Affiniti 70 Diagnostic Ultrasound System 795210 (UDI 00884838097766)Recalls & reported problems
Philips Ultrasound LLC · Class 2
All models of Affiniti 70 Diagnostic Ultrasound System (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00884838097766), received between Jul 21, 2022 and Apr 17, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 6, 2021Class IICompleted
Due to a software defect that can intermittently cause the system to lock-up which exiting Review Mode while performing an exam.
Z-2484-2021 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Application Program Freezes or Fails to Launch1
- Electrical /Electronic Property Problem1
- Incorrect Measurement1
- Poor Quality Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2026Application Program Freezes or Fails to Launch
IT WAS REPORTED THAT A CRITICAL PROCEDURE COULD NOT BE COMPLETED ON THE AFFINITI 70 ULTRASOUND SYSTEM. THE SYSTEM DISPLAYED AN ERROR MESSAGE DURING AN ARTERIAL LINE INSERTION PROCEDURE. THE PROCEDURE WAS ABLE TO BE COMPLETED WITH ANOTHER ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE PHILIPS SERVICE ENGINEER INITIATED THE REPAIR PROCESS TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERN. PHILIPS HAS STARTED AN INVESTIGATION, AND A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24913441 ↗MalfunctionApr 10, 2026Poor Quality Image
IT WAS REPORTED THAT A CRITICAL CARDIAC PROCEDURE WITH THE X7-2T TRANSDUCER COULD NOT BE COMPLETED ON THE AFFINITI 70 ULTRASOUND SYSTEM DUE TO IMAGE ARTIFACT. THE PROCEDURE WAS ABLE TO BE COMPLETED ON ANOTHER ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE PHILIPS SERVICE ENGINEER CLEANED THE CONNECTORS ON THE SYSTEM PORTS AND TRANSDUCER TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERN. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24841763 ↗MalfunctionJul 3, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE INTRACAVITY C10-3V TRANSDUCER FAILED ELECTRICAL SAFETY TESTING. THE TRANSDUCER WAS ASSOCIATED WITH THE AFFINITI 70 ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THE ISSUE WAS DISCOVERED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE PHILIPS SERVICE ENGINEER CONFIRMED THERE WAS NO PHYSICAL DAMAGE TO THE TRANSDUCER. THE TRANSDUCER WAS REPLACED TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22396342 ↗MalfunctionJul 21, 2022Incorrect Measurement; Incorrect Measurement
A CUSTOMER REPORTED THEIR AFFINITI 70 ULTRASOUND SYSTEM HAD PRODUCED INCORRECT VOLUME CALCULATIONS DURING A PROSTATE EXAM DUE TO AN INCORRECT VOLUME CALCULATION ENTRY WHEN COMPOSING A CUSTOM PRESET MANUALLY. THE ISSUE WAS OBVIOUS TO THE USER AND THERE WAS NO HARM TO THE USER OR PATIENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15079643 ↗
Other System, imaging, pulsed doppler, ultrasonic devices
- Transducer C5-1 989605412042Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic11 recalls0 reports
- Transducer X5-1 989605400802Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic10 recalls0 reports
- EPIQ 7 Diagnostic Ultrasound System 989605386721Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls2 deaths146 reports
- Transducer S7-3t 989605406772Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic5 recalls1 death18 reports
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS GF TYPE UCT180Olympus Medical Systems Corp. · System, imaging, pulsed doppler, ultrasonic199 reports
- Transducer S7-3t 989605439232Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls4 reports
- Transducer 3D9-3v 989605395622Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls1 report
- ACUSON Redwood Diagnostic Ultrasound System 11503314Siemens Medical Solutions USA, Inc. · System, imaging, pulsed doppler, ultrasonic3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page