Ingenia recallZ-3311-2026 · Philips North America · initiated Aug 28, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
FDA root cause category: Software design
What the company is doing
On August 28, 2026, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. To prevent the issue from occurring while Philips is developing the software solution, use the Classic DICOM transfer option for exporting DTI scans. 2. You may continue to use your system(s) in accordance with the intended use. 3. Circulate this notice to all users of this device so that they are aware of the issue and associated hazard/harm. 4. Please retain this Urgent Medical Device Correction letter with your system(s) until the software upgrade is installed; ensure the notice is in a place likely to be seen/viewed.
Recall details
- Recall number
- Z-3311-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 28, 2026
- Posted by the FDA
- Sep 25, 2026
- Recalling firm
- Philips North America
- Quantity
- 340
- Product code
- LNH
- FDA submissions
- K193215, K213516, K223458, K230972, K251397, K251808
- Distribution
- Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.
Product description
Ingenia Elition S and Ingenia Elition X.; Model number (REF): 1. 781357 (S), 2. 782106 (S), 3. 782137 (S), 4. 782157 (S), 5. 781358 (X), 6. 782107 (X), 7. 782119 (X), 8. 782136 (X), 9. 782158 (X), 10. 782198 (X); Software Versions: R11, R12 and R13;
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
1. Model number (REF): 781357 (S); UDI-DI: 00884838088108; Serial Numbers: 46002, 46012, 46013, 46017, 46018, 46020, 46021, 46023, 46025, 46027, 46029, 46036, 46039, 46041, 46042, 46044, 46053, 46054, 46082, 46086, 46092; 2. Model Number (REF): 782106 (S); UDI-DI: 00884838098329; Serial Numbers: 46203, 46204, 46207, 46208, 46209, 46219, 46220, 46225, 46235, 46243, 46249, 46258, 46269, 46270, 46286, 46302, 46315, 46325, 46375; 3. Model Number (REF): 782137 (S); UDI-DI: 00884838108615; Serial Numbers: 64001, 64002, 64003, 64004, 64005, 64014, 64015, 64029, 64032, 64039, 64040, 64041, 64042, 64049, 64050, 64051, 64052; 4. Model Number (REF): 782157 (S); UDI-DI: 00884838115682; Serial Numbers: 8900, 89003; 5. Model Number (REF): 781358 (X); UDI-DI: 00884838088115; Serial Numbers: 45008, 45009, 45010, 45011, 45024, 45026, 45028, 45029, 45032, 45033, 45034, 45035, 45036, 45037, 45040, 45041, 45046, 45048, 45049, 45054, 45055, 45056, 45060, 45062, 45066, 45070, 45072, 45073, 45075, 45076, 45078, 45086, 45087, 45089, 45090, 45091, 45093, 45094, 45096, 45097, 45098, 45099, 45100, 45105, 45108, 45112, 45113, 45115, 45116, 45119, 45120, 45127, 45129, 45132, 45136, 45138, 45140, 45146, 45147, 45148, 45152, 45154, 45156, 45157, 45160, 45162, 45168, 45174, 45181, 45186, 45187, 45189, 45190, 45191, 45194, 45205, 45215, 45231, 45235, 45237, 45243, 45245, 45246, 45250, 45252, 45264, 45272, 45277, 45278, 45290, 45296, 45316, 45317, 45319, 45323, 45328, 45333, 45335, 45340, 45342, 45374, 45379, 45395, 45398, 45399, 45406, 45407, 45414, 45423, 45441; 6. Model Number (REF): 782107 (X); UDI-DI: 00884838098336; Serial Numbers: 45500, 45503, 45506, 45507, 45508, 45509, 45510, 45514, 45519, 45523, 45525, 45530, 45532, 45533, 45534, 45536, 45542, 45545, 45547, 45551, 45560, 45561, 45573, 45574, 45578, 45579, 45589, 45594, 62019, 62024, 62026, 62030, 62031, 62033, 62035, 62037, 62040, 62041, 62048, 62051, 62052, 62056, 62058, 62059, 62061, 62062, 62063, 62065, 62067, 62070, 62071, 62073, 62080, 62081, 62082, 62088, 62092, 62098, 62134, 62136, 62137, 62148, 62151, 62152, 62153, 62159, 62178, 62181, 62182, 62197; 7. Model Number (REF): 782119 (X); UDI-DI: 00884838104129; Serial Number: 66001; 8. Model Number (REF): 782136 (X); UDI-DI: 00884838108608; Serial Numbers: 28500, 28502, 28503, 28504, 28506, 28507, 28509, 28510, 28514, 28515, 28516, 28521, 28524, 28525, 28526, 28527, 28528, 28529, 28532, 28533, 28535, 28536, 28538, 28540, 28542, 28543, 28544, 28546, 28547, 28548, 28549, 28550, 28555, 28558, 28559, 28560, 28563, 28564, 28565, 28566, 28568, 28572, 28573, 28576, 28577, 28579, 28581, 28583, 28585, 28586, 28587, 28591, 28592, 28593, 28594, 28595, 28596, 28597, 28598, 28599, 28604, 28605, 28608, 28611, 28612, 28613, 28614, 28615, 28616, 28618, 28619, 28623, 28625, 28628, 28629, 28630, 28631, 28632, 28635, 28636, 28638; 9. Model Number (REF): 782158 (X); UDI-DI: 00884838115699; Serial Numbers: 90001, 90002, 90004, 90006, 90007, 90010, 90011, 90015, 90016, 90017, 90018, 90023, 90024, 90027, 90036; 10. Model Number (REF): 782198 (X); UDI-DI: 00884838128460; Serial Numbers: 57003, 57004, 57005, 57009;
Devices named in this recall
10 devices in the FDA device registry whose identifier appears in the recall's code information.
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging14 recalls12 reports
- Ingenia Elition SPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging13 recalls2 reports
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls5 reports
- Ingenia Ingenia Elition SPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging11 recalls3 reports
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, Nuclear Magnetic Resonance Imaging10 recalls1 report
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging7 recalls0 reports
- Ingenia Ingenia Elition SPhilips Medical Systems Nederland B.V. · System, Nuclear Magnetic Resonance Imaging7 recalls0 reports
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, Nuclear Magnetic Resonance Imaging2 recalls0 reports
- Ingenia Ingenia Elition SPhilips Medical Systems Nederland B.V. · System, Nuclear Magnetic Resonance Imaging1 recall0 reports
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, Nuclear Magnetic Resonance Imaging1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99834).
- Z-3307-2026 · Open, ClassifiedIngenia Blue Seal SE and Blue Seal XE; Model number (REF): 1. 782185 (SE), 2. 782192 (XE); Software Versions: R11, R12 and R13;
- Z-3308-2026 · Open, ClassifiedIngenia 1.5T, 1.5T CX, and 1.5T S; Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), 4. 782115 (1.5T), 5. 782140 (1.5T), 6. 781262 (1.5T CX), 7. 781347 (1.5…
- Z-3309-2026 · Open, ClassifiedIngenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software Versions: R11, R12 and R13;
- Z-3310-2026 · Open, ClassifiedIngenia Ambition S and Ingenia Ambition X.; Model number (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 782159 (S), 5. 782194 (S), 6. 781356 (X), 7. 782109 (X), 8. 782138 (…
- Z-3312-2026 · Open, ClassifiedMR 5300; Model number (REF): 1. 782110, 2. 782152, 3. 782156, 4. 782196; Software Versions: R11, R12 and R13;
- Z-3313-2026 · Open, ClassifiedMR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Software Versions: R11, R12 and R13;
- Z-3314-2026 · Open, ClassifiedSmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, and SmartPath to Ingenia Elition X; Model number (REF): 1. 781260 (1.5T), 2. 78…
- Z-3315-2026 · Open, ClassifiedSmartPath to MR 7700; Model number (REF): 1. 782178, 2. 782207; Software Versions: R11, R12 and R13;
- Z-3316-2026 · Open, ClassifiedUpgrade dStream; Model number (REF): 782127; Software Versions: R11, R12 and R13;
- Z-3317-2026 · Open, ClassifiedUpgrade to MR 7700; Model number (REF): 782130; Software Versions: R11, R12 and R13;
- Z-3318-2026 · Open, ClassifiedEvolution upgrade 1.5T amd Evolution upgrade 3.0T; Model number (REF): 1. 782116 (1.5T), 2. 782148 (1.5T), 3. 782166 (1.5T), 4. 782201(1.5T), 5. 782117 (3.0T), 6. 782143 (3.0T), 7.…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page