Ingenia Ingenia 3.0T CX (REF 782105)Recalls & reported problems

Philips Medical Systems Nederland B.V. · Enables trained physicians to obtain cross-sectional and spectroscopic images of internal structures of the head, body or extremities, in any orientation, representing the spatial distribution of protons or other nuclei with spin. · Class 2

All models of Ingenia (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

7 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Philips Ingenia 3.0T MRI system makes internal body images. The 40 adverse event reports describe a scanner body-coil seal breaking or coming loose, exposing sharp edges; there were no deaths and no injuries. Reports fell from 36 to 4 in the past year. This device has 7 open recalls covering software that may end exams during reconstruction, overheating gradient coil parts, loose screws, mains-unit hotspots, table plate misinstallation, and seal adhesive failure.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
41
Deaths reported
0
Injuries reported
0
Recalls naming it
7

Summary

The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 00884838098312), received between Nov 26, 2024 and Sep 18, 2026 — classified as 41 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (41 reports).

7 FDA recall records name this device's identifier, 7 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 22 reportsFeb 2025: 7 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202622
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 21, 2026Class IIOpen, Classified

    A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

    Z-3238-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Apr 14, 2026Class IIOpen, Classified

    The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

    Z-1959-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Jun 30, 2025Class IIOpen, Classified

    The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

    Z-2207-2025 · Root cause: Device Design · Recall details · FDA record ↗

  4. Sep 3, 2024Class IIOpen, Classified

    Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

    Z-3196-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  5. May 3, 2024Class IIOpen, Classified

    The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

    Z-1849-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  6. Mar 12, 2024Class IIOpen, Classified

    Patient support table floor plate may be incorrectly installed.

    Z-1567-2024 · Root cause: Process design · Recall details · FDA record ↗

  7. Jan 3, 2024Class IIOpen, Classified

    The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.

    Z-0951-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component41

Patient problems

  • No Clinical Signs, Symptoms or Conditions41

Recent adverse event reports

The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2026Detachment of Device or Device Component

    Philips received a report that the rubber on the Quadrature Body Coil (QBC) was peeling off. This seal is used for protection, to avoid the patient from potentially contacting the sharp edges of the Magnet covers.If this seal is damaged/comes off, it is likely that this could lead to serious injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50814013 ↗
  • MalfunctionMar 25, 2026Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL HAD BEEN DAMAGED AS THE MAGNET HOLE ADHESIVE STRIP HAD COME OFF. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS DAMAGED IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24684439 ↗
  • MalfunctionMar 11, 2026Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL BROKEN. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS BROKEN, IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24566973 ↗
  • MalfunctionFeb 2, 2026Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL UPPER WAS DAMAGED. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24232406 ↗
  • MalfunctionNov 14, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL WAS LOOSE. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS GASKET CAME OUT, IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23551964 ↗
  • MalfunctionOct 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL WAS BROKEN. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS BROKEN IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23434660 ↗
  • MalfunctionOct 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) RUBBER STRIP WAS FELL OFF. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS FALLEN OFF IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23414552 ↗
  • MalfunctionOct 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) RUBBER (FRONT BODY COIL SEAL ) WAS FELL OFF. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS FALLEN OFF IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23348346 ↗
  • MalfunctionAug 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL WAS LOOSE. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22923752 ↗
  • MalfunctionAug 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL WAS COMING OFF. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE, IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22881900 ↗
  • MalfunctionJul 18, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL FELL OFF. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22539778 ↗
  • MalfunctionFeb 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL IS BROKEN. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS DAMAGED IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21353466 ↗
  • MalfunctionFeb 10, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL FELL OFF. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21342017 ↗
  • MalfunctionFeb 10, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL WAS BROKEN. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21342014 ↗
  • MalfunctionFeb 10, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL WAS LOOSE. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE, IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21342011 ↗
  • MalfunctionFeb 4, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QBC SEAL FELL OFF (LOOSE). THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS.IF THIS SEAL IS LOOSE IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21301402 ↗
  • MalfunctionFeb 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL FELL OFF. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS BROKEN IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21292062 ↗
  • MalfunctionFeb 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) RUBBER STRIP IS DAMAGED. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS STRIP IS DAMAGED, IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21291476 ↗
  • MalfunctionJan 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL MAGNET TAPE WAS LOOSE. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21283037 ↗
  • MalfunctionJan 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL WAS LOOSE. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21282088 ↗
  • MalfunctionJan 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL CAME OFF. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21281378 ↗
  • MalfunctionJan 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL CAME OFF. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21281377 ↗
  • MalfunctionJan 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL WAS BROKEN. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS BROKEN IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21281372 ↗
  • MalfunctionJan 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL CAME OFF. THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21281371 ↗
  • MalfunctionJan 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    PHILIPS RECEIVED A REPORT THAT THE QUADRATURE BODY COIL (QBC) SEAL WAS BROKEN (DAMAGED). THIS SEAL IS USED FOR PROTECTION, TO AVOID THE PATIENT FROM POTENTIALLY CONTACTING THE SHARP EDGES OF THE MAGNET COVERS. IF THIS SEAL IS LOOSE, IT IS LIKELY THAT THIS COULD LEAD TO SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21281336 ↗

Other System, nuclear magnetic resonance imaging devices

All System, nuclear magnetic resonance imaging devices →