Ingenia Ingenia Elition X (REF 782158)Recalls & reported problems

Philips Medical Systems Nederland B.V. · Enables trained physicians to obtain cross-sectional and spectroscopic images of internal structures of the head, body or extremities, in any orientation, representing the spatial distribution of protons or other nuclei with spin. · Class 2

All models of Ingenia (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 28, 2026Class IIOpen, Classified

    When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

    Z-3311-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Aug 21, 2026Class IIOpen, Classified

    A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

    Z-3239-2026 · Root cause: Software design · Recall details · FDA record ↗

Other System, Nuclear Magnetic Resonance Imaging devices

All System, Nuclear Magnetic Resonance Imaging devices →