EPIQ Elite Diagnostic Ultrasound System 989606706051Recalls & reported problems
Philips Ultrasound LLC · Class 2
All models of EPIQ Elite Diagnostic Ultrasound System (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Philips EPIQ Elite diagnostic ultrasound system is used for imaging exams including cardiac, vascular and breast scans. The 32 reports mostly describe the console or system moving on its own, blurry or poor quality images, and problems with attached transducers, including probes cracking, separating, or failing leakage tests; one report mentions a system crashing mid-exam and another describes a minor shock when unplugging it. There were no deaths and 2 injuries, with 30 malfunctions. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 32
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 00884838099111), received between Jan 26, 2021 and Sep 25, 2026 — classified as 2 injuries, 30 malfunctions.
The most frequently reported specific device problem is “Unintended System Motion” (16 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended System Motion16
- Poor Quality Image8
- Electrical /Electronic Property Problem1
- False Negative Result1
- Fracture1
- Image Orientation Incorrect1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions30
- Insufficient Information2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026Unintended System Motion
It was reported the control panel arm on an EPIQ Elite Diagnostic Ultrasound System's dropped quickly. The problem occurred outside of clinical use and no patient or user was harmed.Philips has started an investigation, and upon completion, a follow up report will be submitted.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50899389 ↗MalfunctionJan 28, 2026Fracture
IT WAS REPORTED THE C10-3V TRANSDUCER HAD A HOLE IN THE LENS WITH EXPOSED ELECTRONICS. THE TRANSDUCER WAS ASSOCIATED WITH AN EPIQ ELITE ULTRASOUND SYSTEM. THIS ISSUE WAS FOUND OUTSIDE OF CLINICAL USE AND THERE WAS NO PATIENT OR USER HARM. A REPLACEMENT TRANSDUCER WAS SHIPPED DIRECTLY TO THE CUSTOMER TO ADDRESS THEIR IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24198615 ↗MalfunctionOct 23, 2025Material Separation; Material Separation
IT REPORTED A 3D9-3V TRANSDUCER HAD SEPARATION ON THE PROBE. THIS PROBE IS ASSOCIATED WITH THE EPIQ ELITE ULTRASOUND SYSTEM. THE ISSUE WAS FOUND OUTSIDE OF PATIENT USE, AND THERE WAS NO PATIENT OR USER HARM. THE SERVICE ENGINEER EVALUATED THE TRANSDUCER TO CONFIRM THE REPORTED PROBLEM AND REPLACED THE TRANSDUCER TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23368040 ↗MalfunctionSep 18, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE C10-3V TRANSDUCER FAILED LEAKAGE TESTING. THE TRANSDUCER WAS ASSOCIATED WITH THE EPIQ ELITE ULTRASOUND SYSTEM AT THE CUSTOMER¿S SITE. THE ISSUE WAS DISCOVERED OUTSIDE OF CLINICAL USE, DURING PREVENTATIVE MAINTENANCE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. PHILIPS FIELD SERVICE ENGINEER REPLACED THE TRANSDUCER TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23092586 ↗MalfunctionJul 2, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down
IT WAS REPORTED THE EPIQ ELITE ULTRASOUND SYSTEM WAS NOT AVAILABLE DURING A CRITICAL RENAL ARTERY EXAM IN THE EMERGENCY DEPARTMENT. THE SYSTEM CRASHED DURING THE PROCEDURE AND WAS UNABLE TO BE RECOVERED. ANOTHER SYSTEM WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22382563 ↗MalfunctionJun 2, 2025Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ ELITE ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A SERVICE ENGINEER PERFORMED A SYSTEM INSPECTION AND TESTED THE CONTROL PANEL LOCKING MECHANISM. THE ISSUE COULD NOT BE REPRODUCED. THE LOCKING MECHANISM WAS WORKING WITHOUT ISSUE, AND NO PROBLEMS WERE OBSERVED. THE SYSTEM HAS BEEN RETURNED TO SERVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22128544 ↗InjuryFeb 19, 2025Poor Quality Image; Poor Quality Image
IT WAS REPORTED THE EPIQ ELITE ULTRASOUND SYSTEM ALONG WITH AN EL18-4 TRANSDUCER WERE PRODUCING IMAGES DURING A BREAST SCAN THAT WERE NOT ABLE TO DIAGNOSE ANECHOIC CYSTS ACCURATELY. PATIENTS HAD UNNECESSARY ASPIRATIONS OF CYSTS. THE CUSTOMER IS CURRENTLY USING AN L12-5 TRANSDUCER AND OBTAINING ADDITIONAL IMAGES TO PREVENT TREATMENT OF ANECHOIC CYSTS. NO FURTHER INFORMATION WAS PROVIDED. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: Insufficient Information
MAUDE report 21412239 ↗InjuryFeb 19, 2025Poor Quality Image; Poor Quality Image
IT WAS REPORTED THE EPIQ ELITE ULTRASOUND SYSTEM ALONG WITH AN EL18-4 TRANSDUCER WERE PRODUCING IMAGES DURING A BREAST SCAN THAT WERE NOT ABLE TO DIAGNOSE ANECHOIC CYSTS ACCURATELY. PATIENTS HAD UNNECESSARY ASPIRATIONS OF CYSTS. THE CUSTOMER IS CURRENTLY USING AN L12-5 TRANSDUCER AND OBTAINING ADDITIONAL IMAGES TO PREVENT TREATMENT OF ANECHOIC CYSTS. NO FURTHER INFORMATION WAS PROVIDED. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: Insufficient Information
MAUDE report 21411989 ↗MalfunctionFeb 6, 2025Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED HAVING CHALLENGES WITH THE 2D IMAGE QUALITY IN THE CALF VEINS; THEY ALLEGED THEY COULD NOT DISTINGUISH A CLOTTED VESSEL FROM A NON-CLOTTED ONE WHEN USING THE EPIQ ELITE ULTRASOUND SYSTEM. ALTHOUGH THERE WAS NO MISDIAGNOSIS AS THE PATIENT HAD A PRIOR EXAM TO COMPARE, THEY COULD NOT DIAGNOSE PROPERLY BASED ON THE STUDY. THERE WAS NO ALLEGATION OF PATIENT/USER HARM, INJURY, OR DEATH. AN INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE OF THE ISSUE. A FOLLOW-UP REPORT WILL BE PROVIDED UPON INVESTIGATION COMPLETION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21326989 ↗MalfunctionFeb 6, 2025Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED HAVING CHALLENGES WITH THE 2D IMAGE QUALITY IN THE CALF VEINS; THEY ALLEGED THEY COULD NOT DISTINGUISH A CLOTTED VESSEL FROM A NON-CLOTTED ONE WHEN USING THE EPIQ ELITE ULTRASOUND SYSTEM. ALTHOUGH THERE WAS NO MISDIAGNOSIS AS THE PATIENT HAD A PRIOR EXAM TO COMPARE, THEY COULD NOT DIAGNOSE PROPERLY BASED ON THE STUDY. THERE WAS NO ALLEGATION OF PATIENT/USER HARM, INJURY, OR DEATH. AN INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE OF THE ISSUE. A FOLLOW-UP REPORT WILL BE PROVIDED UPON INVESTIGATION COMPLETION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21324160 ↗MalfunctionAug 6, 2024Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ ELITE ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A PHILIPS SERVICE ENGINEER REALIGNED THE BUSHING TO RESOLVE THE ISSUE. A THOROUGH ENGINEERING INVESTIGATION HAS BEEN PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19924275 ↗MalfunctionJul 26, 2024Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ ELITE ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A PHILIPS SERVICE ENGINEER REALIGNED THE BUSHING TO RESOLVE THE ISSUE. A THOROUGH ENGINEERING INVESTIGATION HAS BEEN PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19836423 ↗MalfunctionJul 3, 2024Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ ELITE ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY.THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE FIELD SERVICE ENGINEER EVALUATED THE SYSTEM AND REPLACED THE SOLENOID AND CPA BEARING TO REPLACE THE SYSTEM. AN INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE OF THE ISSUE.RESULTS OF THE INVESTIGATION WILL BE INCLUDED IN A FOLLOW UP REPORT UPON ITS COMPLETION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19665408 ↗MalfunctionMay 10, 2024Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ ELITE ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A PHILIPS SERVICE ENGINEER REALIGNED THE BUSHING TO RESOLVE THE ISSUE. A THOROUGH ENGINEERING INVESTIGATION HAS BEEN PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19292095 ↗MalfunctionFeb 28, 2024Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ ELITE ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE PHILIPS SERVICE ENGINEER REPAIRED THE SYSTEM BY RESEATING AND SECURING THE BUSHING. A THOROUGH INVESTIGATION WAS PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS CAUSED BY A HARDWARE ISSUE IN THE ARTICULATING ARM'S LATCHING MECHANISM, WHICH PREVENTED THE LOCKING SOLENOID FROM FULLY ENGAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18797429 ↗MalfunctionJan 23, 2024Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ ELITE ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. THE PHILIPS SERVICE ENGINEER REPAIRED THE SYSTEM BY RESEATING AND SECURING THE BUSHING. A THOROUGH INVESTIGATION WAS PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS CAUSED BY A HARDWARE ISSUE IN THE ARTICULATING ARM'S LATCHING MECHANISM, WHICH PREVENTED THE LOCKING SOLENOID FROM FULLY ENGAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18564480 ↗MalfunctionNov 21, 2023Unintended System Motion; Unintended System Motion
THE CUSTOMER REPORTED THAT THE EPIQ ULTRASOUND SYSTEM CONTROL PANEL WAS SWIVELING AND NOT LOCKING. IT IS UNKNOWN IF THE ISSUE WAS DISCOVERED DURING TRANSPORT. THERE WAS NO INJURY OR CLINICAL USE ASSOCIATED WITH THIS EVENT. A PHILIPS SERVICE ENGINEER CLEANED AND ADJUSTED PIN AND SOLENOID. THE SYSTEM WAS PUT BACK INTO SERVICE WITH NO ADDITIONAL ISSUES REPORTED POST REPAIR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18186054 ↗MalfunctionOct 27, 2023Poor Quality Image; Poor Quality Image
A CUSTOMER REPORTED THEY WERE NOT SEEING THE NEEDLE TIP DURING LIVER/RENAL INTERVENTIONAL PROCEDURES WITH THE C5-1 TRANSDUCER AND ON THE EPIQ ELITE ULTRASOUND. THE PROCEDURE WAS NOT CONTINUED AND WAS RESCHEDULED. A PHILIPS FIELD SERVICE ENGINEER REPLACED THE SYSTEM HARD DRIVES AND UPDATED THE SYSTEM SOFTWARE TO RESOLVE THIS ISSUE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. AN INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE OF THE ISSUE. RESULTS OF THE INVESTIGATION WILL BE INCLUDED IN A FOLLOW UP REPORT UPON ITS COMPLETION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18025397 ↗MalfunctionJun 26, 2023Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE CONTROL PANEL REMAINS UNLOCKED ON THE EPIQ ELITE ULTRASOUND SYSTEM. THE FAILURE OCCURRED WHILE IN CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17200047 ↗MalfunctionMay 19, 2023Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ ELITE ULTRASOUND SYSTEM¿S CONTROL PANEL WAS UNABLE TO LOCK WHEN ATTEMPTING TO TRANSPORT THE UNIT WITHIN THE USER FACILITY. THEY WERE MOVING THE CART BACK INTO A ROOM, AND WHEN THEY WENT TO ADJUST THE POSITION OF THE CONTROL PANEL, IT WOULD NOT LOCK IN PLACE. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A THOROUGH INVESTIGATION WAS PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. A PHILIPS SERVICE ENGINEER DETERMINED THE FAILURE WAS CAUSED BY A GREASE BUILD-UP ON THE SMALL PIN ACTUATOR. THE GREASE BUILD-UP IS NOT ALLOWING THE PIN TO LOCK BACK ON ITS OWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16963368 ↗MalfunctionMay 10, 2023Unintended System Motion; Unintended System Motion
THE CUSTOMER REPORTED THE TOP HALF OF THE EPIQ ELITE ULTRASOUND SYSTEM IS NOT LOCKING, SWIVELS ALL AROUND WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16901060 ↗MalfunctionApr 13, 2023Poor Quality Image; Poor Quality Image
IT WAS REPORTED THERE WAS A SIGNIFICANT CEUS (CONTRAST-ENHANCED ULTRASOUND) PROBLEM IN NEARFIELD ANECHOIC LESIONS WHEN SCANNING WITH THE EPIQ ELITE ULTRASOUND SYSTEM CONTRAST IST ENHANCED SUCH THAT ANECHOIC LESIONS IN THE B MODE ARE FILLED WITH CONTRAST EVEN THOUGH ONLY A SMALL AMOUNT OF CONTRAST 1.0-1.2ML BOLUS IST USED. THE EXAM BEING PERFORMED WAS AN ABDOMEN, CEUS EXAMINATION, SUPINE. THE MEDICAL INDICATION FOR THIS ULTRASOUND WAS UNCLEAR ABDOMEN WITH CYST. THE CYST WOULD HAVE BEEN OVERLOOKED IF NOT KNOWN BEFORE EXAM. THERE WAS NO MEDICAL INTERVENTION, NO CHANGE IN THE PATIENT¿S STATE OF HEALTH. THE ALLEGED INJURY WAS A DIAGNOSIS THAT ARE NOT PRESENTED AS CHARACTERISTIC. THE USER ALLEGED THE PATIENT¿S SAFETY IS AT RISK WHEN INCORRECT DIAGNOSIS IS MADE DUE TO POOR CEUS QUALITY. THE IMAGE ISSUE WAS DESCRIBED AS: THE CEUS CONTRAST IS SO STRONG IN THE NEARFIELD THAT ANECHOIC LESIONS SUDDENLY APPEAR FILLED WITH CONTRAST MEDIUM OR ARE MASSIVELY SUPERIMPOSED WITH ARTIFACTS. THE EXAM WAS NOT COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16731596 ↗MalfunctionFeb 16, 2023Unintended System Motion; Unintended System Motion
THE CUSTOMER REPORTED THE CONTROL PANEL IS NOT ABLE TO LOCK ON THEIR EPIQ ELITE ULTRASOUND SYSTEM. IT IS UNKNOWN AT THIS TIME IF THE SYSTEM WAS MOBILE WHEN FAILURE OCCURRED. THERE IS NO DEATH OR SERIOUS INJURY WAS REPORTED. PHILIPS HAS MADE MULTIPLE ATTEMPTS TO CONTACT THE CUSTOMER FOR FURTHER INFORMATION TO NO AVAIL. IF FURTHER INFORMATION BECOMES AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16386932 ↗MalfunctionDec 15, 2022False Negative Result; False Negative Result
THE CUSTOMER REPORTED A THYROID MASS APPEARED CYSTIC ON THEIR EPIQ ULTRASOUND MACHINES. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. AN INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE OF THE ISSUE. A FOLLOW UP REPORT WILL BE PROVIDED UPON INVESTIGATION COMPLETION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15986352 ↗MalfunctionDec 14, 2022Poor Quality Image; Poor Quality Image
THE CUSTOMER REPORTED AN EPIQ ELITE ULTRASOUND SYSTEM MISCHARACTERIZED THE INTERNAL LESION RESULTING IN A MISDIAGNOSIS OF CANCER IN A CYST. NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. AN INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE OF THE ISSUE. A FOLLOW UP REPORT WILL BE PROVIDED UPON INVESTIGATION COMPLETION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976104 ↗
Other System, imaging, pulsed doppler, ultrasonic devices
- Transducer C5-1 989605412042Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic11 recalls0 reports
- Transducer X5-1 989605400802Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic10 recalls0 reports
- EPIQ 7 Diagnostic Ultrasound System 989605386721Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls2 deaths146 reports
- Transducer S7-3t 989605406772Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic5 recalls1 death18 reports
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS GF TYPE UCT180Olympus Medical Systems Corp. · System, imaging, pulsed doppler, ultrasonic199 reports
- Transducer S7-3t 989605439232Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls4 reports
- Transducer 3D9-3v 989605395622Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls1 report
- ACUSON Redwood Diagnostic Ultrasound System 11503314Siemens Medical Solutions USA, Inc. · System, imaging, pulsed doppler, ultrasonic3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page