Skip to content

EPIQ 5 DIAGNOSTIC ULTRASOUND SYSTEM 795204Recalls & reported problems

Philips Ultrasound LLC · Class 2

All models of EPIQ 5 DIAGNOSTIC ULTRASOUND SYSTEM (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00884838097896), received between Apr 8, 2025 and Nov 14, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1
  • Material Integrity Problem1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2025Fracture; Material Separation; Fracture; Material Separation

    IT WAS REPORTED THE C10-3V TRANSDUCER HAD SEPARATION ON THE SEAM. THIS WAS ASSOCIATED WITH AN EPIQ 5G ULTRASOUND SYSTEM. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE SERVICE ENGINEER CONFIRMED THE REPORTED ISSUE AND REPLACED THE TRANSDUCER TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23554544 ↗
  • MalfunctionApr 8, 2025Material Integrity Problem; Material Integrity Problem

    AS PART OF CAPA 5677695, IT WAS IDENTIFIED THAT RETROSPECTIVE REVIEW ACTIVITIES WERE WARRANTED FOR INTRACAVITY TRANSDUCER DAMAGE IF PHYSICALLY SPLIT, FRACTURED, OR CONTAINS SHARP EDGES ON THE TRANSDUCER, REGARDLESS OF WHETHER THE DEVICE IS IN CLINICAL USE. PHILIPS ULTRASOUND IS REPORTING THIS COMPLAINT AS PART OF THE RETROSPECTIVE REVIEW COMPLETED FOR INTRACAVITY TRANSDUCER DAMAGE. IT WAS REPORTED THE C10-3V TRANSDUCER HAD PHYSICAL DAMAGE TO THE LENS CAUSING IT TO FAIL THE ELECTRICAL LEAKAGE TEST. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE SERVICE ENGINEER CONFIRMED THE REPORTED ISSUE AND REPLACED THE TRANSDUCER TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21800705 ↗

Other System, imaging, pulsed doppler, ultrasonic devices

All System, imaging, pulsed doppler, ultrasonic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page