Skip to content

NA B.01Recalls & reported problems

Philips Medical Systems Hsg · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
1
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00884838081284), received between Apr 26, 2018 and Oct 10, 2023 — classified as 1 death, 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Alarm Not Visible” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Alarm Not Visible1
  • Communication or Transmission Problem1
  • Device Alarm System1
  • Device Displays Incorrect Message1
  • Wireless Communication Problem1

Patient problems

  • Cardiopulmonary Arrest2
  • No Clinical Signs, Symptoms or Conditions2
  • Death1
  • Loss Of Pulse1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 10, 2023Wireless Communication Problem; Wireless Communication Problem

    PHILIPS RECEIVED A COMPLAINT ON THE CAREEVENT (866345 INDICATING THAT DEVICE SENT A FALSE SIGNAL (PROBLEM WITH THE RE-UP). PATIENT INVOLVEMENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17906586 ↗
  • MalfunctionJan 19, 2021Device Alarm System

    THE CUSTOMER REPORTED AN ALARM PROBLEM, THERE IS NO CALL FORWARDING. THE DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11195271 ↗
  • InjuryNov 5, 2020Communication or Transmission Problem

    THE CUSTOMER REPORTED THAT A CODE BLUE WAS NOT PAGED TO THE NURSES' PHONES. THE DEVICE WAS IN USE. PATIENT DETAILS WERE NOT PROVIDED.

    Patient problems: Cardiopulmonary Arrest

    MAUDE report 10794638 ↗
  • InjuryFeb 5, 2020Alarm Not Visible

    THE CUSTOMER REPORTED THAT AN ALARM WAS NOT RECEIVED ON (B)(6) 2020 AT 19:49:40 FOR A PATIENT IN ROOM 589A. THE OUTPUT WAS ON ARUBA ACCESS POINTS (AP) WITH PIVOT 8742 SMART PHONES. ERROR "NO DEVICE ASSIGNMENT" LISTED IN EVENT TRANSCRIPT REPORT. THE PATIENT HAD SUFFERED A CARDIAC ARREST. THE STAFF ENTERED THE PATIENT ROOM FOR ROUTINE BLOOD DRAW AND FOUND THE PATIENT HAD ARRESTED WITH A FLAT LINE HEART RATE AND NOT REGISTERING A PULSE OXYGENATION. THE PATIENT WAS ABLE TO BE RESUSCITATED.

    Patient problems: Cardiopulmonary Arrest; Loss Of Pulse

    MAUDE report 9670774 ↗
  • DeathApr 26, 2018Device Displays Incorrect Message

    IT WAS REPORTED THAT THE HOSPITAL STAFF AT THE SITE WAS PERFORMING AN INVESTIGATION OF A PATIENT EVENT WHERE THE PATIENT EXPIRED BETWEEN 6:40AM -7:40AM. THE CUSTOMER WANTED TO UNDERSTAND IF THERE WAS A DELAY IN CARE EVENT SENDING ALERTS TO THE MOBIL PHONES. INVESTIGATION IS NEEDED TO UNDERSTAND IF ANY PHILIPS PRODUCT CAUSED OR CONTRIBUTED TO THE PATIENT EVENT.

    Patient problems: Death

    MAUDE report 7464322 ↗

FDA clearance and classification

The FDA regulates product code MSX (system,network and communication,physiological monitors) under 21 CFR 870.2300 as a Class II device. 21 CFR 870.2300 ↗

Devices of this type normally need 510(k) clearance before sale.

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA B.01

Has the NA B.01 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00884838081284), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA B.01?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Apr 26, 2018 and Oct 10, 2023. In those reports, the most common specific device problems are “Alarm Not Visible” (1), “Communication or Transmission Problem” (1) and “Device Alarm System” (1); the most common patient problems are “Cardiopulmonary Arrest” (2), “Death” (1) and “Loss Of Pulse” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA B.01?

1 report is classified as a death, 2 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the NA B.01, and is it still sold?

The FDA's device database lists Philips Medical Systems Hsg as the company responsible for it (brand NA, model B.01). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SYSTEM,NETWORK AND COMMUNICATION,PHYSIOLOGICAL MONITORS devices

All SYSTEM,NETWORK AND COMMUNICATION,PHYSIOLOGICAL MONITORS devices →

Sources and updates