Skip to content

XperGuide XperGuide - 1Recalls & reported problems

Philips Medical Systems Nederland B.V. · Software medical device, used with a Philips Interventional X-Ray System. Extension of XperCT. Intended to assist the user during percutaneous interventions by providing live 3D needle image guidance. · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838059252), received between Apr 21, 2023 and Apr 21, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Image Orientation Incorrect” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Image Orientation Incorrect1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 21, 2023Image Orientation Incorrect; Image Orientation Incorrect

    IT WAS REPORTED TO PHILIPS THAT DURING A CBCT SCAN THE ORIENTATION OF THE XPERGUIDE DISPLAY WAS UPSIDE DOWN. THE PROCEDURE WAS ABORTED, HOWEVER NO HARM WAS REPORTED TO PHILIPS. THE USER LATER IDENTIFIED THAT THE TABLE TOP WAS POSITIONED BACKWARDS ON THE TABLE BASE (THE HEAD END AT THE FEET AND THE FEET END AT THE HEAD). PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16784718 ↗

Other Interventional Fluoroscopic X-Ray System devices

All Interventional Fluoroscopic X-Ray System devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.