Skip to content

Ultrasound System 5500W 795144Recalls & reported problems

Philips Ultrasound LLC · Compact Series Ultrasound System · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 23, 2024Class IIOpen, Classified

    Ultrasound system with: 1) transesophageal echocardiography transducer (TEE) connected/not selected 2) 1, 2 non-TEE connected 3) Imaging with non-TEE 4) Selected/not selected imaging transducer removed 5) Transducer connected 6) Non-TEE imaging continues, Then though TEE is unselected, it may have power, no temperature monitoring/control, and may result in temperature above limit and patient burns

    Z-1757-2024 · Root cause: Component change control · Recall details · FDA record ↗

Other System, Imaging, Pulsed Doppler, Ultrasonic devices

All System, Imaging, Pulsed Doppler, Ultrasonic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page