Skip to content

Stress ST80i Stress Test SystemRecalls & reported problems

Philips Medical Systems Hsg · Philips ST80i Stress System · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00884838094840), received between May 23, 2022 and Jul 22, 2025 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Fail-Safe Did Not Operate” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fail-Safe Did Not Operate3
  • Failure to Auto Stop2
  • Unintended System Motion2
  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Insufficient Information1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 22, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THE CUSTOMER REPORTED THAT THEY HIT THE EMERGENCY STOP BUTTON; HOWEVER, THE TREADMILL KEPT GOING. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT. NO ADVERSE EVENT TO THE PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22569522 ↗
  • MalfunctionOct 9, 2024Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    IT WAS REPORTED THAT THE EMERGENCY STOP BUTTON WAS NOT WORKING ON TREADMILL. THE DEVICE WAS REPORTED TO BE IN USE ON A PATIENT. NO ADVERSE EVENT TO THE PATIENT OR USER WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20406151 ↗
  • MalfunctionMar 18, 2024Failure to Auto Stop; Failure to Auto Stop

    THE CUSTOMER REPORTED THE EMERGENCY STOP IS NOT STOPPING. THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18924653 ↗
  • MalfunctionMar 16, 2023Failure to Auto Stop; Failure to Auto Stop

    PHILIPS RECEIVED A COMPLAINT ON THE OEM TRACKMASTER TMX428 TREADMILL 220V INDICATING THAT THE EMERGENCY STOP (E-STOP) BUTTON WAS NO LONGER WORKING. IT IS UNKNOWN IF THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT, NO ADVERSE EVENT OR PATIENT HARM WAS REPORTED. THE BIOMED ENGINEER AT THE CUSTOMER FACILITY EVALUATED THE DEVICE AND DETERMINED THAT THE E-STOP BUTTON REQUIRES REPLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16556083 ↗
  • MalfunctionFeb 22, 2023Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THE CUSTOMER REPORTED THAT THE TREADMILL EMERGENCY STOP BUTTON WAS NOT FUNCTIONING PROPERLY ON THE PHILIPS ST80I STRESS TEST SYSTEM. IT IS UNKNOWN IF THE DEVICE WAS IN USE ON A PATIENT AT THE TIME OF THE REPORTED ISSUE. NO DEATH OR PATIENT / USER INJURY OR HARM WAS REPORTED. A PHILIPS REMOTE SERVICE ENGINEER (RSE) SPOKE TO THE BIOMED AT THE CUSTOMER SITE. DURING THE TESTING IT WAS CONFIRMED THAT THE CAUSE FOR THE REPORTED ISSUE WAS A FAILED EMERGENCY STOP BUTTON. THE BIOMED ORDERED AND WAS PROVIDED WITH A REPLACEMENT EMERGENCY STOP ASSEMBLY TO REPAIR THE DEVICE AND TO RESOLVE THE ISSUE. THE DEVICE FAILED TO WORK AS INTENDED. THIS ISSUE WAS CAUSED BY A MALFUNCTION OF THE DEVICE AND WAS RESOLVED BY THE CUSTOMER REPLACING THE EMERGENCY STOP ASSEMBLY. IT HAS BEEN CONCLUDED THAT NO FURTHER ACTION IS REQUIRED. THE DEVICE REMAINS AT THE CUSTOMER SITE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16415707 ↗
  • MalfunctionFeb 17, 2023Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED THAT WHILE A STRESS STUDY WAS BEING PERFORMED, THE TREADMILL E-STOP BUTTON WAS UNRESPONSIVE. THE PATIENT HAD TO JUMP OFF THE TREADMILL TO SAFELY EXIT THE TREADMILL. THERE WAS NO REPORT OF A DEATH OR SERIOUS INJURY, NOR WAS THERE A REPORT OF ANY ADVERSE IMPACT TO ANY USER OR PATIENT. A PHILIPS AUTHORIZED THIRD PARTY FIELD SERVICE ENGINEER (FSE) VISITED THE CUSTOMER FACILITY TO EVALUATE THE DEVICE. THE FSE CONFIRMED THAT THEY WERE UNABLE TO REPLICATE THE E-STOP ISSUE. THE FSE STATES THAT A FACILITY TECHNICIAN REMOVED AND REINSTALLED THE E-STOP BUTTON WHICH RESULTED IN THE E-STOP BUTTON STOPPING THE TREADMILL WHEN ENGAGED, RESOLVING THE ISSUE. THE FSE WAS UNABLE TO CONFIRM THAT THE E-STOP BUTTON HAD FAILED, AS IT WAS WORKING AS SPECIFIED DURING HIS EVALUATION. BASED ON THE INFORMATION AVAILABLE AND THE TESTING CONDUCTED WE WERE UNABLE TO REPLICATE THE REPORTED PROBLEM. THE REPORTED PROBLEM WAS NOT CONFIRMED. THE DEVICE WAS CONFIRMED TO BE OPERATING PER SPECIFICATIONS AND NO FAILURE WAS IDENTIFIED. THE INVESTIGATION CONCLUDES THAT NO FURTHER ACTION IS REQUIRED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16395474 ↗
  • InjuryJan 18, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED THAT THEY¿VE GONE TO SWITCH IT ON THE DEVICE WITH A PATIENT IN SITU, AND THE BELT HAS THROWN THIS PERSON BACKWARDS AND TREADMILL GONE INTO REVERSE. THE PATIENT IS HARMED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16194404 ↗
  • MalfunctionJun 21, 2022Unintended System Motion; Insufficient Device Problem Information; Unintended System Motion; Insufficient Device Problem Information

    THE CUSTOMER REPORTED THAT RED BUTTON IS NOT STOPPING/ PIM NOT COMMUNICATING WITH COMPUTER. IT UNKNOWN IF THE DEVICE WAS IN USE AT THE TIME OF THE EVENT, NO REPORTED PATENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14758728 ↗
  • MalfunctionMay 23, 2022Unintended System Motion; Unintended System Motion

    THE CUSTOMER REPORTED THAT THE TREADMILL CONTINUED TO RUN AFTER THE COMPLETION OF THE TEST, AND THE STOP BUTTON WAS NOT AVAILABLE. IT WAS ALSO STATED THAT THE DEVICE FROZE FROM ALLOWING TO PUT PATIENT COMMENTS. THE CUSTOMER ADVISED THAT THE ISSUE HAPPENED THE PRIOR WEEK AND COULD NOT RECREATE THE ISSUE. IT IS UNKNOWN IF THE DEVICE WAS IN CLINICAL USE AT THE TIME THE REPORTED ISSUE WAS DISCOVERED. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14470994 ↗

Other Computer, diagnostic, programmable devices

All Computer, diagnostic, programmable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page