Brilliance iCTRecalls & reported problems across all models

Philips Medical Systems Nederland B.V. · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

5 open recalls name at least one Brilliance iCT model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
3
Recalls
8

We found 2 models of Brilliance iCT by Philips Medical Systems Nederland B.V. in the FDA's device registry (GUDID). Together they are named in 9 adverse event reports in the FDA's MAUDE database, received between Dec 8, 2017 and Sep 22, 2026 — classified as 3 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Device Inoperable” (2 reports).

8 FDA recall records name at least one of these models by its device identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Brilliance iCT recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 7, 2026Open, Classified

    Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.

    Z-3040-2026 · Root cause: Radiation Control for Health and Safety Act · Recall details · FDA record ↗

  2. Jun 11, 2026Open, Classified

    The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.

    Z-2650-2026 · Root cause: Radiation Control for Health and Safety Act · Recall details · FDA record ↗

  3. Jan 7, 2026Open, Classified

    As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

    Z-1315-2026 · Root cause: Process change control · Recall details · FDA record ↗

  4. Aug 13, 2025Open, Classified

    The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

    Z-2598-2025 · Root cause: Process control · Recall details · FDA record ↗

  5. Jun 8, 2021Terminated

    When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.

    Z-2314-2021 · Root cause: Software design · Recall details · FDA record ↗

  6. May 1, 2021Open, Classified

    Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

    Z-1990-2021 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  7. Nov 10, 2017Terminated

    Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.

    Z-0391-2018 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  8. Nov 10, 2017Terminated

    Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.

    Z-0388-2018 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported Brilliance iCT problems

Device problems

  • Device Inoperable2
  • No Audible Alarm2
  • Adverse Event Without Identified Device or Use Problem1
  • Arcing1
  • Defective Component1
  • Degraded1
  • Signal Artifact/Noise1
  • Unintended System Motion1

Patient problems

  • No Known Impact Or Consequence To Patient5
  • Bone Fracture(s)1
  • Insufficient Information1
  • Laceration(s)1
  • No Clinical Signs, Symptoms or Conditions1

Brilliance iCT models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Brilliance iCT adverse event reports

The 9 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Philips Medical Systems Nederland B.V. products

All Philips Medical Systems Nederland B.V. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page