Skip to content

NA B.01Recalls & reported problems

Philips Medical Systems Hsg · Class 2

Adverse event reports
5
Deaths reported
1
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00884838081284), received between Apr 26, 2018 and Oct 10, 2023 — classified as 1 death, 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Alarm Not Visible” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Alarm Not Visible1
  • Communication or Transmission Problem1
  • Device Alarm System1
  • Device Displays Incorrect Message1
  • Wireless Communication Problem1

Patient problems

  • Cardiopulmonary Arrest2
  • No Clinical Signs, Symptoms or Conditions2
  • Death1
  • Loss Of Pulse1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 10, 2023Wireless Communication Problem; Wireless Communication Problem

    PHILIPS RECEIVED A COMPLAINT ON THE CAREEVENT (866345 INDICATING THAT DEVICE SENT A FALSE SIGNAL (PROBLEM WITH THE RE-UP). PATIENT INVOLVEMENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT OR USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17906586 ↗
  • MalfunctionJan 19, 2021Device Alarm System

    THE CUSTOMER REPORTED AN ALARM PROBLEM, THERE IS NO CALL FORWARDING. THE DEVICE WAS NOT IN USE ON A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11195271 ↗
  • InjuryNov 5, 2020Communication or Transmission Problem

    THE CUSTOMER REPORTED THAT A CODE BLUE WAS NOT PAGED TO THE NURSES' PHONES. THE DEVICE WAS IN USE. PATIENT DETAILS WERE NOT PROVIDED.

    Patient problems: Cardiopulmonary Arrest

    MAUDE report 10794638 ↗
  • InjuryFeb 5, 2020Alarm Not Visible

    THE CUSTOMER REPORTED THAT AN ALARM WAS NOT RECEIVED ON (B)(6) 2020 AT 19:49:40 FOR A PATIENT IN ROOM 589A. THE OUTPUT WAS ON ARUBA ACCESS POINTS (AP) WITH PIVOT 8742 SMART PHONES. ERROR "NO DEVICE ASSIGNMENT" LISTED IN EVENT TRANSCRIPT REPORT. THE PATIENT HAD SUFFERED A CARDIAC ARREST. THE STAFF ENTERED THE PATIENT ROOM FOR ROUTINE BLOOD DRAW AND FOUND THE PATIENT HAD ARRESTED WITH A FLAT LINE HEART RATE AND NOT REGISTERING A PULSE OXYGENATION. THE PATIENT WAS ABLE TO BE RESUSCITATED.

    Patient problems: Cardiopulmonary Arrest; Loss Of Pulse

    MAUDE report 9670774 ↗
  • DeathApr 26, 2018Device Displays Incorrect Message

    IT WAS REPORTED THAT THE HOSPITAL STAFF AT THE SITE WAS PERFORMING AN INVESTIGATION OF A PATIENT EVENT WHERE THE PATIENT EXPIRED BETWEEN 6:40AM -7:40AM. THE CUSTOMER WANTED TO UNDERSTAND IF THERE WAS A DELAY IN CARE EVENT SENDING ALERTS TO THE MOBIL PHONES. INVESTIGATION IS NEEDED TO UNDERSTAND IF ANY PHILIPS PRODUCT CAUSED OR CONTRIBUTED TO THE PATIENT EVENT.

    Patient problems: Death

    MAUDE report 7464322 ↗

Other SYSTEM,NETWORK AND COMMUNICATION,PHYSIOLOGICAL MONITORS devices

All SYSTEM,NETWORK AND COMMUNICATION,PHYSIOLOGICAL MONITORS devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.