EPIQ Diagnostic Ultrasound System 795231 (UDI 00884838114135)Recalls & reported problems

Philips Ultrasound LLC · EPIQ CVx Diagnostic Ultrasound System Software version 10.0.X · Class 2

All models of EPIQ Diagnostic Ultrasound System (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00884838114135), received between May 20, 2025 and May 4, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 9, 2023Class IICompleted

    Their is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image.

    Z-0283-2024 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Application Program Freezes or Fails to Launch1
  • Mechanical Jam1
  • Unintended System Motion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 4, 2026Application Program Freezes or Fails to Launch

    IT WAS REPORTED THE EPIQ CVX ULTRASOUND SYSTEM, IN USE WITH AN X8-2T TRANSDUCER, DISPLAYED ERROR MESSAGES DURING A VALUE REPLACEMENT SURGERY. THE SYSTEM WAS EXCHANGED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE FIELD SERVICE ENGINEER EVALUATED THE SYSTEM BASED ON THE REPORTED PROBLEM AND DETERMINED THE EVENT WAS RELATED TO THE X8-2T TRANSDUCER. THE SERVICE ENGINEER INITIATED THE REPLACEMENT PROCESS TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25071627 ↗
  • MalfunctionFeb 13, 2026Unintended System Motion

    IT WAS REPORTED THE SWIVEL MECHANISM OF THE EPIQ CVX ULTRASOUND SYSTEM¿S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE UNIT WITHIN THE USER FACILITY. THERE WAS NO PATIENT OR USER HARM AS A RESULT OF THE ISSUE. THE SERVICE ENGINEER VERIFIED THE REPORTED ISSUE. INFORMATION TO REPLACE THE SOLENOID WAS PROVIDED TO THE CUSTOMER TO ADDRESSED THEIR IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24338812 ↗
  • MalfunctionMay 20, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THE X8-2T TRANSDUCER WOULD NOT ARTICULATE DURING USE. THE TRANSDUCER WAS BEING USED WITH AN EPIQ CVX ULTRASOUND SYSTEM. THE TRANSDUCER WAS EXCHANGED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE PHILIPS SERVICE ENGINEER INSPECTED AND TESTED THE TRANSDUCER AND NOT ABLE TO VERIFY THE REPORTED ISSUE. THE TRANSDUCER WAS REPLACED TO RESOLVE THE CUSTOMER¿S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22055481 ↗