EPIQ CVx Ultrasound Circular Edition 795122Recalls & reported problems
Philips Ultrasound LLC · Software Version 9.x · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838113923), received between Sep 3, 2024 and Sep 3, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Unintended System Motion” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended System Motion1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2024Unintended System Motion; Unintended System Motion
A CUSTOMER REPORTED THE SWIVEL MECHANISM OF THEIR EPIQ CVX ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE, AND NO PATIENT OR USER WAS HARMED AS A RESULT OF THE ISSUE. A PHILIPS SERVICE ENGINEER REPAIRED THE SYSTEM BY RESEATING AND SECURING THE BUSHING WHICH ALLOWS THE LOCKING MECHANISM TO BE FULLY ENGAGED. A THOROUGH INVESTIGATION WAS PERFORMED TO IDENTIFY THE ROOT CAUSE OF THE REPORTED ISSUE. THE ENGINEERING TEAM DETERMINED THE FAILURE WAS CAUSED BY A HARDWARE ISSUE IN THE ARTICULATING ARM'S LATCHING MECHANISM, WHICH PREVENTED THE LOCKING SOLENOID FROM FULLY ENGAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20127781 ↗
Other System, Imaging, Pulsed Doppler, Ultrasonic devices
- Transducer C5-1 989605412042Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic11 recalls0 reports
- Transducer X5-1 989605400802Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic10 recalls0 reports
- EPIQ 7 Diagnostic Ultrasound System 989605386721Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls2 deaths146 reports
- Transducer S7-3t 989605406772Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic5 recalls1 death18 reports
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS GF TYPE UCT180Olympus Medical Systems Corp. · System, imaging, pulsed doppler, ultrasonic199 reports
- Transducer S7-3t 989605439232Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls4 reports
- Transducer 3D9-3v 989605395622Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls1 report
- ACUSON Redwood Diagnostic Ultrasound System 11503314Siemens Medical Solutions USA, Inc. · System, imaging, pulsed doppler, ultrasonic3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page