EPIQ Diagnostic Ultrasound System 795232 (UDI 00884838114142)Recalls & reported problems
Philips Ultrasound LLC · EPIQ CVxi Diagnostic Ultrasound System Software version 10.0.X · Class 2
All models of EPIQ Diagnostic Ultrasound System (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 2
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00884838114142), received between Mar 26, 2025 and Dec 3, 2025 — classified as 2 deaths, 4 malfunctions.
The most frequently reported specific device problem is “Unintended Application Program Shut Down” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended Application Program Shut Down4
- Application Program Freezes or Fails to Launch2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Cardiac Arrest2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 3, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down
A CUSTOMER REPORTED THEIR EPIQ CVXI ULTRASOUND SYSTEM CRASHED DURING A MITRAL CLIP TREATMENT. THE SYSTEM WAS EXCHANGED TO COMPLETE THE PROCEDURE. THERE WAS NO USER OR PATIENT HARM AS A RESULT OF THE ISSUE. THE SERVICE ENGINEER WAS NOT ABLE TO REPRODUCE THE REPORTED ISSUE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23706885 ↗MalfunctionNov 4, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down
IT WAS REPORTED THE EPIQ CVXI ULTRASOUND SYSTEM WAS NOT AVAILABLE DURING AN INTRACARDIAC ECHOCARDIOGRAPHY (ICE) PROCEDURE WHILE CONNECTED TO AN EXTERNAL MONITOR. THE SYSTEM SHUT DOWN AND WAS REBOOTED SEVERAL TIMES. THERE WAS NO REPORTED PATIENT NOR USER HARM AS A RESULT OF THE ISSUE. A PHILIPS FIELD SERVICE ENGINEER ADVISED THE CUSTOMER TO PLUG IN THE EXTERNAL MONITOR WHILE THE SYSTEM IS OFF TO RESOLVE THE ISSUE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23472700 ↗MalfunctionMay 28, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down
IT WAS REPORTED THE EPIQ CVXI ULTRASOUND SYSTEM WAS NOT AVAILABLE DURING A CRITICAL CORONARY PROCEDURE. THE SYSTEM CRASHED DURING THE PROCEDURE AND WAS UNABLE TO BE RECOVERED. ANOTHER SYSTEM WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22099784 ↗DeathMay 6, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
IT WAS REPORTED THAT THE EPIQ CVXI ULTRASOUND SYSTEM BECAME UNRESPONSIVE AND LOCKED UP DURING AN EXAMINATION WHERE A PATIENT WAS CODING. THE CLINICIAN HAD TO BRING IN ANOTHER SYSTEM IN AN ATTEMPT TO COMPLETE THE EXAMINATION. UNFORTUNATELY, THE PATIENT DID NOT RECOVER FROM THE CODE AND ULTIMATELY EXPIRED; HOWEVER, THE CUSTOMER HAS CLEARLY STATED THE ULTRASOUND DID NOT CAUSE THE PATIENT INCIDENT. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: Cardiac Arrest
MAUDE report 21971899 ↗DeathMay 6, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
IT WAS REPORTED THAT THE EPIQ CVXI ULTRASOUND SYSTEM BECAME UNRESPONSIVE AND LOCKED UP DURING AN EXAMINATION WHERE A PATIENT WAS CODING. THE CLINICIAN HAD TO BRING IN ANOTHER SYSTEM IN AN ATTEMPT TO COMPLETE THE EXAMINATION. UNFORTUNATELY, THE PATIENT DID NOT RECOVER FROM THE CODE AND ULTIMATELY EXPIRED; HOWEVER, THE CUSTOMER HAS CLEARLY STATED THE ULTRASOUND DID NOT CAUSE THE PATIENT INCIDENT. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: Cardiac Arrest
MAUDE report 21971231 ↗MalfunctionMar 26, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down
IT WAS REPORTED THE SYSTEM WAS NOT AVAILABLE DURING A CRITICAL PROCEDURE. THE EPIQ CVX ULTRASOUND SYSTEM SHUT DOWN MULTIPLE TIMES DURING AN UNSPECIFIED OPEN-HEART SURGERY. THE TRANSDUCER WAS EXCHANGED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21700902 ↗
Other System, Imaging, Pulsed Doppler, Ultrasonic devices
- Transducer C5-1 989605412042Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic11 recalls0 reports
- Transducer X5-1 989605400802Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic10 recalls0 reports
- EPIQ 7 Diagnostic Ultrasound System 989605386721Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls2 deaths146 reports
- Transducer S7-3t 989605406772Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic5 recalls1 death18 reports
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS GF TYPE UCT180Olympus Medical Systems Corp. · System, imaging, pulsed doppler, ultrasonic199 reports
- Transducer S7-3t 989605439232Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls4 reports
- Transducer 3D9-3v 989605395622Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls1 report
- ACUSON Redwood Diagnostic Ultrasound System 11503314Siemens Medical Solutions USA, Inc. · System, imaging, pulsed doppler, ultrasonic3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.