Azurion Azurion 7 M20 (REF 722079)Recalls & reported problems
Philips Medical Systems Nederland B.V. · Monoplane Interventional X-Ray System. A ceiling or floor-mounted monoplane system with a 20-inch flat detector. Optionally the ceiling-mounted configuration has an interface for an OR table · Class 2
All models of Azurion (22)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
27 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Philips Azurion 7 M20 X-ray system has 2,194 reports, including 1 death and 8 injuries. Many describe the system failing to produce X-rays or to activate. 28 recalls name this device, covering software, table movement and loose parts, and the most recent ones are still open. Reports fell to 252 in the last 12 months from 778 in the previous 12.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,176
- Deaths reported
- 1
- Injuries reported
- 8
- Recalls naming it
- 28
Summary
The FDA's MAUDE database contains 2,176 adverse event reports that list this exact device (DI 00884838085268), received between Dec 20, 2019 and Sep 28, 2026 — classified as 1 death, 8 injuries, 2,167 malfunctions.
The most frequently reported specific device problem is “Radiation Output Failure” (894 reports).
28 FDA recall records name this device's identifier, 27 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 16, 2026Class IIOpen, Classified
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
Z-2931-2026 · Root cause: Software design · Recall details · FDA record ↗
- Jun 17, 2026Class IIOpen, Classified
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
Z-2892-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Apr 30, 2026Class IIOpen, Classified
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
Z-2258-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Apr 28, 2026Class IIOpen, Classified
Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
Z-2225-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Apr 28, 2026Class IIOpen, Classified
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
Z-2227-2026 · Root cause: Component design/selection · Recall details · FDA record ↗
- Mar 19, 2026Class IIOpen, Classified
Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.
Z-1805-2026 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 3, 2026Class IIOpen, Classified
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Z-1678-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Jan 26, 2026Class IIOpen, Classified
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Z-1349-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Dec 12, 2025Class IIOpen, Classified
Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.
Z-1091-2026 · Root cause: Software design · Recall details · FDA record ↗
- Aug 1, 2025Class IIOpen, Classified
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
Z-2502-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jun 4, 2025Class IIOpen, Classified
Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.
Z-2134-2025 · Root cause: Device Design · Recall details · FDA record ↗
- May 28, 2025Class IIOpen, Classified
A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).
Z-1969-2025 · Root cause: Software design · Recall details · FDA record ↗
- Jan 13, 2025Class IIOpen, Classified
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
Z-1094-2025 · Root cause: Device Design · Recall details · FDA record ↗
- Jan 13, 2025Class IOpen, Classified
Possibility of the patient falling from the table related to the mattress used on the patient table.
Z-1160-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Aug 22, 2024Class IIOpen, Classified
Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
Z-3165-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Aug 6, 2024Class IIOpen, Classified
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel.
Z-2953-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 15, 2024Class IIOpen, Classified
Potential Loss of Imaging (X-ray) Functionality and/or Longer Time to Perform Cold Restart
Z-2534-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Jan 2, 2024Class IIOpen, Classified
Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
Z-1145-2024 · Root cause: Process design · Recall details · FDA record ↗
- Jan 2, 2024Class IIOpen, Classified
Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Z-1152-2024 · Root cause: Process design · Recall details · FDA record ↗
- Dec 12, 2023Class IIOpen, Classified
System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the procedure.
Z-0835-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Dec 8, 2023Class IIOpen, Classified
Loss of connectivity between the FlexArm and the Table due to a software issue.
Z-0860-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Dec 8, 2023Class IIOpen, Classified
Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
Z-0861-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Dec 7, 2023Class IIOpen, Classified
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Z-2291-2026 · Root cause: Software design · Recall details · FDA record ↗
- Nov 3, 2023Class IIOpen, Classified
Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose
Z-0559-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Jun 12, 2023Class IIOpen, Classified
The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
Z-2284-2023 · Root cause: Component design/selection · Recall details · FDA record ↗
- Mar 21, 2023Class IIOpen, Classified
Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure
Z-1647-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Feb 2, 2022Class IIOpen, Classified
Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.
Z-0874-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Sep 24, 2021Class IITerminated
When the user presses or releases both the APC (accept) button and the Float Tabletop (panning) button at the same time, the geometry may stop reacting on movement requests (table lock-up) and X-ray imaging becomes unavailable.
Z-0213-2022 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Radiation Output Failure894
- Activation Failure731
- Positioning Failure307
- No Display/Image124
- Detachment of Device or Device Component32
- Failure to Power Up20
- Unintended System Motion15
- Key or Button Unresponsive/not Working7
Patient problems
- No Clinical Signs, Symptoms or Conditions2,146
- Insufficient Information25
- No Known Impact Or Consequence To Patient8
- No Consequences Or Impact To Patient5
- Hemorrhage/Blood Loss/Bleeding2
- Abrasion1
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Bone Fracture(s)1
Recent adverse event reports
The 25 most recent of 2,176 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Radiation Output Failure
It was reported to Philips that the system was unable to expose, preventing imaging.The system was in clinical use at the time of the reported event. There was no harm reported. Philips has started an investigation of the complaint.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50916729 ↗MalfunctionSep 28, 2026Activation Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50916003 ↗MalfunctionSep 24, 2026Positioning Failure
It was reported to Philips that there was a remote alert indicating a motor assembly error within stand. There was no harm reported. Philips has started the investigation of the complaint.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50882664 ↗MalfunctionSep 23, 2026Activation Failure
It was reported to Philips that the system was not booting up and generating a message as "power failure".The system was not in use at the time of the reported event. There was no harm reported. Philips has started an investigation of the complaint.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50865467 ↗MalfunctionSep 23, 2026Radiation Output Failure
It was reported to Philips that the system could not emit X-ray.The device was in clinical use at the time of the event. No harm to the patient or user was reported.Philips has started an investigation of this complaint.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50864817 ↗MalfunctionMay 28, 2026Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S FLEXVISION COMPUTER WAS UNABLE TO BOOT UP. THE SYSTEM WAS OUTSIDE OF CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25317394 ↗MalfunctionMay 25, 2026Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT. THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME OF EVENT. NO HARM TO USER OR PATIENT WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267530 ↗MalfunctionMay 22, 2026Positioning Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S C-ARM WAS UNABLE TO PERFORM GEOMETRY MOVEMENTS. THE SYSTEM WAS IN CLINICAL USE DURING THE TIME OF THE REPORTED MALFUNCTION. THERE WAS NO HARM REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THE COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25252285 ↗MalfunctionMay 19, 2026Positioning Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS STUCK IN AP (ANTEROPOSTERIOR) POSITION, IMPACTING GEOMETRY MOVEMENT. THE SYSTEM WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25220064 ↗MalfunctionMay 15, 2026Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO PRODUCE X-RAY. THE SYSTEM WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25180351 ↗MalfunctionMay 14, 2026Positioning Failure
IT WAS REPORTED TO PHILIPS THAT THE GEOMETRY MOVEMENT WAS NOT AVAILABLE. THERE WAS NO HARM REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THE COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25170832 ↗MalfunctionMay 13, 2026Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE CIRCUIT BREAKERS TRIPPED WHEN THE POWER SOURCE WAS SWITCHED TO THE BACKUP POWER, THAT IMPACTED BOOTING. THE SYSTEM WAS NOT IN CLINICAL USE DURING THE TIME OF THE REPORTED MALFUNCTION. THERE WAS NO HARM REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THE COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25160381 ↗MalfunctionMay 8, 2026Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THERE WAS A ERROR STATING X-RAY WAS NOT POSSIBLE. THE USER PERFORMED A FULL RESTART, BUT THE ISSUE PERSISTED. THE SYSTEM WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25122779 ↗MalfunctionApr 30, 2026Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE USER PERFORMED A REBOOT, BUT THE ISSUE PERSISTED. THE SYSTEM WAS NOT IN CLINICAL USE DURING THE TIME OF THE REPORTED MALFUNCTION. THERE WAS NO HARM REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THE COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25042054 ↗MalfunctionApr 30, 2026Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT-UP. THE SYSTEM WAS NOT IN CLINICAL USE DURING THE TIME OF THE REPORTED MALFUNCTION. THERE WAS NO HARM REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THE COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25041624 ↗MalfunctionApr 30, 2026Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THERE WAS AN ERROR MESSAGE INDICATING THE SYSTEM WAS UNABLE TO PERFORM FLUOROSCOPY. THE USER PERFORMED A REBOOT, BUT THE ISSUE PERSISTED. THE SYSTEM WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25039297 ↗MalfunctionApr 24, 2026Radiation Output Failure
IT HAS BEEN REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO PRODUCE X-RAYS. NO HARM HAS BEEN REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24978670 ↗InjuryApr 21, 2026Unintended Collision
IT WAS REPORTED TO PHILIPS THAT THE C-ARM CAME INTO CONTACT WITH THE PATIENT¿S NOSE. THE REPORT INDICATED THAT THE PATIENT¿S NOSE WAS SCRAPED AND BLEEDING; HOWEVER, NO FURTHER INFORMATION REGARDING THE INCIDENT HAS BEEN PROVIDED TO DATE. AN INVESTIGATION INTO THIS COMPLAINT HAS BEEN INITIATED BY PHILIPS.
Patient problems: Insufficient Information; Unspecified Tissue Injury
MAUDE report 24942048 ↗MalfunctionApr 14, 2026Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE X-RAY STOPPED WORKING WHEN THE C-ARM WAS IN THE RIGHT ANTERIOR OBLIQUE (RAO) POSITION. THE SYSTEM WAS IN CLINICAL USE AT THE TIME OF THE REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24876357 ↗MalfunctionApr 9, 2026Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS OUTSIDE OF USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24830149 ↗MalfunctionApr 6, 2026Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE USE WAS UNKNOWN AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24790588 ↗MalfunctionApr 3, 2026Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO PRODUCE X-RAYS. THE PATIENT WAS MOVED TO ANOTHER SYSTEM. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24781190 ↗MalfunctionApr 3, 2026Image Display Error/Artifact
IT WAS REPORTED TO PHILIPS THAT THE VERTEBRAL BODY IMAGE WAS UNCLEAR, NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24773772 ↗MalfunctionApr 2, 2026Detachment of Device or Device Component
IT WAS REPORTED TO PHILIPS THAT THE CASING THAT COVERS THE C-ARM WAS COMING OFF. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24772046 ↗MalfunctionMar 30, 2026Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S X-RAY TUBE NEEDED TO BE REPLACED, WHICH CAN IMPACT IMAGING NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24730334 ↗
Other Interventional Fluoroscopic X-Ray System devices
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System25 recalls2 deaths3,984 reports
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths3,614 reports
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System32 recalls2 deaths2,308 reports
- O-ARM® BI70002000Medtronic Navigation, Inc. · Interventional fluoroscopic x-ray system2 recalls1 death3,667 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths2,015 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths1,837 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths1,087 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths835 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page