NA B.01Recalls & reported problems
Philips Medical Systems Hsg · Class 2
- Adverse event reports
- 5
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00884838081284), received between Apr 26, 2018 and Oct 10, 2023 — classified as 1 death, 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Alarm Not Visible” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Alarm Not Visible1
- Communication or Transmission Problem1
- Device Alarm System1
- Device Displays Incorrect Message1
- Wireless Communication Problem1
Patient problems
- Cardiopulmonary Arrest2
- No Clinical Signs, Symptoms or Conditions2
- Death1
- Loss Of Pulse1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 10, 2023Wireless Communication Problem; Wireless Communication Problem
PHILIPS RECEIVED A COMPLAINT ON THE CAREEVENT (866345 INDICATING THAT DEVICE SENT A FALSE SIGNAL (PROBLEM WITH THE RE-UP). PATIENT INVOLVEMENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17906586 ↗MalfunctionJan 19, 2021Device Alarm System
THE CUSTOMER REPORTED AN ALARM PROBLEM, THERE IS NO CALL FORWARDING. THE DEVICE WAS NOT IN USE ON A PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11195271 ↗InjuryNov 5, 2020Communication or Transmission Problem
THE CUSTOMER REPORTED THAT A CODE BLUE WAS NOT PAGED TO THE NURSES' PHONES. THE DEVICE WAS IN USE. PATIENT DETAILS WERE NOT PROVIDED.
Patient problems: Cardiopulmonary Arrest
MAUDE report 10794638 ↗InjuryFeb 5, 2020Alarm Not Visible
THE CUSTOMER REPORTED THAT AN ALARM WAS NOT RECEIVED ON (B)(6) 2020 AT 19:49:40 FOR A PATIENT IN ROOM 589A. THE OUTPUT WAS ON ARUBA ACCESS POINTS (AP) WITH PIVOT 8742 SMART PHONES. ERROR "NO DEVICE ASSIGNMENT" LISTED IN EVENT TRANSCRIPT REPORT. THE PATIENT HAD SUFFERED A CARDIAC ARREST. THE STAFF ENTERED THE PATIENT ROOM FOR ROUTINE BLOOD DRAW AND FOUND THE PATIENT HAD ARRESTED WITH A FLAT LINE HEART RATE AND NOT REGISTERING A PULSE OXYGENATION. THE PATIENT WAS ABLE TO BE RESUSCITATED.
Patient problems: Cardiopulmonary Arrest; Loss Of Pulse
MAUDE report 9670774 ↗DeathApr 26, 2018Device Displays Incorrect Message
IT WAS REPORTED THAT THE HOSPITAL STAFF AT THE SITE WAS PERFORMING AN INVESTIGATION OF A PATIENT EVENT WHERE THE PATIENT EXPIRED BETWEEN 6:40AM -7:40AM. THE CUSTOMER WANTED TO UNDERSTAND IF THERE WAS A DELAY IN CARE EVENT SENDING ALERTS TO THE MOBIL PHONES. INVESTIGATION IS NEEDED TO UNDERSTAND IF ANY PHILIPS PRODUCT CAUSED OR CONTRIBUTED TO THE PATIENT EVENT.
Patient problems: Death
MAUDE report 7464322 ↗
Other SYSTEM,NETWORK AND COMMUNICATION,PHYSIOLOGICAL MONITORS devices
- Xhibit Xhibit CentralSpacelabs Healthcare (Washington), Inc · System, network and communication, physiological monitors3 deaths218 reports
- Infinity M540 MS26372Draeger Medical Systems, Inc. · System,Network And Communication,Physiological Monitors2 recalls9 reports
- CareEventPhilips Medical Systems Hsg · System, network and communication, physiological monitors2 recalls4 deaths7 reports
- XTR 96280Spacelabs Healthcare (Washington), Inc · SYSTEM,NETWORK AND COMMUNICATION,PHYSIOLOGICAL MONITORS2 recalls0 reports
- CNS-2101 Central MonitorNihon Kohden Corporation · System, Network And Communication, Physiological Monitors3 deaths76 reports
- BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd. · System, Network And Communication, Physiological Monitors1 recall2 deaths5 reports
- eCareManager 4.5Visicu, Inc. · System, network and communication, physiological monitors1 recall1 death1 report
- eCareManager 4.3Visicu, Inc. · System, network and communication, physiological monitors1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.