EPIQ Diagnostic Ultrasound System 795232 (UDI 00884838114142)Recalls & reported problems

Philips Ultrasound LLC · EPIQ CVxi Diagnostic Ultrasound System Software version 10.0.X · Class 2

All models of EPIQ Diagnostic Ultrasound System (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
2
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00884838114142), received between Mar 26, 2025 and Dec 3, 2025 — classified as 2 deaths, 4 malfunctions.

The most frequently reported specific device problem is “Unintended Application Program Shut Down” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Application Program Shut Down4
  • Application Program Freezes or Fails to Launch2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Cardiac Arrest2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 3, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down

    A CUSTOMER REPORTED THEIR EPIQ CVXI ULTRASOUND SYSTEM CRASHED DURING A MITRAL CLIP TREATMENT. THE SYSTEM WAS EXCHANGED TO COMPLETE THE PROCEDURE. THERE WAS NO USER OR PATIENT HARM AS A RESULT OF THE ISSUE. THE SERVICE ENGINEER WAS NOT ABLE TO REPRODUCE THE REPORTED ISSUE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23706885 ↗
  • MalfunctionNov 4, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down

    IT WAS REPORTED THE EPIQ CVXI ULTRASOUND SYSTEM WAS NOT AVAILABLE DURING AN INTRACARDIAC ECHOCARDIOGRAPHY (ICE) PROCEDURE WHILE CONNECTED TO AN EXTERNAL MONITOR. THE SYSTEM SHUT DOWN AND WAS REBOOTED SEVERAL TIMES. THERE WAS NO REPORTED PATIENT NOR USER HARM AS A RESULT OF THE ISSUE. A PHILIPS FIELD SERVICE ENGINEER ADVISED THE CUSTOMER TO PLUG IN THE EXTERNAL MONITOR WHILE THE SYSTEM IS OFF TO RESOLVE THE ISSUE. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23472700 ↗
  • MalfunctionMay 28, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down

    IT WAS REPORTED THE EPIQ CVXI ULTRASOUND SYSTEM WAS NOT AVAILABLE DURING A CRITICAL CORONARY PROCEDURE. THE SYSTEM CRASHED DURING THE PROCEDURE AND WAS UNABLE TO BE RECOVERED. ANOTHER SYSTEM WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22099784 ↗
  • DeathMay 6, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    IT WAS REPORTED THAT THE EPIQ CVXI ULTRASOUND SYSTEM BECAME UNRESPONSIVE AND LOCKED UP DURING AN EXAMINATION WHERE A PATIENT WAS CODING. THE CLINICIAN HAD TO BRING IN ANOTHER SYSTEM IN AN ATTEMPT TO COMPLETE THE EXAMINATION. UNFORTUNATELY, THE PATIENT DID NOT RECOVER FROM THE CODE AND ULTIMATELY EXPIRED; HOWEVER, THE CUSTOMER HAS CLEARLY STATED THE ULTRASOUND DID NOT CAUSE THE PATIENT INCIDENT. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: Cardiac Arrest

    MAUDE report 21971899 ↗
  • DeathMay 6, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    IT WAS REPORTED THAT THE EPIQ CVXI ULTRASOUND SYSTEM BECAME UNRESPONSIVE AND LOCKED UP DURING AN EXAMINATION WHERE A PATIENT WAS CODING. THE CLINICIAN HAD TO BRING IN ANOTHER SYSTEM IN AN ATTEMPT TO COMPLETE THE EXAMINATION. UNFORTUNATELY, THE PATIENT DID NOT RECOVER FROM THE CODE AND ULTIMATELY EXPIRED; HOWEVER, THE CUSTOMER HAS CLEARLY STATED THE ULTRASOUND DID NOT CAUSE THE PATIENT INCIDENT. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: Cardiac Arrest

    MAUDE report 21971231 ↗
  • MalfunctionMar 26, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down

    IT WAS REPORTED THE SYSTEM WAS NOT AVAILABLE DURING A CRITICAL PROCEDURE. THE EPIQ CVX ULTRASOUND SYSTEM SHUT DOWN MULTIPLE TIMES DURING AN UNSPECIFIED OPEN-HEART SURGERY. THE TRANSDUCER WAS EXCHANGED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21700902 ↗

Other System, Imaging, Pulsed Doppler, Ultrasonic devices

All System, Imaging, Pulsed Doppler, Ultrasonic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page