS NO IMPACT TO THE PATIENT. THE PROCEDURE HAD ALREADY BEEN COMPLETED WHEN THE INCIDENT OCCURRED AND HOSPITAL STAFF WERE TAKING THE PATIENT OFF OF THE TABLE. A PHILIPS FIELD SERVICE ENGINEER (FSE) INSPECTED THE SYSTEM ONSITE AND REPAIRED THE STEERING SYSTEM THAT SLID DUE TO THE C-ARM PHYSICALLY TIPPING OVER. THE SYSTEM WAS RETURNED TO USE IN GOOD WORKING ORDER.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 17459815 ↗
  • MalfunctionMar 15, 2023Radiation Output Failure; Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE DEVICE COULD NOT TAKE LIVE IMAGES. THE DEVICE WAS IN CLINICAL USE. PHILIPS HAS STARTED AN INVESTIGATION OF THE COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16543182 ↗
  • InjuryDec 29, 2022Positioning Failure; Device Handling Problem; Positioning Failure; Device Handling Problem

    IT HAS BEEN REPORTED TO PHILIPS THAT THE RADIOGRAPHER GOT FINGER INJURY WHILE USING THE LOCKS ON THE C-ARM. INTERVENTION IS UNKNOWN. PHILIPS HAS STARTED AN INVESTIGATION OF THE COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16068697 ↗
  • MalfunctionOct 10, 2022Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED TO PHILIPS THAT THE DEVICE SWITCHED OFF DURING A PROCEDURE AND DISPLAYED "GENERATOR ERROR". IT IS UNKNOWN IF THE PROCEDURE WAS ABLE TO BE COMPLETED. NO PATIENT HARM WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION FOR THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15569064 ↗
  • MalfunctionSep 14, 2022Smoking; Smoking

    IT HAS BEEN REPORTED TO PHILIPS THAT SMOKE AND SPARKS APPEARED FROM THE SYSTEM IN THE CART AND POWER CABLE AREA. THE SYSTEM WAS NOT IN CLINICAL USE AT THE TIME THE ISSUE WAS OBSERVED. THE SYSTEM WAS TURNED OFF AND DISCONNECTED FROM THE POWER SUPPLY OF THE HOSPITAL. NO HARM HAS BEEN REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15418625 ↗
  • MalfunctionSep 8, 2022Smoking; Smoking

    IT HAS BEEN REPORTED TO PHILIPS THAT DURING A PLANNED NON-EMERGENCY TREATMENT PROCEDURE, A BURNING SMELL WAS COMING FROM THE ON/OFF BOARD. SMOKE WAS DETECTED BY THE CUSTOMER AND THERE WERE VISIBLE BURN MARKS ON THE BOARD. THE USER STOPPED USING THE SYSTEM AND COMPLETED THE PROCEDURE WITH ANOTHER SYSTEM. NO HARM HAS BEEN REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15380065 ↗

Other Interventional Fluoroscopic X-Ray System devices

All Interventional Fluoroscopic X-Ray System devices →